Total body irradiation using helical tomotherapy: Set-up experience and in-vivo dosimetric evaluation

被引:9
|
作者
Cleuziou, J. P. [1 ]
Desgranges, C. [1 ]
Henry, I [1 ]
Jaumot, M. [1 ]
Chartier, P. [1 ]
Sihanath, R. [1 ]
Carre, M. [2 ]
Bulabois, C. E. [2 ]
Cahn, J. Y. [2 ]
Pasteris, C. [1 ]
Balosso, J. [1 ]
Gabelle-Flandin, I [1 ]
Verry, C. [1 ]
Giraud, J. Y. [1 ]
机构
[1] Ctr Hosp Univ Grenoble Alpes CHUGA, Serv Radiotherapie, CS 10217, Grenoble 9, France
[2] Ctr Hosp Univ Grenoble Alpes CHUGA, Serv Hematol, CS 10217, Grenoble 9, France
来源
CANCER RADIOTHERAPIE | 2021年 / 25卷 / 03期
关键词
Helical tomotherapy; Total body irradiation; FEASIBILITY; TBI;
D O I
10.1016/j.canrad.2020.07.009
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose. - Helical Tomotherapy (HT) appears as a valuable technique for total body irradiation (TBI) to create highly homogeneous and conformal dose distributions with more precise repositioning than conventional TBI techniques. The aim of this work is to describe the technique implementation, including treatment preparation, planning and dosimetric monitoring of TBI delivered in our institution from October 2016 to March 2019. Material and method. - Prior to patient care, irradiation protocol was set up using physical phantoms. Gafchromic films were used to assess dose distribution homogeneity and evaluate imprecise patient positioning impact. Sixteen patients' irradiations with a prescribed dose of 12 Gy were delivered in 6 fractions of 2 Gy over 3 days. Pre-treatment quality assurance (QA) was performed for the verification of dose distributions at selected positions. In addition, in-vivo dosimetry was carried out using optically stimulated luminescence dosimeters (OSLD). Results. - Planning evaluation, as well as results of pre-treatment verifications, are presented. In-vivo dosimetry showed the strong consistency of OSLD measured doses. OSLD mean relative dose differences between measurement and calculation were respectively +0,96% and -2% for armpit and hands locations, suggesting better reliability for armpit OSLD positioning. Repercussion of both longitudinal and transversal positioning inaccuracies on phantoms is depicted up to 2 cm shifts. Conclusion. - The full methodology to set up TBI protocol, as well as dosimetric evaluation and pretreatment QA, were presented. Our investigations reveal strong correspondence between planned and delivered doses shedding light on the dose reliability of OSLD for HT based TBI in-vivo dosimetry. (C) 2020 Societe francaise de radiotherapie oncologique (SFRO). Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:213 / 221
页数:9
相关论文
共 50 条
  • [1] Total body irradiation using Helical Tomotherapy®: Treatment technique, dosimetric results and initial clinical experience
    Sarradin, V.
    Simon, L.
    Huynh, A.
    Gilhodes, J.
    Filleron, T.
    Izar, F.
    CANCER RADIOTHERAPIE, 2018, 22 (01): : 17 - 24
  • [2] Dosimetric Evaluation of Total Body Irradiations with Helical Tomotherapy: Phantom Study
    Ucar Alveroglu, Gokce
    Kingir Celtik, Inci
    Koksal Akbas, Canan
    Becerir, Hatice Bilge
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2023, 38 (01): : 104 - 111
  • [3] Dosimetric verification of helical tomotherapy for total scalp irradiation
    Hardcastle, Nicholas
    Soisson, Emilie
    Metcalfe, Peter
    Rosenfeld, Anatoly B.
    Tome, Wolfgang A.
    MEDICAL PHYSICS, 2008, 35 (11) : 5061 - 5068
  • [4] A dosimetric comparison of conventional radiotherapy and helical tomotherapy (HT) for total body irradiation (TBI)
    Faught, S. R.
    Firat, S. Y.
    Lawton, C. A.
    Kainz, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S496 - S496
  • [5] Total Skin Irradiation using helical tomotherapy: First experience
    Haraldsson, A.
    Engleson, J.
    Engelholm, S.
    Enstrom, P. E.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1204 - S1205
  • [6] Dosimetric Evaluation of Helical Tomotherapy Low-Dose Total Body Irradiation (TBI) for Mini Allogeneic Transplants
    Kelly, D. A.
    Lollar, D.
    Sen, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S683 - S683
  • [7] DOSIMETRIC STUDY AND VERIFICATION OF TOTAL BODY IRRADIATION USING HELICAL TOMOTHERAPY AND ITS COMPARISON TO EXTENDED SSD TECHNIQUE
    Zhuang, Audrey H.
    Liu, An
    Schultheiss, Timothy E.
    Wong, Jeffrey Y. C.
    MEDICAL DOSIMETRY, 2010, 35 (04) : 243 - 249
  • [8] FIRST FRENCH EXPERIENCES OF TOTAL BODY IRRADIATION USING HELICAL TOMOTHERAPY
    Sun, R.
    Chea, M.
    Jenny, C.
    Itti, R.
    Vernant, J. P.
    Mazeron, J. J.
    Cuenca, X.
    HAEMATOLOGICA, 2014, 99 : 432 - 433
  • [9] Investigation of the Dosimetric Accuracy for OSLD In-Vivo Measurements of Total Body Irradiation
    Eldib, A.
    Lin, T.
    Veltchev, I.
    Hossain, M.
    Price, R.
    Ma, C.
    MEDICAL PHYSICS, 2018, 45 (06) : E287 - E288
  • [10] Procedure for total body irradiation (TBI) with Helical Tomotherapy
    Ferrer, C.
    Huertas, C.
    Plaza, R.
    de la Monja, P.
    Minguez, C.
    Garcia, D.
    Aznar, A.
    Saez, M.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1597 - S1598