Calibration and application of radiometer for low-temperature infrared radiation source

被引:0
|
作者
Shi, Qingfeng [1 ,2 ]
Gao, Yanze [1 ,2 ]
Li, Zhuo [1 ,2 ]
Shi, Rui [1 ,2 ]
机构
[1] Beijing Inst Technol, Sch Opt & Photon, Beijing 100081, Peoples R China
[2] Beijing Key Lab Precis Optoelect Measurement Inst, Beijing 100081, Peoples R China
基金
中国博士后科学基金;
关键词
radiation calibration; low-temperature infrared radiation; infrared radiometer;
D O I
10.1117/12.2606178
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In response to the current increasing demand for calibration of low-temperature infrared radiation sources with a large dynamic range, this paper studies a calibration method for low-temperature infrared radiation sources with a temperature range of 100-450K. The improvement of the signal-to-noise ratio of the input signal by related detection techniques is analyzed. Developed a radiation calibration device, also called a radiometer, analyzed the calibration and measurement process of the radiometer, and established a corresponding mathematical model. The experimental results show that the minimum radiation intensity that can be measured by the radiometer is 2.47x10(-12) W, and the response rate of the radiometer is 102878.38 V/W. Use the calibrated radiometer to measure the target simulator. The source and contribution of the uncertainty of the radiometer in the measurement process are analyzed.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] TEMPERATURE CALIBRATION OF AN INFRARED RADIATION SOURCE
    SCHOLL, MS
    APPLIED OPTICS, 1980, 19 (21): : 3622 - 3625
  • [2] TEMPERATURE CALIBRATION OF LOW-TEMPERATURE INFRARED CELLS
    FORD, TA
    SETO, PF
    FALK, M
    SPECTROCHIMICA ACTA PART A-MOLECULAR SPECTROSCOPY, 1969, A 25 (09): : 1650 - &
  • [3] Low-temperature infrared radiation thermometry at NMIJ
    Ishii, J
    Ono, A
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 7, PTS 1 AND 2, 2003, 684 : 657 - 662
  • [4] LOW-TEMPERATURE INFRARED SOURCE CALIBRATION AND TRACEABILITY AT ARNOLD-ENGINEERING-DEVELOPMENT-CENTER (AEDC)
    LITTLE, HR
    HIATT, J
    LIENEMANN, KA
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1983, 416 : 136 - 144
  • [5] LOW-TEMPERATURE SOURCE RADIATION FOR GREENHOUSE HEATING
    ALIG, CS
    ALDRICH, RA
    TRANSACTIONS OF THE ASAE, 1974, 17 (03): : 521 - 525
  • [6] Low-temperature radiator for verification and calibration of radiation thermometers
    Zuev, A. V.
    Rozhkov, V. A.
    MEASUREMENT TECHNIQUES, 2013, 56 (07) : 805 - 808
  • [7] Low-temperature radiator for verification and calibration of radiation thermometers
    A. V. Zuev
    V. A. Rozhkov
    Measurement Techniques, 2013, 56 : 805 - 808
  • [8] Application of the blackbody radiation source for characteristics measurement of submillimeter low-temperature direct detectors
    Vystavkin, AN
    Pestriakov, AV
    Vinogradov, EA
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 559 (02): : 570 - 572
  • [9] Gallium Fixed-point Blackbody Radiation Source for Calibration of Sea Surface Temperature Radiometer
    Zhang Kailin
    Zhao Zhiyuan
    Ren Chuanyun
    Yang Minglun
    Qu Liqin
    ACTA PHOTONICA SINICA, 2024, 53 (01)
  • [10] COLLIMATED CALIBRATION SOURCE OF INFRARED RADIATION
    PAVLYUKOV, AK
    KHOLOPOV, GK
    PARAMONOV, PI
    FILIPKOV, VD
    KHALIULLIN, FM
    SOVIET JOURNAL OF OPTICAL TECHNOLOGY, 1981, 48 (03): : 150 - 153