Detection of anti-dsDNA antibodies by computer-aided automated immunofluorescence analysis

被引:12
|
作者
Lakos, Gabriella [1 ]
Gonzalez, Marlene [1 ]
Flaherty, David [1 ]
Bentow, Chelsea [1 ]
Ibarra, Claudia [2 ]
Stimson, Deborah [2 ]
Nacario, Lori [2 ]
Hiemann, Rico [3 ]
Dervieux, Thierry [2 ]
机构
[1] Inova Diagnost Inc, San Diego, CA 92182 USA
[2] Exagen Diagnost, Vista, CA USA
[3] Brandenburg Univ Technol Cottbus Senftenberg, Fac Nat Sci, Senftenberg, Germany
关键词
dsDNA; Autoantibodies; Indirect immunofluorescence; Systemic lupus erythematosus; Crithidia luciliae; DOUBLE-STRANDED DNA; CRITHIDIA-LUCILIAE; CLASSIFICATION; ASSAYS; FARR;
D O I
10.1016/j.jim.2016.02.019
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Introduction: NOVA View is a computer aided fluorescence microscope that is used for the automated reading and interpretation of indirect immunofluorescent tests in diagnostic immunology. The objective of the present study was to evaluate the performance of the NOVA View system for the measurement of anti-dsDNA antibodies using the Crithidia luciliae indirect immunofluorescence test (CLIFT) technology. Methods: Analytical performance of NOVA View CLIFF was assessed in repeatability (within run) and reproducibility (between runs and instruments) studies. Two hundred-fifty patient samples (N = 200 consecutive samples and N = 50 samples from systemic lupus erythematosus patients) were tested to evaluate the agreement between results generated with NOVA View CLIFF, and those obtained with manual microscopic reading of the same slides. Positivity rate in SLE was assessed on the 50 SLE samples. Results: The NOVA View system showed high level of repeatability and reproducibility within runs, between runs, and between instruments. Agreement of NOVA View software interpretation and digital image reading results with manual microscopic reading results was 96.0%, and the same positivity rate was obtained on SLE samples by NOVA View digital image reading as that of manual microscopic reading (36.0% vs. 38.0%, respectively). Conclusion: Results generated by NOVA View CLIFF were equivalent to those obtained by manual microscopic reading on a large routine sample set. NOVA View demonstrated consistency within and between runs, and between instruments. Automation of CLIFF provides reliability and is a suitable alternative for routine clinical laboratories. (C) 2016 The Authors. Published by Elsevier B.V.
引用
收藏
页码:17 / 22
页数:6
相关论文
共 50 条
  • [31] Computer-aided detection for the automated evaluation of pulmonary embolism
    Li, Yan
    Dai, Yongliang
    Deng, Lei
    Yu, Nan
    Guo, Youmin
    TECHNOLOGY AND HEALTH CARE, 2017, 25 : S135 - S142
  • [32] Production and characterization of human monoclonal anti-idiotype antibodies to anti-dsDNA antibodies
    Zhang, W
    Frank, MB
    Reichlin, M
    LUPUS, 2002, 11 (06) : 362 - 369
  • [33] Automated Computer-aided Diagnosis for Brain Tumor Detection
    Pranav, P.
    Samhita, P.
    13TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2021), 2018,
  • [34] Clinical comparison of five assays for the detection of anti-dsdna antibodies and their correlation with complement consumption
    Ricchiuti, V.
    Obney, J. R.
    Holloway, B.
    Aure, M.
    Mahler, M.
    CLINICAL CHEMISTRY, 2024, 70
  • [35] DETECTION OF ANTI-DSDNA AS A DIAGNOSTIC-TOOL - A PROSPECTIVE-STUDY IN 441 NON-SYSTEMIC LUPUS-ERYTHEMATOSUS PATIENTS WITH ANTI-DSDNA ANTIBODY (ANTI-DSDNA)
    SWAAK, T
    SMEENK, R
    ANNALS OF THE RHEUMATIC DISEASES, 1985, 44 (04) : 245 - 251
  • [36] Detection of Anti-ENA and anti-dsDNA Antibodies Using Line Immunoassay in Systemic Autoimmune Diseases
    Kim, Ji Myung
    Ilm, Chun Hwa
    Sin, Dong Hyuk
    Ihm, Mi Kyung
    Sim, Seung Chul
    KOREAN JOURNAL OF LABORATORY MEDICINE, 2008, 28 (05): : 353 - 361
  • [37] The predictive value of ANA and anti-dsDNA antibodies for flares in SLE
    Bizzaro, N
    Villalta, D
    RHEUMATOLOGY, 2001, 40 (12) : 1422 - 1423
  • [38] Clinical evaluation of a new automated anti-dsDNA fluorescent immunoassay
    Hernando, M
    González, C
    Sánchez, A
    Guevara, P
    Navajo, JA
    Papisch, W
    González-Buitrago, JM
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (10) : 1056 - 1060
  • [39] Crossreactivity of human anti-dsDNA antibodies to phosphorylcholine: Clues to their origin
    Sharma, A
    Isenberg, DA
    Diamond, B
    JOURNAL OF AUTOIMMUNITY, 2001, 16 (04) : 479 - 484
  • [40] Anti-dsDNA antibodies in the classification criteria of systemic lupus erythematosus
    Infantino, Maria
    Nagy, Eszter
    Bizzaro, Nicola
    Fischer, Katarzyna
    Bossuyt, Xavier
    Damoiseaux, Jan
    JOURNAL OF TRANSLATIONAL AUTOIMMUNITY, 2022, 5