Reverse total shoulder arthroplasty baseplate stability with locking vs. non-locking peripheral screws

被引:0
|
作者
Martin, Elise J. [1 ,2 ]
Duquin, Thomas R. [2 ]
Ehrensberger, Mark T. [1 ,2 ,3 ]
机构
[1] SUNY Buffalo, Dept Biomed Engn, Buffalo, NY USA
[2] SUNY Buffalo, Dept Orthopaed Surg, Buffalo, NY USA
[3] SUNY Buffalo, Dept Biomed Engn, 3435 Main St,445 Biomed Res Bldg, Buffalo, NY 14214 USA
关键词
Total shoulder arthroplasty; Reverse total shoulder arthroplasty; Glenoid baseplate; Biomechanics; 3D digital image correlation; Peripheral screw; GLENOID COMPONENT FIXATION; KNEE ARTHROPLASTY; BONE INGROWTH; POSITION; STRAIN; MOVEMENT; DESIGN; METAL;
D O I
10.1016/j.clinbiomech.2022.105665
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: There are many options for glenosphere baseplate fixation commercially available, yet there is little biomechanical evidence supporting one type of fixation over another. In this study, we compared the biomechanical fixation of a reverse total shoulder glenoid baseplate secured with locking or non-locking peripheral screws. Methods: Both a non-augmented mini baseplate with full backing support and an augmented baseplate were testing after implantation in solid rigid polyurethane foam. Each baseplate was implanted with a 30 mm central compression screw and four peripheral screws, either locking or non-locking (15 mm anterior/posterior and 30 mm superior/inferior). A 1 Hz cyclic force of 0-750 N was applied at a 60 degrees angle for 5000 cycles. Throughout the test, the displacement of the baseplate was measured using a 3D Digital Image Correlation System. Findings: The amount of migration measured in the both the non-augmented and augment cases shows no significant differences between locking and non-locking cases at the final cycle count (non-augment: 5.66 +/- 2.29 mu m vs. 3.71 +/- 1.23 mu m; p = 0.095, augment: 15.43 +/- 8.49 mu m vs. 12.46 +/- 3.24 mu m; p = 0.314). Additionally, the amount of micromotion measured for both sample types shows the same lack of significant difference (non-augment: 10.79 +/- 5.22 mu m vs. 10.16 +/- 7.61 mu m; p = 0.388, augment: 55.03 +/- 10.13 mu m vs. 54.84 +/- 10.65 mu m; p = 0.968). Interpretation: The presence of locking versus non-locking peripheral screws does not make a significant difference on the overall stability of a glenoid baseplate, in both a no defect case with a non-augmented baseplate and a bone defect case with an augmented baseplate.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Superior Baseplate Inclination Is Associated With Instability After Reverse Total Shoulder Arthroplasty
    McFarland, Edward G.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2018, 476 (08) : 1630 - 1632
  • [22] Risk of penetration of the baseplate peg in reverse total shoulder arthroplasty for an Asian population
    Eiji Tashiro
    Naohide Takeuchi
    Naoya Kozono
    Akira Nabeshima
    Ei Teshima
    Yasuharu Nakashima
    International Orthopaedics, 2022, 46 : 1063 - 1071
  • [23] How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability
    Diaz, Miguel A.
    Hutchinson, Adam J.
    Ricchetti, Eric T.
    Hsu, Jason E.
    Garrigues, Grant E.
    Gutierrez, Sergio
    Frankle, Mark A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2024, 33 (02) : 389 - 398
  • [24] Inlay vs. onlay humeral components in reverse total shoulder arthroplasty: a biorobotic shoulder simulator study
    Zitnay, Jared L.
    Tashjian, Robert Z.
    Walch, Gilles
    Chalmers, Peter N.
    Joyce, Christopher D.
    Henninger, Heath B.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2024, 33 (06) : 1377 - 1386
  • [25] Aseptic glenoid baseplate loosening after reverse total shoulder arthroplasty with a single prosthesis
    Schell, Lauren E.
    Roche, Christopher P.
    Eichinger, Josef K.
    Flurin, Pierre-Henri
    Wright, Thomas W.
    Zuckerman, Joseph D.
    Friedman, Richard J.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2023, 32 (08) : 1584 - 1593
  • [26] Relationship Between Insertion Torque and Compression Strength in the Reverse Total Shoulder Arthroplasty Baseplate
    Diaz, Miguel A.
    Garrigues, Grant E.
    Ricchetti, Eric T.
    Gutierrez, Sergio
    Frankle, Mark A.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2020, 38 (04) : 871 - 879
  • [27] Reverse Shoulder Arthroplasty Baseplate Stability Is Affected by Bone Density and the Type and Amount of Augmentation
    Ritter, Daniel
    Raiss, Patric
    Denard, Patrick J.
    Werner, Brian C.
    Kistler, Manuel
    Lesnicar, Celina
    van der Merwe, Micheal
    Mueller, Peter E.
    Woiczinski, Matthias
    Wijdicks, Coen A.
    Bachmaier, Samuel
    BIOENGINEERING-BASEL, 2025, 12 (01):
  • [28] Reverse total shoulder glenoid baseplate stability with superior glenoid bone loss
    Martin, Elise J.
    Duquin, Thomas R.
    Ehrensberger, Mark T.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2017, 26 (10) : 1748 - 1755
  • [29] Evaluation of a new baseplate in reverse total shoulder arthroplasty - comparison of biomechanical testing of stability with roentgenological follow up criteria
    Irlenbusch, U.
    Kohut, G.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2015, 101 (02) : 185 - 190
  • [30] Functional Outcomes Of Hemiarthroplasty Vs. Reverse Polarity Total Shoulder Arthroplasty For Fracture Of The Proximal Humerus
    O'Sullivan, H.
    Cassidy, J. T.
    Molony, D.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2018, 187 : S109 - S109