Dosimetric comparison of peripheral NSCLC SBRT using Acuros XB and AAA calculation algorithms

被引:15
|
作者
Ong, Chloe C. H. [1 ]
Ang, Khong Wei [2 ]
Soh, Roger C. X. [3 ]
Tin, Kah Ming [2 ]
Yap, Jerome H. H. [2 ]
Lee, James C. L. [2 ,3 ]
Bragg, Christopher M. [1 ]
机构
[1] Sheffield Hallam Univ, Fac Hlth & Wellbeing, Sheffield, S Yorkshire, England
[2] Natl Canc Ctr Singapore, Div Radiat Oncol, Singapore, Singapore
[3] Nanyang Technol Univ, Sch Phys & Math Sci, Div Phys & Appl Phys, Singapore, Singapore
关键词
SBRT; AXB; AAA; Calculation algorithms; NSCLC; DOSE CALCULATION; RADIATION-THERAPY; LUNG; CANCER;
D O I
10.1016/j.meddos.2017.05.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
There is a concern for dose calculation in highly heterogenous environments such as the thorax region. This study compares the quality of treatment plans of peripheral non small cell lung cancer (NSCLC) stereotactic body radiation therapy (SBRT) using 2 calculation algorithms, namely, Eclipse Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB), for 3-dimensional conformal radiation therapy (3DCRT) and volumetric-modulated arc therapy (VMAT). Four-dimensional computed tomography (4DCT) data from 20 anonymized patients were studied using Varian Eclipse planning system, AXB, and MA version 10.0.28. A 3DCRT plan and a VMAT plan were generated using AAA and AXB with constant plan parameters for each patient. The prescription and dose constraints were benchmarked against Radiation Therapy Oncology Group (RTOG) 0915 protocol. Planning parameters of the plan were compared statistically using Mann-Whitney U tests. Results showed that 3DCRT and VMAT plans have a lower target coverage up to 8% when calculated using AXB as compared with AAA. The conformity index (CI) for AXB plans was 4.7% lower than AAA plans, but was closer to unity, which indicated better target conformity. AXB produced plans with global maximum doses which were, on average, 2% hotter than AAA plans. Both 3DCRT and VMAT plans were able to achieve D95%. VMAT plans were shown to be more conformal (CI = 1.01) and were at least 3.2% and 1.5% lower in terms of PTV maximum and mean dose, respectively. There was no statistically significant difference for doses received by organs at risk (OARs) regardless of calculation algorithms and treatment techniques. In general, the difference in tissue modeling for AXB and AAA algorithm is responsible for the dose distribution between the AXB and the AAA algorithms. The AXB VMAT plans could be used to benefit patients receiving peripheral NSCLC SBRT. (C) 2017 American Association of Medical Dosimetrists.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 50 条
  • [41] Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water
    Fogliata, Antonella
    Nicolini, Giorgia
    Clivio, Alessandro
    Vanetti, Eugenio
    Mancosu, Pietro
    Cozzi, Luca
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (06): : 1879 - 1904
  • [42] Dosimetric accuracy and clinical quality of Acuros XB and AAA dose calculation algorithm for stereotactic and conventional lung volumetric modulated arc therapy plans
    Petra S Kroon
    Sandra Hol
    Marion Essers
    Radiation Oncology, 8
  • [43] Dosimetric accuracy and clinical quality of Acuros XB and AAA dose calculation algorithm for stereotactic and conventional lung volumetric modulated arc therapy plans
    Kroon, Petra S.
    Hol, Sandra
    Essers, Marion
    RADIATION ONCOLOGY, 2013, 8
  • [44] Virtual bronchoscopy-guided lung SAbR: dosimetric implications of using AAA versus Acuros XB to calculate dose in airways
    Kinkopf, P.
    Modiri, A.
    Yu, Kun-Chang
    Yan, Y.
    Mohindra, P.
    Timmerman, R.
    Sawant, A.
    Vicente, E.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (06)
  • [45] Accuracy of Acuros XB and AAA dose calculation for small fields with reference to RapidArc® stereotactic treatments
    Fogliata, Antonella
    Nicolini, Giorgia
    Clivio, Alessandro
    Vanetti, Eugenio
    Cozzi, Luca
    MEDICAL PHYSICS, 2011, 38 (11) : 6228 - 6237
  • [46] New Dosimetric Guidelines for Acuros Through Comparative Evaluation of SBRT Lung Treatment Plans Against AAA
    Pacella, M.
    Webster, M.
    Tanny, S.
    Joyce, N.
    Constine, L.
    Milano, M.
    Yeo, I.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [47] Evaluation of Acuros XB Algorithm for Lung SBRT Using Flattening Filter Free
    Sandrini, E.
    da Silva, A.
    Silva, C.
    MEDICAL PHYSICS, 2017, 44 (06) : 2963 - 2963
  • [48] Validation of Photon Dose Calculation Using Mobius3D System Compared to AAA and Acuros XB Systems
    Majithia, L.
    DiCostanzo, D.
    Weldon, M.
    Gupta, N.
    Rong, Y.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [49] Comparison between Ethos measurements and dosimetric parametres calculated with Acuros XB Beam Model
    Camba, Francisco
    Honorato, Sergio
    Perez-Gomez, Nuria
    Sancho-Kolster, Ismael
    Martin-Vaello, Roberto
    de la Cruz, Manuel
    Pages-Perez, Alex
    Picon, Cristina
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3430 - S3433
  • [50] Evaluating small field dosimetry with the Acuros XB (AXB) and analytical anisotropic algorithm (AAA) dose calculation algorithms in the eclipse treatment planning system
    Behinaein, Sepideh
    Osei, Ernest
    Darko, Johnson
    Charland, Paule
    Bassi, Dylan
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (04) : 353 - 364