An online replanning method using warm start optimization and aperture morphing for flattening-filter-free beams

被引:13
|
作者
Ahunbay, Ergun E. [1 ]
Ates, O. [1 ]
Li, X. A. [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
关键词
flattening-filter-free beams; adaptive radiotherapy; online replanning; MR-linac; INTERFRACTIONAL VARIATIONS; RADIOTHERAPY; PROSTATE; THERAPY; CANCER;
D O I
10.1118/1.4955439
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In a situation where a couch shift for patient positioning is not preferred or prohibited (e. g., MR-linac), segment aperture morphing (SAM) can address target dislocation and deformation. For IMRT /VMAT with flattening-filter-free (FFF) beams, however, SAM method would lead to an adverse translational dose effect due to the beam unflattening. Here the authors propose a new two-step process to address both the translational effect of FFF beams and the target deformation. Methods: The replanning method consists of an offline and an online step. The offline step is to create a series of preshifted-plans (PSPs) obtained by a so-called "warm start" optimization (starting optimization from the original plan, rather than from scratch) at a series of isocenter shifts. The PSPs all have the same number of segments with very similar shapes, since the warm start optimization only adjusts the MLC positions instead of regenerating them. In the online step, a new plan is obtained by picking the closest PSP or linearly interpolating the MLC positions and the monitor units of the closest PSPs for the shift determined from the image of the day. This two-step process is completely automated and almost instantaneous (no optimization or dose calculation needed). The previously developed SAM algorithm is then applied for daily deformation. The authors tested the method on sample prostate and pancreas cases. Results: The two-step interpolation method can account for the adverse dose effects from FFF beams, while SAM corrects for the target deformation. Plan interpolation method is effective in diminishing the unflat beam effect and may allow reducing the required number of PSPs. The whole process takes the same time as the previously reported SAM process (5-10 min). Conclusions: The new two-step method plus SAM can address both the translation effects of FFF beams and target deformation, and can be executed in full automation except the delineation of target contour required by the SAM process. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4575 / 4584
页数:10
相关论文
共 50 条
  • [1] Applicability of an Aperture-Morphing Based Replanning Algorithm for Flattening Filter-Free Beams
    Ahunbay, E. E.
    Li, A. X.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S930 - S931
  • [2] Technical Note: A fast online adaptive replanning method for VMAT using flattening filter free beams
    Ates, Ozgur
    Ahunbay, Ergun E.
    Moreau, Michel
    Li, X. Allen
    MEDICAL PHYSICS, 2016, 43 (06) : 2756 - 2764
  • [3] Implementation of An Online Replanning Tool for VMAT Using Flattening Filter-Free Beams
    Ates, O.
    Ahunbay, E.
    Li, X.
    MEDICAL PHYSICS, 2015, 42 (06) : 3293 - 3293
  • [4] Clinical Subtleties of Flattening-Filter-Free Beams
    Corns, Robert
    Thomas, Steven
    Huang, Vicky
    MEDICAL PHYSICS, 2014, 41 (08) : 21 - 22
  • [5] A Fast Adaptive Replanning Method for VMAT Using Flattening Filter-Free Beams
    Ates, O.
    Ahunbay, E. E.
    Moreau, M.
    Li, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S31 - S31
  • [6] Plan Optimization for IMRT with Flattening-filter-free (FFF) Beams by using Compressed Sensing Technique
    Kim, T.
    Wang, L.
    Mok, E.
    Suh, T.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S721 - S721
  • [7] Comparison of Aperture Complexity Between VMAT Plans Using Conventionally Flattened (FF) and Flattening-Filter-Free (FFF) Beams
    Liu, Q.
    Liang, J.
    MEDICAL PHYSICS, 2022, 49 (06) : E422 - E423
  • [8] Study of the ionization chamber response to flattening-filter-free photon beams
    Bresolin, A.
    Bonfantini, F.
    Stucchi, C. G.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    RADIATION MEASUREMENTS, 2017, 97 : 47 - 53
  • [9] Comparison of Head Scatters in Flattening-Filter-Free and Flattened Photon Beams
    Chen, Y.
    Shih, R.
    Afghan, M.
    Gopal, A.
    Chen, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [10] An Online Replanning Technique with Deformation- Based Aperture Morphing and Weight Optimization
    Ahunbay, E.
    Godley, A.
    Li, X.
    MEDICAL PHYSICS, 2009, 36 (06)