Comparison of the morphological and biomechanical characteristics of keratoconus, forme fruste keratoconus, and normal corneas

被引:20
|
作者
Guo, Li-Li [1 ]
Tian, Lei [1 ,2 ,3 ]
Cao, Kai [1 ]
Li, Yu-Xin [1 ]
Li, Na [1 ]
Yang, Wei-Qiang [1 ]
Jie, Ying [1 ]
机构
[1] Capital Med Univ, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Hosp, Beijing Inst Ophthalmol,Beijing Tongren Eye Ctr, Beijing, Peoples R China
[2] Beihang Univ, Beijing Adv Innovat Ctr Big Data Based Precis Med, Beijing, Peoples R China
[3] Capital Med Univ, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
keratoconus; forme fruste keratoconus; corneal morphology; corneal biomechanics; Corvis ST;
D O I
10.1080/08820538.2021.1896752
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To explore the feasibility of corneal morphological and biomechanical parameters for keratoconus and forme fruste keratoconus diagnosis. Methods: This case-control study included a total of 517 eyes from 408 keratoconus patients (KC group), 83 eyes from 83 forme fruste keratoconus patients (FFKC group), and 158 eyes from 158 patients with normal corneas (NL group). All subjects underwent routine ophthalmologic examinations. Pentacam and Corneal Visualization Scheimpflug Technology (Corvis ST) were used to obtain corneal morphological and biomechanical parameters. Differences between groups were compared using receiver operating characteristic (ROC) curve analysis. Results: ROC analysis showed that all corneal morphological parameters and most biomechanical parameters distinguished KC from NL, with an area under the curve (AUC) greater than 0.80, of which Belin-Ambrosio enhanced ectasia total deviation index (BAD-D) and tomographic and biomechanical index (TBI) were most efficient. The AUC for distinguishing KC from NL of the BAD-D was 0.989 and the TBI was 0.993, which were not statistically significant (DeLong et al., P= .232). The BAD-D cut-off point of 1.595 provided 95.9% sensitivity for distinguishing KC from NL with 100% specificity. The TBI cut-off point of 0.515 provided 96.7% sensitivity for distinguishing KC from NL with 100% specificity. The ability of other parameters to distinguish KC from NL was lower than that of BAD and TBI. Except for central astigmatism from the anterior corneal surface (AstigF), the AUC that distinguished FFKC from NL was 0.862. The AstigF cut-off point of 4.65 provided 73.5% sensitivity for distinguishing FFKC from NL with 99.3% specificity. Other parameters distinguished FFKC from NL with low efficiency. Among them, the AUC for distinguishing FFKC from NL of the TBI was 0.722, whose cut-off point of 0.273 provided 55.4% sensitivity for distinguishing KC from NL with 79.7% specificity. Conclusion: BAD-D and TBI have the highest efficiency, sensitivity, and specificity for distinguishing KC from NL. Except for AstigF, other corneal morphological and biomechanical parameters have a relatively low ability to distinguish FFKC from NL.
引用
收藏
页码:671 / 678
页数:8
相关论文
共 50 条
  • [21] Screening of Forme Fruste Keratoconus with the Ocular Response Analyzer
    Schweitzer, Cedric
    Roberts, Cynthia J.
    Mahmoud, Ashraf M.
    Colin, Joseph
    Maurice-Tison, Sylvie
    Kerautret, Julien
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (05) : 2403 - 2410
  • [22] Corneal and Total Aberrations in Patients With Fruste Forme Keratoconus
    Schlegel, Z.
    Gatinel, D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [23] Bilateral severe keratoconus after laser in situ keratomileusis in a patient with forme fruste keratoconus
    Lafond, G
    Bazin, R
    Lajoie, C
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (07): : 1115 - 1118
  • [24] Ocular, corneal, and internal aberrations in eyes with keratoconus, forme fruste keratoconus, and healthy eyes
    Naderan, Mohammad
    Jahanrad, Ali
    Farjadnia, Mahgol
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (04) : 1565 - 1573
  • [25] Transiency of Fleischer's Rings in Forme-Fruste Keratoconus
    Downie, Laura E.
    OPHTHALMOLOGY, 2013, 120 (05) : 1101 - +
  • [26] An exploratory analysis of forme fruste keratoconus sensitivity diagnostic parameters
    Hui Zhang
    Xue Zhang
    Lin Hua
    Lin Li
    Lei Tian
    Xinxin Zhang
    Haixia Zhang
    International Ophthalmology, 2022, 42 : 2473 - 2481
  • [27] COMPARISON OF MECHANICAL-PROPERTIES OF KERATOCONUS AND NORMAL CORNEAS
    NASH, IS
    GREENE, PR
    FOSTER, CS
    EXPERIMENTAL EYE RESEARCH, 1982, 35 (05) : 413 - 424
  • [28] Diagnosis of forme fruste keratoconus with scheimpflug photography in Ghanaian patients
    Lartey, Seth
    Appiagyei, Emmanuel
    BMC OPHTHALMOLOGY, 2024, 24 (01)
  • [29] An exploratory analysis of forme fruste keratoconus sensitivity diagnostic parameters
    Zhang, Hui
    Zhang, Xue
    Hua, Lin
    Li, Lin
    Tian, Lei
    Zhang, Xinxin
    Zhang, Haixia
    INTERNATIONAL OPHTHALMOLOGY, 2022, 42 (08) : 2473 - 2481
  • [30] Iatrogenic keratectasia after LASIK in a case of forme fruste keratoconus
    Seiler, T
    Quurke, AW
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (07): : 1007 - 1009