Kalman-Based Calibration Algorithm for AgaMatrix Continuous Glucose Monitoring System

被引:6
|
作者
Singh, Abhinoy Kumar [1 ,2 ]
Rebec, Mihailo V. [3 ]
Haidar, Ahmad [1 ]
机构
[1] McGill Univ, Dept Biomed Engn, Montreal, PQ H3A 2B4, Canada
[2] IIT Indore, Dept Elect Engn, Indore 453552, India
[3] WaveForm Technol Inc, Wilsonville, OR 97070 USA
关键词
Glucose; Calibration; Kalman filters; Blood; Noise measurement; Current measurement; Estimation; Continuous glucose monitoring; glucose sensor; Kalman filter; parameter estimation; INTEGRAL EQUATIONS; DECONVOLUTION; ACCURACY; SENSORS; STATE;
D O I
10.1109/TCST.2020.3003450
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A continuous glucose monitoring system is composed of a glucose sensor and an estimation (calibration) algorithm. The glucose sensor is composed of an electrode that is inserted under the skin and generates a noisy electrical current in response to interstitial glucose levels. The relationship between the electrical current and the interstitial glucose levels varies between individuals and within the same individual (with sensor wear time). The estimation algorithm infers the glucose levels in the blood from the noisy electrical current signal and uses intermittent capillary glucose measurements to account for intra-and inter-individual variability. In this article, we propose a novel real-time Kalman filter-based estimation algorithm that is composed of three steps: 1) noise filtering step; 2) compartment matching step using sequential Kalman filter; and 3) parameters estimation step using the conventional Kalman filter. The initial estimate of the parameters and covariance matrix are extracted offline using a nonlinear cubature Kalman filter. Our algorithm is compared with four alternative algorithms using 20 sensor data sets. Each data set was generated over a seven-day sensor wear period, during which patients were tested on three in-clinic days (days 1, 4, and 7). The comparison is based on the mean absolute relative difference (MARD) between frequent reference glucose measurements and the sensor glucose levels. MARD for days 1, 4, and 7 were 10.3%, 10.7%, and 11.9% for our algorithm and 12.46%, 15.75%, and 14.17% for the alternative algorithm, respectively. Our algorithm is currently being integrated into a commercial continuous glucose monitoring system (AgaMatrix, Inc., Salem, NH, USA).
引用
收藏
页码:1257 / 1267
页数:11
相关论文
共 50 条
  • [21] Calibration algorithms for continuous glucose monitoring systems based on interstitial fluid sensing
    Sun, Tianyi
    Liu, Jentsai
    Chen, Ching Jung
    BIOSENSORS & BIOELECTRONICS, 2024, 260
  • [22] A dual-rate Kalman filter for continuous glucose monitoring
    Kuure-Kinsey, Matthew
    Palerm, Cesar C.
    Bequette, B. Wayne
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 3305 - +
  • [23] Continuous glucose monitoring system: dawn period calibration does not change accuracy of the method
    Augusto, Gustavo A.
    Sousa, Andre G. P.
    Perazo, Marcela N. A.
    Correa-Giannella, Maria L. C.
    Nery, Marcia
    de Melo, Karla F. S.
    ARQUIVOS BRASILEIROS DE ENDOCRINOLOGIA E METABOLOGIA, 2009, 53 (04) : 425 - 428
  • [24] Continuous Glucose Monitoring System Based on Percutaneous Microneedle Array
    Chien, Ming-Nan
    Chen, Yu-Jen
    Bai, Chin-Han
    Huang, Jung-Tung
    MICROMACHINES, 2022, 13 (03)
  • [25] A novel camera calibration algorithm based on Kalman filter
    Stringa, E
    Regazzoni, CS
    15TH INTERNATIONAL CONFERENCE ON PATTERN RECOGNITION, VOL 1, PROCEEDINGS: COMPUTER VISION AND IMAGE ANALYSIS, 2000, : 872 - 875
  • [26] Microdialysis system for continuous glucose monitoring
    Steinkuhl, R
    Sundermeier, C
    Hinkers, H
    Dumschat, C
    Cammann, K
    Knoll, M
    SENSORS AND ACTUATORS B-CHEMICAL, 1996, 33 (1-3) : 19 - 24
  • [27] Eversense Continuous Glucose Monitoring System
    Crowner, Jamila
    Gabenski, Brandon
    Okere, Arinzechukwu N.
    Honeywell, Marlon S.
    US PHARMACIST, 2020, 45 (09) : 30 - 32
  • [28] MiniMed continuous glucose monitoring system
    不详
    BIOSENSORS & BIOELECTRONICS, 2000, 14 (10-11): : AR1 - AR1
  • [29] Enhanced Accuracy of Continuous Glucose Monitoring by Online Extended Kalman Filtering
    Facchinetti, Andrea
    Sparacino, Giovanni
    Cobelli, Claudio
    DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 (05) : 353 - 363
  • [30] COMPARISON OF DIFFERENT CALIBRATION STRATEGIES FOR CONTINUOUS GLUCOSE MONITORING SENSORS
    Vettoretti, M.
    Facchinetti, A.
    Sparacino, G.
    Del Favero, S.
    Cobelli, C.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 : A16 - A16