A Monte Carlo calculation of dosimetric parameters of 90Sr/90Y and 192Ir SS sources for intravascular brachytherapy

被引:53
|
作者
Wang, RQ [1 ]
Li, XA [1 ]
机构
[1] Univ Maryland, Dept Radiat Oncol, Baltimore, MD 21201 USA
关键词
Monte Carlo; dosimetry; intravascular brachytherapy;
D O I
10.1118/1.1319374
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The AAPM TG-60 report has proposed various dose calculation parameters for intravascular brachytherapy (IVBT). These parameters include the dose rate constant (or the dose rate at a reference position for a beta-particle emitting source), the radial dose function, and the anisotropy function. In this work, we have used a modified EGS4 Monte Carlo system to calculate these parameters for the two most commonly used IVBT sources (the beta-particle emitting Sr-90/Y-90 source and the photon emitting Ir-192 SS source). To ensure the calculation accuracy, the present calculation was compared with several measurements and calculations reported by other authors. Excellent agreement was found for the results with the photon source. For beta-particle source calculation, the present results for a variety of point sources agree very well with a previous work. The presently calculated radial dose functions for the Sr-90/Y-90 source are consistent with those of a published work for intermediate radial distances. The dose uniformity in the axial direction was also studied. The contributions of bremsstrahlung photons to total doses for the Sr-90/Y-90 beta source, and the influence of ignoring electron transport on calculated doses for the Ir-192 SS photon source are discussed. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01211-6].
引用
收藏
页码:2528 / 2535
页数:8
相关论文
共 50 条
  • [41] The effects of non-centred catheter and guidewire on the dose distribution around source in catheter-based intravascular brachytherapy with 90Sr/90Y beta source
    Demir, B
    Demir, M
    Ahmed, AS
    RADIATION MEASUREMENTS, 2006, 41 (03) : 317 - 322
  • [42] Experimental determination of the radial dose function of 90Sr/90Y IVBT sources (vol 33, pg 3225, 2006)
    Holmes, Shannon M.
    DeWerd, Larry A.
    Micka, John A.
    MEDICAL PHYSICS, 2007, 34 (02) : 790 - 790
  • [43] Calculation of Dosimetry Parameters for 192Ir and 125I Brachytherapy Sources Using Geant4
    Fonseca-Rodrigues, Sane S. O.
    Martins, Maximiano C.
    Begalli, Marcia
    Queiroz Filho, Pedro P.
    de Souza-Santos, Denison
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 1190 - 1192
  • [44] EXPOSURE INHOMOGENEITY FROM 241AM AND 90SR/90Y SOURCES IN TERMS OF THE EYE LENS MONITORING IN THE NUCLEAR FACILITIES
    Yoshitomi, H.
    Kowatari, M.
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (02) : 191 - 198
  • [45] Monte Carlo simulation for assessing absorbed dose effects of low-dose β-radiation (90Sr/90Y) on cytotoxicity and apoptotic death in K562 cells
    Alizadeh, Zeynab
    Khodabakhsh, Rasoul
    Mohammadzadeh, Mehdi
    Pazhang, Yaghub
    Mohammadi, Seyed Mostafa
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2022, 18 (01) : 200 - 208
  • [46] Anisotropy function for 192Ir low-dose-rate brachytherapy sources:: an EGS4 Monte Carlo study
    Capote, R
    Mainegra, E
    López, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05): : 1487 - 1499
  • [47] GATE MONTE CARLO DOSIMETRY IN 90Y TARE PLANNING: INFLUENCE OF SIMULATION PARAMETERS AND IMAGE RESAMPLING ON DOSIMETRIC ACCURACY AND OPTIMIZATION OF COMPUTATIONAL TIMES
    Pistone, Daniele
    Italiano, Antonio
    Auditore, Lucrezia
    Amato, Ernesto
    Campenni, Alfredo
    Baldari, Sergio
    ATTI ACCADEMIA PELORITANA DEI PERICOLANTI-CLASSE DI SCIENZE FISICHE MATEMATICHE E NATURALI, 2021, 99 (02):
  • [48] Verification and uniformity control of doses for Sr-90/ Y-90 intravascular brachytherapy sources using radiochromic film dosimetry
    Demir, Bayram
    Ahmed, Asm Sabbir
    Babalik, Erhan
    Demir, Mustafa
    Gurmen, Tevfi K.
    JOURNAL OF MEDICAL PHYSICS, 2008, 33 (02) : 54 - 59
  • [49] Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study
    Ranjbar, Sahar
    Gholami, Somayeh
    Longo, Francesco
    Sharp, Ryan
    Meigooni, Ali S.
    APPLIED RADIATION AND ISOTOPES, 2020, 159
  • [50] Rapid separation, purification of 90Sr and feasibility studies for the preparation of 90Y resin microsphere brachytherapy source from FBTR irradiated fuel dissolver solution
    Saha, Debasish
    Vithya, J.
    Kalaiarasu, T.
    Bola Sankar, D.
    Rajeswari, S.
    Arulkumari, A.
    Manoravi, P.
    Sundararajan, K.
    Jayaraman, V.
    Sivaraman, N.
    Venkatraman, B.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2024, 333 (04) : 2027 - 2038