Reverse total shoulder arthroplasty baseplate stability with locking vs. non-locking peripheral screws

被引:0
|
作者
Martin, Elise J. [1 ,2 ]
Duquin, Thomas R. [2 ]
Ehrensberger, Mark T. [1 ,2 ,3 ]
机构
[1] SUNY Buffalo, Dept Biomed Engn, Buffalo, NY USA
[2] SUNY Buffalo, Dept Orthopaed Surg, Buffalo, NY USA
[3] SUNY Buffalo, Dept Biomed Engn, 3435 Main St,445 Biomed Res Bldg, Buffalo, NY 14214 USA
关键词
Total shoulder arthroplasty; Reverse total shoulder arthroplasty; Glenoid baseplate; Biomechanics; 3D digital image correlation; Peripheral screw; GLENOID COMPONENT FIXATION; KNEE ARTHROPLASTY; BONE INGROWTH; POSITION; STRAIN; MOVEMENT; DESIGN; METAL;
D O I
10.1016/j.clinbiomech.2022.105665
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: There are many options for glenosphere baseplate fixation commercially available, yet there is little biomechanical evidence supporting one type of fixation over another. In this study, we compared the biomechanical fixation of a reverse total shoulder glenoid baseplate secured with locking or non-locking peripheral screws. Methods: Both a non-augmented mini baseplate with full backing support and an augmented baseplate were testing after implantation in solid rigid polyurethane foam. Each baseplate was implanted with a 30 mm central compression screw and four peripheral screws, either locking or non-locking (15 mm anterior/posterior and 30 mm superior/inferior). A 1 Hz cyclic force of 0-750 N was applied at a 60 degrees angle for 5000 cycles. Throughout the test, the displacement of the baseplate was measured using a 3D Digital Image Correlation System. Findings: The amount of migration measured in the both the non-augmented and augment cases shows no significant differences between locking and non-locking cases at the final cycle count (non-augment: 5.66 +/- 2.29 mu m vs. 3.71 +/- 1.23 mu m; p = 0.095, augment: 15.43 +/- 8.49 mu m vs. 12.46 +/- 3.24 mu m; p = 0.314). Additionally, the amount of micromotion measured for both sample types shows the same lack of significant difference (non-augment: 10.79 +/- 5.22 mu m vs. 10.16 +/- 7.61 mu m; p = 0.388, augment: 55.03 +/- 10.13 mu m vs. 54.84 +/- 10.65 mu m; p = 0.968). Interpretation: The presence of locking versus non-locking peripheral screws does not make a significant difference on the overall stability of a glenoid baseplate, in both a no defect case with a non-augmented baseplate and a bone defect case with an augmented baseplate.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] A quantitative analysis of the effect of baseplate and glenosphere position on deltoid lengthening in reverse total shoulder arthroplasty
    Wright, Jonathan
    Potts, Christopher
    Smyth, Mark P.
    Ferrara, Lisa
    Sperling, John W.
    Throckmorton, Thomas W.
    INTERNATIONAL JOURNAL OF SHOULDER SURGERY, 2015, 9 (02): : 33 - 37
  • [32] Outcomes of Reverse Total Shoulder Arthroplasty for Proximal Humeral Fractures: Primary Arthroplasty Versus Secondary Arthroplasty After Failed Proximal Humeral Locking Plate Fixation
    Sebastia-Forcada, Emilio
    Lizaur-Utrilla, Alejandro
    Cebrian-Gomez, Roman
    Miralles-Munoz, Francisco A.
    Lopez-Prats, Fernando A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (08) : E236 - E240
  • [33] A Biomechanical Comparison of Locked Unicortical with Non-Locking Bicortical Screws for Metacarpal Shaft Fractures in a Randomized Human Cadaveric Model
    Minogue, Febe
    Reilly, Keryn
    Walker, Cameron
    Malak, Sharif
    Mayo, Andrew
    Heiss-Dunlop, Wolfgang
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2021, 26 (01): : 10 - 16
  • [34] A two-year comparison of reverse total shoulder arthroplasty mini-humeral tray and augmented mini-glenoid baseplate implants vs. standard implants
    Krupp, Ryan
    O'Grady, Christopher
    Werner, Brian
    Wiater, J. Michael
    Nyland, John
    Duquin, Thomas
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2024, 144 (05) : 1925 - 1935
  • [35] A two-year comparison of reverse total shoulder arthroplasty mini-humeral tray and augmented mini-glenoid baseplate implants vs. standard implants
    Ryan Krupp
    Christopher O’Grady
    Brian Werner
    J. Michael Wiater
    John Nyland
    Thomas Duquin
    Archives of Orthopaedic and Trauma Surgery, 2024, 144 : 1925 - 1935
  • [36] Cement augmentation of glenoid baseplate screws does not improve primary stability in reversed shoulder arthroplasty: A cadaveric study
    Bockmann, Benjamin
    Baeumlein, Martin
    Koch, Rainer
    Figiel, Jens
    Lechler, Philipp
    Aigner, Rene
    Bliemel, Christopher
    Oberkircher, Ludwig
    Ruchholtz, Steffen
    Frink, Michael
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2019, 105 (02) : 219 - 223
  • [37] Clinical and Radiologic Outcomes of Small Glenoid Baseplate in Reverse Total Shoulder Arthroplasty: A Prospective Multicenter Study
    Kim, Min-Su
    Rhee, Yong Girl
    Oh, Joo Han
    Yoo, Jae Chul
    Noh, Kyu-Cheol
    Shin, Sang-Jin
    CLINICS IN ORTHOPEDIC SURGERY, 2022, 14 (01) : 119 - 127
  • [38] Are proximal screws necessary for osteosynthesis of stable-stem periprosthetic femoral fractures fixed with non-locking plate and cable?
    Yang, Tzu-Cheng
    Tzeng, Yun-Hsuan
    Wang, Chien-Shun
    Lin, Chun-Cheng
    Chang, Ming-Chau
    Chiang, Chao-Ching
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2019, 50 (10): : 1739 - 1744
  • [39] Outcomes and revision rates of primary vs. secondary reverse total shoulder arthroplasty for proximal humeral fractures
    Katthagen, J. Christoph
    Hesse, Elina
    Lill, Helmut
    Schliemann, Benedikt
    Ellwein, Alexander
    Raschke, Michael J.
    Imrecke, Julia
    OBERE EXTREMITAET-SCHULTER-ELLENBOGEN-HAND-UPPER EXTREMITY-SHOULDER ELBOW HAND, 2020, 15 (01): : 35 - 41
  • [40] Scapular and humeral elevation coordination patterns used before vs. after Reverse Total Shoulder Arthroplasty
    Zaferiou, Antonia M.
    Knowlton, Christopher B.
    Jang, Suk-Hwan
    Saltzman, Bryan M.
    Verma, Nikhil N.
    Forsythe, Brian
    Nicholson, Gregory P.
    Romeo, Anthony A.
    JOURNAL OF BIOMECHANICS, 2021, 125