Kalman-Based Calibration Algorithm for AgaMatrix Continuous Glucose Monitoring System

被引:6
|
作者
Singh, Abhinoy Kumar [1 ,2 ]
Rebec, Mihailo V. [3 ]
Haidar, Ahmad [1 ]
机构
[1] McGill Univ, Dept Biomed Engn, Montreal, PQ H3A 2B4, Canada
[2] IIT Indore, Dept Elect Engn, Indore 453552, India
[3] WaveForm Technol Inc, Wilsonville, OR 97070 USA
关键词
Glucose; Calibration; Kalman filters; Blood; Noise measurement; Current measurement; Estimation; Continuous glucose monitoring; glucose sensor; Kalman filter; parameter estimation; INTEGRAL EQUATIONS; DECONVOLUTION; ACCURACY; SENSORS; STATE;
D O I
10.1109/TCST.2020.3003450
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A continuous glucose monitoring system is composed of a glucose sensor and an estimation (calibration) algorithm. The glucose sensor is composed of an electrode that is inserted under the skin and generates a noisy electrical current in response to interstitial glucose levels. The relationship between the electrical current and the interstitial glucose levels varies between individuals and within the same individual (with sensor wear time). The estimation algorithm infers the glucose levels in the blood from the noisy electrical current signal and uses intermittent capillary glucose measurements to account for intra-and inter-individual variability. In this article, we propose a novel real-time Kalman filter-based estimation algorithm that is composed of three steps: 1) noise filtering step; 2) compartment matching step using sequential Kalman filter; and 3) parameters estimation step using the conventional Kalman filter. The initial estimate of the parameters and covariance matrix are extracted offline using a nonlinear cubature Kalman filter. Our algorithm is compared with four alternative algorithms using 20 sensor data sets. Each data set was generated over a seven-day sensor wear period, during which patients were tested on three in-clinic days (days 1, 4, and 7). The comparison is based on the mean absolute relative difference (MARD) between frequent reference glucose measurements and the sensor glucose levels. MARD for days 1, 4, and 7 were 10.3%, 10.7%, and 11.9% for our algorithm and 12.46%, 15.75%, and 14.17% for the alternative algorithm, respectively. Our algorithm is currently being integrated into a commercial continuous glucose monitoring system (AgaMatrix, Inc., Salem, NH, USA).
引用
收藏
页码:1257 / 1267
页数:11
相关论文
共 50 条
  • [31] Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring
    Goodarzi, Mohammad
    Sharma, Sandeep
    Ramon, Herman
    Saeys, Wouter
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2015, 67 : 147 - 158
  • [32] Kalman-Based Real-Time Functional Decomposition for the Spectral Calibration in Swept Source Optical Coherence Tomography
    Zavareh, Amir Tofighi
    Hoyos, Sebastian
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2020, 14 (02) : 257 - 273
  • [33] CONTINUOUS BLOOD GLUCOSE MONITORING (CGM) SYSTEM BASED ON INTRAVENOUS MICRODIALYSIS AND GLUCOSE SENSING
    Greiner, H.
    Schaller-Ammann, R.
    Huber, A.
    O'Connell, M.
    Krejci, J.
    Porro, G.
    Korsatko, S.
    Gurban, J.
    Deller, S.
    Brunner, M.
    Berghofer, A.
    Priedl, J.
    Zijlstra, E.
    Heise, T.
    Pieber, T. R.
    Schaupp, L.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2013, 58
  • [34] THE NEW MODEL OF THE GLUCOSE DISTURBANCES IN PATIENTS ON HAEMODIALYSIS BASED ON THE CONTINUOUS GLUCOSE MONITORING SYSTEM
    Klimm, Wojciech
    Pietrzak, Bozena
    Smoszna, Jerzy
    Niemczyk, Stanislaw
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2017, 32
  • [35] Performance of a Microdialysis-Based Continuous Glucose Monitoring (CGM) System
    Zijlstra, Eric
    Heise, Tim
    Kuennecke, Wolfgang
    DIABETES, 2012, 61 : A2 - A2
  • [36] Electromagnetic-based Implantable Sensor for Continuous Glucose Monitoring System
    Malik, J.
    Kim, S.
    Jeong, J. Y.
    Bien, F.
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2023, 197
  • [37] Performance of a microdialysis-based continuous glucose monitoring (CGM) system
    Zijlstra, E.
    Heise, T.
    Kuennecke, W.
    DIABETOLOGIA, 2012, 55 : S429 - S429
  • [38] IoT-based continuous glucose monitoring system: A feasibility study
    Tuan Nguyen Gia
    Ali, Mai
    Ben Dhaou, Imed
    Rahmani, Amir M.
    Westerlund, Tomi
    Liljeberg, Pasi
    Tenhunen, Hannu
    8TH INTERNATIONAL CONFERENCE ON AMBIENT SYSTEMS, NETWORKS AND TECHNOLOGIES (ANT-2017) AND THE 7TH INTERNATIONAL CONFERENCE ON SUSTAINABLE ENERGY INFORMATION TECHNOLOGY (SEIT 2017), 2017, 109 : 327 - 334
  • [39] Non-invasive impedance based continuous glucose monitoring system
    Talary, M. S.
    Dewarrat, F.
    Huber, D.
    Falco-Jonasson, L.
    Caduff, A.
    13TH INTERNATIONAL CONFERENCE ON ELECTRICAL BIOIMPEDANCE AND THE 8TH CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY 2007, 2007, 17 : 636 - +
  • [40] Human Feasibility of a Semiconductor-Based Continuous Glucose Monitoring System
    Mujeeb-U-Rahman, Muhammad
    Nazari, Meisam Honarvar
    Sencan, Mehmet
    Nadir, Abdullah
    Heithaus, John H.
    Marcus, Alan O.
    DIABETES, 2023, 72