Dosimetric comparison of peripheral NSCLC SBRT using Acuros XB and AAA calculation algorithms

被引:15
|
作者
Ong, Chloe C. H. [1 ]
Ang, Khong Wei [2 ]
Soh, Roger C. X. [3 ]
Tin, Kah Ming [2 ]
Yap, Jerome H. H. [2 ]
Lee, James C. L. [2 ,3 ]
Bragg, Christopher M. [1 ]
机构
[1] Sheffield Hallam Univ, Fac Hlth & Wellbeing, Sheffield, S Yorkshire, England
[2] Natl Canc Ctr Singapore, Div Radiat Oncol, Singapore, Singapore
[3] Nanyang Technol Univ, Sch Phys & Math Sci, Div Phys & Appl Phys, Singapore, Singapore
关键词
SBRT; AXB; AAA; Calculation algorithms; NSCLC; DOSE CALCULATION; RADIATION-THERAPY; LUNG; CANCER;
D O I
10.1016/j.meddos.2017.05.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
There is a concern for dose calculation in highly heterogenous environments such as the thorax region. This study compares the quality of treatment plans of peripheral non small cell lung cancer (NSCLC) stereotactic body radiation therapy (SBRT) using 2 calculation algorithms, namely, Eclipse Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB), for 3-dimensional conformal radiation therapy (3DCRT) and volumetric-modulated arc therapy (VMAT). Four-dimensional computed tomography (4DCT) data from 20 anonymized patients were studied using Varian Eclipse planning system, AXB, and MA version 10.0.28. A 3DCRT plan and a VMAT plan were generated using AAA and AXB with constant plan parameters for each patient. The prescription and dose constraints were benchmarked against Radiation Therapy Oncology Group (RTOG) 0915 protocol. Planning parameters of the plan were compared statistically using Mann-Whitney U tests. Results showed that 3DCRT and VMAT plans have a lower target coverage up to 8% when calculated using AXB as compared with AAA. The conformity index (CI) for AXB plans was 4.7% lower than AAA plans, but was closer to unity, which indicated better target conformity. AXB produced plans with global maximum doses which were, on average, 2% hotter than AAA plans. Both 3DCRT and VMAT plans were able to achieve D95%. VMAT plans were shown to be more conformal (CI = 1.01) and were at least 3.2% and 1.5% lower in terms of PTV maximum and mean dose, respectively. There was no statistically significant difference for doses received by organs at risk (OARs) regardless of calculation algorithms and treatment techniques. In general, the difference in tissue modeling for AXB and AAA algorithm is responsible for the dose distribution between the AXB and the AAA algorithms. The AXB VMAT plans could be used to benefit patients receiving peripheral NSCLC SBRT. (C) 2017 American Association of Medical Dosimetrists.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 50 条
  • [21] Dosimetric Validation of Acuros XB Algorithm for Photon Dose Calculation in Water
    Kumar, L.
    Yadav, G.
    Kishore, V.
    Bhushan, M.
    Samuvel, K.
    Suhail, M.
    MEDICAL PHYSICS, 2016, 43 (06) : 3562 - 3562
  • [22] COMPARISON OF ACUROS XB AND AAA DOSE CALCULATIONS FOR ADVANCED LUNG TREATMENTS
    Vanetti, E.
    Fogliata, A.
    Clivio, A.
    Nicolini, G.
    Cozzi, L.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (08) : 516 - 516
  • [23] Comparison between Acuros XB and Brainlab Monte Carlo algorithms for photon dose calculation
    Misslbeck, M.
    Kneschaurek, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (07) : 599 - 605
  • [24] Acuros XB and AAA Dose Calculation Accuracy for Small Fields in RapidArc Treatments
    Fogliata, A.
    Nicolini, G.
    Clivio, A.
    Vanetti, E.
    Cozzi, L.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [25] Dose Calculation Accuracy Comparison of BrainLab Monte Carlo and Eclipse Acuros XB for SRS/SBRT Planning
    Yoon, J.
    Gray, A.
    MEDICAL PHYSICS, 2020, 47 (06) : E748 - E748
  • [26] Dosimetric Validation for Pinnacle, Acuros, AAA, and Brainlab Algorithms with Induced Inhomogenieties
    Lopez, P.
    LaFontaine, R.
    Burns, L.
    Tambasco, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 254 - 254
  • [27] Dosimetric impact of Acuros XB dose calculation algorithm in prostate cancer treatment using RapidArc
    Rana, Suresh
    Rogers, Kevin
    Lee, Terry
    Reed, Daniel
    Biggs, Christopher
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2013, 9 (03) : 430 - 435
  • [28] ADVANCED LUNG TREATMENT USING DIFFERENT MODALITIES: COMPARISON OF ACUROS XB AND AAA DOSE CALCULATIONS
    Fogliata, A.
    Clivio, A.
    Nicolini, G.
    de Palma, E. Vanetti
    Cozzi, L.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S575 - S575
  • [29] Effects of metal implants and a metal artifact reduction tool on calculation accuracy of AAA and Acuros XB algorithms in small fields
    Akdeniz, Yucel
    Yegingil, Ilhami
    Yegingil, Zehra
    MEDICAL PHYSICS, 2019, 46 (11) : 5326 - 5335
  • [30] Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media
    Fogliata, Antonella
    Nicolini, Giorgia
    Clivio, Alessandro
    Vanetti, Eugenio
    Cozzi, Luca
    RADIATION ONCOLOGY, 2011, 6