Detection of anti-dsDNA antibodies by computer-aided automated immunofluorescence analysis

被引:12
|
作者
Lakos, Gabriella [1 ]
Gonzalez, Marlene [1 ]
Flaherty, David [1 ]
Bentow, Chelsea [1 ]
Ibarra, Claudia [2 ]
Stimson, Deborah [2 ]
Nacario, Lori [2 ]
Hiemann, Rico [3 ]
Dervieux, Thierry [2 ]
机构
[1] Inova Diagnost Inc, San Diego, CA 92182 USA
[2] Exagen Diagnost, Vista, CA USA
[3] Brandenburg Univ Technol Cottbus Senftenberg, Fac Nat Sci, Senftenberg, Germany
关键词
dsDNA; Autoantibodies; Indirect immunofluorescence; Systemic lupus erythematosus; Crithidia luciliae; DOUBLE-STRANDED DNA; CRITHIDIA-LUCILIAE; CLASSIFICATION; ASSAYS; FARR;
D O I
10.1016/j.jim.2016.02.019
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Introduction: NOVA View is a computer aided fluorescence microscope that is used for the automated reading and interpretation of indirect immunofluorescent tests in diagnostic immunology. The objective of the present study was to evaluate the performance of the NOVA View system for the measurement of anti-dsDNA antibodies using the Crithidia luciliae indirect immunofluorescence test (CLIFT) technology. Methods: Analytical performance of NOVA View CLIFF was assessed in repeatability (within run) and reproducibility (between runs and instruments) studies. Two hundred-fifty patient samples (N = 200 consecutive samples and N = 50 samples from systemic lupus erythematosus patients) were tested to evaluate the agreement between results generated with NOVA View CLIFF, and those obtained with manual microscopic reading of the same slides. Positivity rate in SLE was assessed on the 50 SLE samples. Results: The NOVA View system showed high level of repeatability and reproducibility within runs, between runs, and between instruments. Agreement of NOVA View software interpretation and digital image reading results with manual microscopic reading results was 96.0%, and the same positivity rate was obtained on SLE samples by NOVA View digital image reading as that of manual microscopic reading (36.0% vs. 38.0%, respectively). Conclusion: Results generated by NOVA View CLIFF were equivalent to those obtained by manual microscopic reading on a large routine sample set. NOVA View demonstrated consistency within and between runs, and between instruments. Automation of CLIFF provides reliability and is a suitable alternative for routine clinical laboratories. (C) 2016 The Authors. Published by Elsevier B.V.
引用
收藏
页码:17 / 22
页数:6
相关论文
共 50 条
  • [21] ADSORPTION OF ANTI-DSDNA ANTIBODIES BY IMMOBILIZED POLYANIONIC COMPOUNDS
    AOTSUKA, S
    FUNAHASHI, T
    TANI, N
    OKAWATAKATSUJI, M
    KINOSHITA, M
    YOKOHARI, R
    CLINICAL AND EXPERIMENTAL IMMUNOLOGY, 1990, 79 (02): : 215 - 220
  • [22] Lowering anti-dsDNA antibodies - what's new?
    Schiffer, LE
    Hussain, N
    Wang, X
    Huang, W
    Sinha, J
    Ramanujam, M
    Davidson, A
    LUPUS, 2002, 11 (12) : 885 - 894
  • [23] The Therapeutic Strategies for SLE by Targeting Anti-dsDNA Antibodies
    Yaqi Wang
    Shengxiang Xiao
    Yumin Xia
    Huixia Wang
    Clinical Reviews in Allergy & Immunology, 2022, 63 : 152 - 165
  • [24] The Therapeutic Strategies for SLE by Targeting Anti-dsDNA Antibodies
    Wang, Yaqi
    Xiao, Shengxiang
    Xia, Yumin
    Wang, Huixia
    CLINICAL REVIEWS IN ALLERGY & IMMUNOLOGY, 2022, 63 (02) : 152 - 165
  • [25] Clinical significance of ELISA positive and immunofluorescence negative anti-dsDNA antibody
    Kim, Kyeong Hee
    Han, Jin Yeong
    Kim, Jeong Man
    Lee, Sung Won
    Chung, Won Tae
    CLINICA CHIMICA ACTA, 2007, 380 (1-2) : 182 - 185
  • [26] Clinical evaluation of a new automated enzyme immunoassay that uses recombinant plasmid dsDNA for the measurement of anti-dsDNA antibodies.
    González, C
    Hernando, M
    Sánchez, A
    Papisch, W
    Guevara, P
    Navajo, JA
    González-Buitrago, JM
    CLINICAL CHEMISTRY, 2001, 47 (06) : A46 - A47
  • [27] The clinical relevance of anti-dsDNA antibodies determined by the Elia™ dsDNA assay in patients with negative indirect immunofluorescence on the HEp-2 cell
    Robier, Christoph
    Haas, Maximiliane
    Quehenberger, Franz
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (03) : 541 - 546
  • [28] DETECTION OF ANTI-DSDNA AS A DIAGNOSTIC-TOOL
    MCHARDY, KC
    RENNIE, JAN
    REID, TMS
    HORNE, CHW
    ANNALS OF THE RHEUMATIC DISEASES, 1985, 44 (08) : 573 - 574
  • [29] Comparative analysis of three methods in anti-dsDNA antibodies detection: implications for Systemic Lupus Erythematosus diagnosis
    Cuomo, Laura
    Vitillo, Marina
    Della Rocca, Massimiliano
    Trivedi, Pankaj
    SCANDINAVIAN JOURNAL OF IMMUNOLOGY, 2022, 95 (02)
  • [30] Clinical significance of anti-dsDNA antibody isotypes:: IgG/IgM ratio of anti-dsDNA antibodies as a prognostic marker for lupus nephritis
    Förger, F
    Matthias, T
    Oppermann, M
    Becker, H
    Helmke, K
    LUPUS, 2004, 13 (01) : 36 - 44