Study of the ionization chamber response to flattening-filter-free photon beams

被引:2
|
作者
Bresolin, A. [1 ]
Bonfantini, F. [2 ]
Stucchi, C. G. [2 ]
Mongioj, V. [2 ]
Carrara, M. [2 ]
Pignoli, E. [2 ]
机构
[1] Univ Milan, Dept Phys, Via Giovanni Celoria 16, I-20133 Milan, Italy
[2] Fdn IRCCS Ist Nazl Tumori, Med Phys Unit, Via Venezian 1, I-20133 Milan, Italy
关键词
Recombination correction factor; Jaffe plot; Flattening-filter-free; Pulse repetition frequency; Dose per pulse; HIGH-ENERGY PHOTON; ION RECOMBINATION; P-ION; RADIATION; COLLECTION; DOSIMETRY;
D O I
10.1016/j.radmeas.2016.12.011
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The commissioning of up-to-date linacs, designed to deliver flattening-filter-free (FFF) photon beams, is performed by measurements using air-vented ionization chambers which are often employed for traditional beams. The aim is to select the ionization chambers which are able to minimize the ion recombination effect, critical for these high intensity beams. In order to provide operators with information about the response to FFF beams of widely used chambers, the recombination correction factor k(s) and the dose-per-pulse(DPP) were determined in different irradiation conditions. The study, performed by using devices available in most radiotherapy departments, confirmed the reliability of the Boag model for FFF beams as well as with-flattening-filter (WFF) beams. The results showed that k(s) factor largely depends on the spacing between the collecting electrodes (d(el))in the ionization chamber and on the DPP of the linac output, which in turn depends on beam energy but is independent of monitor unit rate (MU/min). For plane-parallel and pin-point chambers, whose d(el) is minimal, the k(s) factor is negligible, whereas for Farmer type chambers the recombination could be important and hence the correction is to be applied (also at voltages higher than the recommended one). (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 50 条
  • [1] Ionization Chamber Perturbation in Flattening Filter Free Beams
    Czarnecki, D.
    von Voigts-Rhetz, P.
    Zink, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3456 - 3456
  • [2] Comparison of Head Scatters in Flattening-Filter-Free and Flattened Photon Beams
    Chen, Y.
    Shih, R.
    Afghan, M.
    Gopal, A.
    Chen, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Clinical Subtleties of Flattening-Filter-Free Beams
    Corns, Robert
    Thomas, Steven
    Huang, Vicky
    MEDICAL PHYSICS, 2014, 41 (08) : 21 - 22
  • [4] Assessment of ion recombination correction and polarity effects for specific ionization chambers in flattening-filter-free photon beams
    Martin-Martin, Guadalupe
    Borja Aguilar, Pedro
    Barbes, Benigno
    Guibelalde, Eduardo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 67 : 176 - 184
  • [5] Feasibility Study of Using Flattening-Filter-Free Photon Beams to Deliver Conventional Flat Beam
    Potter, N.
    Lebron, S.
    Yan, G.
    Li, J.
    Liu, C.
    Lu, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3580 - 3581
  • [6] Ion-recombination correction for different ionization chambers in high dose rate flattening-filter-free photon beams
    Lang, Stephanie
    Hrbacek, Jan
    Leong, Aidan
    Kloeck, Stephan
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09):
  • [7] Dose Enhancement for the Flattening-Filter-Free and Flattening-Filter Photon Beams in Nanoparticle-Enhanced Radiotherapy: A Monte Carlo Phantom Study
    Martelli, Stefano
    Chow, James C. L.
    NANOMATERIALS, 2020, 10 (04)
  • [8] Technical Note: Correction for intra-chamber dose gradients in reference dosimetry of flattening-filter-free MV photon beams
    Ruggieri, Ruggero
    Naccarato, Stefania
    Stavrev, Pavel
    Stavreva, Nadejda
    Pasetto, Stefano
    Salamone, Ignazio
    Alongi, Filippo
    MEDICAL PHYSICS, 2016, 43 (08) : 4729 - 4733
  • [9] Use and Choice of Ionization Chambers for the Commissioning of Flattened and Flattening-Filter-Free Photon Beams: Determination of Recombination Correction Factor (ks)
    Stucchi, C.
    Bresolin, A.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    Bonfantini, F.
    MEDICAL PHYSICS, 2014, 41 (06) : 371 - 371
  • [10] Stereotactic Cone Output Factors of Flattening-Filter-Free Mode Photon Beams; a Plastic Scintillation Detector Measurement Study
    Kim, J.
    Kahn, D.
    Qian, X.
    Lu, S.
    Hsia, A. T.
    Xu, Z.