Simultaneous Hip Head-Stem Taper Junction Measurements of Electrochemical Corrosion and Micromotion: A Comparison of Taper Geometry and Stem Material

被引:2
|
作者
Swaminathan, V. [1 ]
Scholl, L. [1 ]
Lee, R. [1 ]
Faizan, A. [1 ]
Thakore, M. [1 ]
TenHuisen, K. [1 ]
Nevelos, J. [1 ]
机构
[1] Stryker Orthopaed, 325 Corp Dr, Mahwah, NJ 07430 USA
关键词
corrosion; micromotion; electrochemistry; onset load; tapers; FRETTING CORROSION; CREVICE CORROSION; INTERFACE;
D O I
10.1520/STP159120140150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fretting corrosion or mechanically assisted crevice corrosion is a current topic of research in total hip arthroplasty highlighted by the modularity of the head-neck taper junction. The extent of corrosion may be influenced by a multitude of factors, and in this study, we examined two specific variables: material combination and taper geometry. Corrosion performance was quantitatively assessed by measuring fretting corrosion at the taper junction as well as relative micromotion between the head and stem components. To examine the effect of material combinations, stems composed of either Ti-6AI-4V or TMZF (R) were tested with CoCrMo femoral heads. To study taper geometry, a V40 (TM) taper design was compared with a C-taper design. Cyclic fatigue loading was applied to the test specimens, increasing load from 100 to 3200 N and then continuing loading to 1 million cycles. All tests were performed in phosphate-buffered saline solution at room temperature and electrochemical data analysis included determination of onset load, average current, average open-circuit potential, peak-to-peak current, and absolute change in potential. Micromotion fixtures were attached rigidly to the head and proximal stem. The micromotion output was used to calculate pistoning of the head along the stem, rocking of the head about the stem, and subsidence (or seating or slipping) of the head against the stem. The electrochemical data showed that the mean onset load for different groups was in the range of 1075 to 1194 N and average currents up to 2 mu A were recorded for different sample groups during cyclic loading. Fretting corrosion and micromotion measurements in this study showed that the variation produced by material combination and taper design within the range of test groups tested were less than the variation that existed in each test group.
引用
收藏
页码:321 / 335
页数:15
相关论文
共 48 条
  • [1] Influence of flexural rigidity on micromotion at the head-stem taper interface of modular hip prostheses
    Haschke, Henning
    Konow, Tobias
    Huber, Gerd
    Morlock, Michael M.
    MEDICAL ENGINEERING & PHYSICS, 2019, 68 : 1 - 10
  • [2] Damage Patterns at the Head-Stem Taper Junction Helps Understand the Mechanisms of Material Loss
    Hothi, Harry S.
    Panagiotopoulos, Andreas C.
    Whittaker, Robert K.
    Bills, Paul J.
    McMillan, Rebecca A.
    Skinner, John A.
    Hart, Alister J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (01): : 291 - 295
  • [3] A computational approach to fretting wear prediction at the head-stem taper junction of total hip replacements
    English, Russell
    Ashkanfar, Ariyan
    Rothwell, Glynn
    WEAR, 2015, 338 : 210 - 220
  • [4] Micromotion at the head-stem taper junction of total hip prostheses is influenced by prosthesis design-, patient- and surgeon-related factors
    Falkenberg, Adrian
    Biller, Sara
    Morlock, Michael M.
    Huber, Gerd
    JOURNAL OF BIOMECHANICS, 2020, 98 (98)
  • [5] Quantification of the Contact Area at the Head-Stem Taper Interface of Modular Hip Prostheses
    Witt, Florian
    Guehrs, Julian
    Morlock, Michael M.
    Bishop, Nicholas E.
    PLOS ONE, 2015, 10 (08):
  • [6] Geometric Variations of Modular Head-Stem Taper Junctions of Total Hip Replacements
    Wade, A.
    Beadling, A. R.
    Neville, A.
    De Villiers, D.
    Cullum, C. J.
    Collins, S.
    Bryant, M. G.
    MEDICAL ENGINEERING & PHYSICS, 2020, 83 : 34 - 47
  • [7] Method for Characterization of Material Loss from Modular Head-Stem Taper Surfaces of Hip Replacement Devices
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Hart, Alister
    Skinner, John
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 132 - 146
  • [8] The effect of using components from different manufacturers on the rate of wear and corrosion of the head-stem taper junction of metal-on-metal hip arthroplasties
    Whittaker, R. K.
    Hothi, H. S.
    Meswania, J. M.
    Berber, R.
    Blunn, G. W.
    Skinner, J. A.
    Hart, A. J.
    BONE & JOINT JOURNAL, 2016, 98B (07): : 917 - 924
  • [9] A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement
    Bitter, Thom
    Khan, Imran
    Marriott, Tim
    Lovelady, Elaine
    Verdonschot, Nico
    Janssen, Dennis
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (09) : 862 - 870
  • [10] Evaluation of surgical impaction technique and how it affects locking strength of the head-stem taper junction
    Scholl, Laura
    Schmidig, Gregg
    Faizan, Ahmad
    TenHuisen, Kevor
    Nevelos, Jim
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (07) : 661 - 667