Toward endobronchial Ir-192 high-dose-rate brachytherapy therapeutic optimization

被引:8
|
作者
Gay, H. A. [1 ]
Allison, R. R.
Downie, G. H.
Mota, H. C.
Austerlitz, C.
Jenkins, T.
Sibata, C. H.
机构
[1] E Carolina Univ, Brody Sch Med, Dept Radiat Oncol, Greenville, NC 27858 USA
[2] E Carolina Univ, Brody Sch Med, Dept Internal Med, Pulm & Crit Care Med Sect, Greenville, NC 27858 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 11期
关键词
D O I
10.1088/0031-9155/52/11/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A number of patients with lung cancer receive either palliative or curative high-dose-rate (HDR) endobronchial brachytherapy. Up to a third of patients treated with endobronchial HDR die from hemoptysis. Rather than accept hemoptysis as an expected potential consequence of HDR, we have calculated the radial dose distribution for an Ir-192 HDR source, rigorously examined the dose and prescription points recommended by the American Brachytherapy Society (ABS), and performed a radiobiological-based analysis. The radial dose rate of a commercially available Ir-192 source was calculated with a Monte Carlo simulation. Based on the linear quadratic model, the estimated palliative, curative and blood vessel rupture radii from the center of an Ir-192 source were obtained for the ABS recommendations and a series of customized HDR prescriptions. The estimated radius at risk for blood vessel perforation for the ABS recommendations ranges from 7 to 9 mm. An optimized prescription may in some situations reduce this radius to 4 mm. The estimated blood perforation radius is generally smaller than the palliative radius. Optimized and individualized endobronchial HDR prescriptions are currently feasible based on our current understanding of tumor and normal tissue radiobiology. Individualized prescriptions could minimize complications such as fatal hemoptysis without sacrificing efficacy. Fiducial stents, HDR catheter centering or spacers and the use of CT imaging to better assess the relationship between the catheter and blood vessels promise to be useful strategies for increasing the therapeutic index of this treatment modality. Prospective trials employing treatment optimization algorithms are needed.
引用
收藏
页码:2987 / 2999
页数:13
相关论文
共 50 条
  • [1] Impact of different Ir-192 source models on dose calculations in high-dose-rate brachytherapy
    Otani, Yuki
    Shimamoto, Shigetoshi
    Sumida, Iori
    Takahashi, Yutaka
    Tani, Shoji
    Oshima, Tetsuya
    Onosaka, Satoshi
    Isohashi, Fumiaki
    Tamari, Keisuke
    Ogawa, Kazuhiko
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 7 : 23 - 26
  • [2] CALIBRATION OF IR-192 HIGH-DOSE-RATE AFTERLOADING SYSTEMS
    GOETSCH, SJ
    ATTIX, FH
    PEARSON, DW
    THOMADSEN, BR
    MEDICAL PHYSICS, 1991, 18 (03) : 462 - 467
  • [3] Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source
    Sharma, SD
    Bianchi, C
    Conte, L
    Novario, R
    Bhatt, BC
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17): : 4065 - 4072
  • [4] Imaging of Ir-192 source using a high energy gamma camera for high-dose-rate brachytherapy
    Yamamoto, Seiichi
    Yogo, Katsunori
    Noguchi, Yumiko
    Nakaya, Takayoshi
    Okudaira, Kuniyasu
    RADIATION MEASUREMENTS, 2019, 126
  • [5] Dose volume assessment of high dose rate Ir-192 endobronchial implants
    Saw, CB
    Korb, LJ
    Pawlicki, T
    Wu, A
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (04): : 917 - 922
  • [6] INTERSTITIAL HIGH-DOSE-RATE BRACHYTHERAPY WITH IR-192 IN PATIENTS WITH ORAL SQUAMOUS-CELL CARCINOMA
    FRIEDRICH, RE
    KRULL, A
    HELLNER, D
    SCHWARZ, R
    HEYER, D
    PLAMBECK, K
    SCHMELZLE, R
    JOURNAL OF CRANIO-MAXILLO-FACIAL SURGERY, 1995, 23 (04): : 238 - 242
  • [7] Trials of transmission imaging using clinically used Ir-192 source for high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Nakanishi, K.
    Noguchi, Y.
    Okudaira, K.
    Kataoka, J.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (06)
  • [8] Transit dose of an Ir-192 high dose rate brachytherapy stepping source
    Wong, TPY
    Fernando, W
    Johnston, PN
    Bubb, IF
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 323 - 331
  • [9] THE HALF-LIFE OF HIGH-DOSE-RATE IR-192 SOURCES
    PODGORSAK, MB
    DEWERD, LA
    PALIWAL, BR
    MEDICAL PHYSICS, 1993, 20 (04) : 1257 - 1259
  • [10] Interstitial brachytherapy for lip carcinomas: Comparison between Ir-192 low-dose-rate and high-dose-rate treatment
    Cuenin, M.
    Salleron, J.
    Peiffert, D.
    Meknaci, E.
    Gallet, P.
    Abushama, Y.
    Py, J. -F
    Renard, S.
    CANCER RADIOTHERAPIE, 2024, 28 (02): : 145 - 151