Manual ventilation devices in neonatal resuscitation: Tidal volume and positive pressure-provision

被引:49
|
作者
Roehr, Charles C. [1 ]
Kelm, Marcus [1 ]
Fischer, Hendrik S. [1 ]
Buehrer, Christoph [1 ]
Schmalisch, Gerd [1 ]
Proquitte, Hans [1 ]
机构
[1] Charite, Clin Naonatol, D-10117 Berlin, Germany
关键词
Resuscitation; Neopuff; T-piece resuscitator; Self-inflating bag; Neonate; Preterm; Pressure; VLBW; CPAP; Respiratory function monitor; AIRWAY PRESSURE; LUNG INJURY; GUIDELINES; EQUIPMENT; INFANTS;
D O I
10.1016/j.resuscitation.2009.10.008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Excessive peak inspiratory pressures (PIP) and high tidal volumes (Vt) during manual ventilation can be detrimental to the neonatal lung. We compared the influence of different manual ventilation devices and individual professional experience oil the extent of applied Vt and PIP in simulated neonatal resuscitation. Material and methods: One hundred and twenty medical professionals were studied. Ail intubated mannequin (equivalent to 1.0 kg neonate) was ventilated using two different devices: a self-inflating bag and a T-piece resuscitator. Target value was a PIP of 20 cm H2O. Applied PIP and the resulting Vt were recorded continuously using a respiratory function monitor (CO2SMO+, Novametrix, USA). Results: Vt and PIP provision was significantly higher in SI-bags, compared to T-piece devices: median (interquartile range) PIP 25.6 (18.2) cm H2O vs 19.7 (3.2) cm H2O (p < 0.0005), and Vt 5.1(3.2) ml vs Vt 3.6 (0.8) ml (p < 0.0005) respectively. The intersubject variability of Vt and PIP provision was distinctly higher in SI-bags, compared to T-piece devices. Professional experience had no significant impact on the level and the variability of Vt or PIP provided. Conclusion: Use of T-piece devices guarantees reliable and constant Vt and PIP provision, irrespective of individual, operator dependent variables. Methods to measure and to avoid excessive tidal volumes in neonatal resuscitation need to be developed. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:202 / 205
页数:4
相关论文
共 50 条
  • [31] Effects of tidal volume and positive end-expiratory pressure during resuscitation of very premature lambs
    Probyn, ME
    Hooper, SB
    Dargaville, PA
    McCallion, N
    Harding, R
    Morley, CJ
    ACTA PAEDIATRICA, 2005, 94 (12) : 1764 - 1770
  • [32] Evolution of tidal volume requirements during the neonatal period in preterm infants supported by pressure limited ventilation
    Greenough, Anne
    Ali, Kamal
    Mughal, Sabina
    Cockar, Iram
    EUROPEAN RESPIRATORY JOURNAL, 2017, 50
  • [33] Effect of tidal volume, sampling duration, and cardiac contractility on pulse pressure and stroke volume variation during positive-pressure ventilation
    Kim, Hyung Kook
    Pinsky, Michael R.
    CRITICAL CARE MEDICINE, 2008, 36 (10) : 2858 - 2862
  • [34] Noninvasive Positive Pressure Ventilation or Conventional Mechanical Ventilation for Neonatal Continuous Positive Airway Pressure Failure
    Badiee, Zohreh
    Nekooie, Babak
    Mohammadizadeh, Majid
    INTERNATIONAL JOURNAL OF PREVENTIVE MEDICINE, 2014, 5 (08) : 1045 - 1053
  • [35] Continuous positive airway pressure during neonatal resuscitation
    Halamek, LP
    Morley, C
    CLINICS IN PERINATOLOGY, 2006, 33 (01) : 83 - +
  • [36] Comparison of low and high positive end-expiratory pressure during low tidal volume ventilation
    Edmark, Lennart
    Ostberg, Erland
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2020, 37 (06) : 516 - 517
  • [37] Volume-targeted mask ventilation during simulated neonatal resuscitation
    Law, Brenda Hiu Yan
    Schmoelzer, Georg M.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2024, 109 (02): : 217 - 220
  • [38] DIFFERENT GAS FLOW RATES AND EFFECTS ON TIDAL VOLUME AND MASK LEAK DURING POSITIVE PRESSURE VENTILATION
    Schilleman, K.
    Schmoelzer, G. M.
    Kamlin, C. O. F.
    Pas, A. B. te
    Morley, C. J.
    Davis, P. G.
    PEDIATRIC RESEARCH, 2010, 68 : 478 - 478
  • [39] Influence of tidal volume and positive end-expiratory pressure on ventilation distribution and oxygenation during one-lung ventilation
    Zhao, Zhanqi
    Wang, Wei
    Zhang, Zuojing
    Xu, Meiying
    Frerichs, Inez
    Wu, Jingxiang
    Moeller, Knut
    PHYSIOLOGICAL MEASUREMENT, 2018, 39 (03)
  • [40] CARDIOPULMONARY-RESUSCITATION WITHOUT INTERMITTENT POSITIVE PRESSURE VENTILATION
    OKAMOTO, K
    KISHI, H
    CHOI, H
    MORIOKA, T
    RESUSCITATION, 1993, 26 (03) : 251 - 260