Optimizing 4D cone-beam CT acquisition protocol for external beam radiotherapy

被引:86
|
作者
Li, Tianfang [1 ]
Xing, Lei [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
关键词
image-guided radiotherapy; four-dimensional cone-beam CT; optimization;
D O I
10.1016/j.ijrobp.2006.10.024
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Four-dimensional cone-beam computed tomography (4D-CBCT) imaging is sensitive to parameters such as gantry rotation speed, number of gantry rotations, X-ray pulse rate, and tube current, as well as a patient's breathing pattern. The aim of this study is to optimize the image acquisition on a patient-specific basis while minimizing the scan time and the radiation dose. Methods and Materials: More than 60 sets of 4D-CBCT images, each with a temporal resolution of 10 phases, were acquired using multiple-gantry rotation and slow-gantry rotation techniques. The image quality was quantified with a relative root mean-square error (RE) and correlated with various acquisition settings; specitically, varying gantry rotation speed, varying both the rotation speed and the number of rotations, and varying both the rotation speed and tube current to keep the radiation exposure constant. These experiments were repeated for three different respiratory periods. Results: With similar radiation dose, 4D-CBCT images acquired with low current and low rotation speed have better quality over images obtained with high current and high rotation speed. In general, a one-rotation low-speed scan is superior to a two-rotation double-speed scan, even though they provide the same number of projections. Furthermore, it is found that the image quality behaves monotonically with the relative speed as defined by the gantry rotation speed and the patient respiratory period. Conclusions: The RE curves established in this work can be used to predict the 4D-CBCT image quality before a scan. This allows the acquisition protocol to be optimized individually to balance the desired quality with the associated scanning time and patient radiation dose. (c) 2007 Elsevier Inc.
引用
收藏
页码:1211 / 1219
页数:9
相关论文
共 50 条
  • [31] A hybrid reconstruction algorithm for fast and accurate 4D cone-beam CT imaginga)
    Yan, Hao
    Zhen, Xin
    Folkerts, Michael
    Li, Yongbao
    Pan, Tinsu
    Cervino, Laura
    Jiang, Steve B.
    Jia, Xun
    MEDICAL PHYSICS, 2014, 41 (07)
  • [32] Simultaneous Motion Estimation and Image Reconstruction (SMEIR) for 4D Cone-Beam CT
    Wang, J.
    Gu, X.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [33] Directional sinogram interpolation for motion weighted 4D cone-beam CT reconstruction
    Zhang, Hua
    Kruis, Matthijs
    Sonke, Jan-Jakob
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (06): : 2254 - 2275
  • [34] Simultaneous Motion Estimation and Image Reconstruction (SMEIR) for 4D Cone-beam CT
    Wang, Jing
    Gu, Xuejun
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [35] Evaluation of 4D CT to 4D Cone-Beam CT Deformable Image Registration for Lung Cancer Adaptive Radiation Therapy
    Balik, S.
    Hugo, G.
    Weiss, E.
    Jan, N.
    Roman, N.
    Sleeman, W.
    Fatyga, M.
    Christensen, G.
    Murphy, M.
    Lu, J.
    Keall, P.
    Williamson, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3875 - 3875
  • [36] Respiratory signal extraction for 4D CT imaging of the thorax from cone-beam CT projections
    Rit, S
    Sarrut, D
    Ginestet, C
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S533 - S534
  • [37] Respiratory signal extraction for 4D CT imaging of the thorax from cone-beam CT projections
    Rit, S
    Sarrut, D
    Ginestet, C
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 1, 2005, 3749 : 556 - 563
  • [38] 4D Cone-beam CT Deformable Registration using Unsupervised Spatial Transformation Network
    Wang, Tonghe
    Lei, Yang
    Tian, Zhen
    Giles, Matt
    Bradley, Jeffrey D.
    Curran, Walter J.
    Liu, Tian
    Yang, Xiaofeng
    MEDICAL IMAGING 2021: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2021, 11600
  • [39] Development of the 4D Cone-Beam CT Modelling Software to Investigate Effects of Imaging Parameters
    Kim, D.
    Kim, T.
    Kang, S.
    Suh, T.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [40] A Method for 4D Cone-Beam CT in Under 20 Seconds with New Generation Linacs
    Blake, S.
    Dillon, O.
    O'Brien, R.
    MEDICAL PHYSICS, 2021, 48 (06)