Spinal navigation for minimally invasive thoracic and lumbosacral spine fixation: implications for radiation exposure, operative time, and accuracy of pedicle screw placement

被引:67
|
作者
Tajsic, T. [1 ]
Patel, K. [1 ]
Farmer, R. [2 ]
Mannion, R. J. [1 ]
Trivedi, R. A. [1 ]
机构
[1] Cambridge Univ Hosp NHS Fdn Trust, Div Neurosurg, Hills Rd,Box 167, Cambridge CB2 0QQ, England
[2] Cambridge Univ Hosp NHS Fdn Trust, Dept Radiol, Cambridge, England
关键词
Minimally invasive (MIS) spinal fixation; 2D fluoroscopy; 3D fluoroscopy; Spinal navigation; C-Arm; O-Arm; IMAGE GUIDANCE; LUMBAR SPINE; ASSISTED FLUOROSCOPY; SURGERY; SYSTEM; INSERTION; OUTCOMES; FUSION;
D O I
10.1007/s00586-018-5587-z
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Navigation is emerging as a useful adjunct in percutaneous, minimally invasive spinal surgery (MIS). The aim of this study was to compare C-Arm navigated, O-Arm navigated and conventional 2D-fluoroscopy assisted MIS thoracic and lumbosacral spine fixation techniques in terms of operating time, radiation exposure and accuracy of pedicle screw (PS) placement. Retrospective observational study of 152 consecutive adults who underwent MIS fixations for spinal instability: 96 2D-fluoroscopy assisted, 39 3D-C-Arm navigated and 27 using O-Arm navigated. O-Arm navigation significantly reduced PS misplacement (1.23%, p) compared to 3D-C-Arm navigation (7.29%, p = 0.0082) and 2D-fluoro guided placement (5.16%, p = 0379). 3D-C-Arm navigation was associated with lower procedural radiation exposure of the patient (0.4 mSv) than O-Arm navigation (3.24 mSv) or 2D-fluoro guidance (1.5 mSv). Operative time was comparable between three modalities. O-Arm navigation provides greater accuracy of percutaneous instrumentation placement with an acceptable procedural radiation dose delivered to the patients and comparable operative times. These slides can be retrieved under Electronic Supplementary material. [GRAPHICS] .
引用
收藏
页码:1918 / 1924
页数:7
相关论文
共 50 条
  • [41] Pedicle screw accuracy in clinical utilization of minimally invasive navigated robot-assisted spine surgery
    Vardiman, Arnold B.
    Wallace, David J.
    Crawford, Neil R.
    Riggleman, Jessica R.
    Ahrendtsen, Leigh A.
    Ledonio, Charles
    JOURNAL OF ROBOTIC SURGERY, 2020, 14 (03) : 409 - 413
  • [42] Pedicle screw accuracy in clinical utilization of minimally invasive navigated robot-assisted spine surgery
    Arnold B. Vardiman
    David J. Wallace
    Neil R. Crawford
    Jessica R. Riggleman
    Leigh A. Ahrendtsen
    Charles G. Ledonio
    Journal of Robotic Surgery, 2020, 14 : 409 - 413
  • [43] A Hybrid Uniplanar Pedicle Screw System with a New Intermediate Screw for Minimally Invasive Spinal Fixation: A Finite Element Analysis
    Li, Jia
    Zhang, Li-Cheng
    Li, Jiantao
    Zhang, Hao
    Zhao, Jing-Xin
    Zhang, Wei
    BIOMED RESEARCH INTERNATIONAL, 2020, 2020
  • [44] Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine
    Vinod V. Rajan
    Vijay Kamath
    Ajoy Prasad Shetty
    S. Rajasekaran
    Indian Journal of Orthopaedics, 2010, 44 : 163 - 168
  • [45] Pedicle screw placement in the thoracic spine: a randomized comparison study of computer-assisted navigation and conventional techniques
    Wu Han
    Gao Zhong-li
    Wang Jin-cheng
    Li Ying-pu
    Xia Peng
    Jiang Rui
    CHINESE JOURNAL OF TRAUMATOLOGY, 2010, 13 (04) : 201 - 205
  • [46] Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine
    Rajan, Vinod V.
    Kamath, Vijay
    Shetty, Ajoy Prasad
    Rajasekaran, S.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2010, 44 (02) : 163 - 168
  • [47] Pedicle screw placement in the thoracic spine:a randomized comparison study of computer-assisted navigation and conventional techniques
    武汉
    高中礼
    王金成
    李英普
    夏鹏
    姜睿
    Chinese Journal of Traumatology, 2010, (04)
  • [48] Does the accuracy of pedicle screw placement differ between the attending surgeon and resident in navigated robotic-assisted minimally invasive spine surgery?
    Arnold B. Vardiman
    David J. Wallace
    Grant A. Booher
    Neil R. Crawford
    Jessica R. Riggleman
    Samantha L. Greeley
    Charles G. Ledonio
    Journal of Robotic Surgery, 2020, 14 : 567 - 572
  • [49] Does the accuracy of pedicle screw placement differ between the attending surgeon and resident in navigated robotic-assisted minimally invasive spine surgery?
    Vardiman, Arnold B.
    Wallace, David J.
    Booher, Grant A.
    Crawford, Neil R.
    Riggleman, Jessica R.
    Greeley, Samantha L.
    Ledonio, Charles G.
    JOURNAL OF ROBOTIC SURGERY, 2020, 14 (04) : 567 - 572
  • [50] Spine Bull's-Eye Robot Guidewire Placement With Pedicle Standard Axis View for Thoracic and Lumbar Pedicle Screw Fixation
    Zhang, Chunlin
    Wang, Zheng
    Zhang, Changsheng
    Chen, Fuan
    Zhang, Hua
    Yan, Xu
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (07): : E191 - E198