Study of the ionization chamber response to flattening-filter-free photon beams

被引:2
|
作者
Bresolin, A. [1 ]
Bonfantini, F. [2 ]
Stucchi, C. G. [2 ]
Mongioj, V. [2 ]
Carrara, M. [2 ]
Pignoli, E. [2 ]
机构
[1] Univ Milan, Dept Phys, Via Giovanni Celoria 16, I-20133 Milan, Italy
[2] Fdn IRCCS Ist Nazl Tumori, Med Phys Unit, Via Venezian 1, I-20133 Milan, Italy
关键词
Recombination correction factor; Jaffe plot; Flattening-filter-free; Pulse repetition frequency; Dose per pulse; HIGH-ENERGY PHOTON; ION RECOMBINATION; P-ION; RADIATION; COLLECTION; DOSIMETRY;
D O I
10.1016/j.radmeas.2016.12.011
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The commissioning of up-to-date linacs, designed to deliver flattening-filter-free (FFF) photon beams, is performed by measurements using air-vented ionization chambers which are often employed for traditional beams. The aim is to select the ionization chambers which are able to minimize the ion recombination effect, critical for these high intensity beams. In order to provide operators with information about the response to FFF beams of widely used chambers, the recombination correction factor k(s) and the dose-per-pulse(DPP) were determined in different irradiation conditions. The study, performed by using devices available in most radiotherapy departments, confirmed the reliability of the Boag model for FFF beams as well as with-flattening-filter (WFF) beams. The results showed that k(s) factor largely depends on the spacing between the collecting electrodes (d(el))in the ionization chamber and on the DPP of the linac output, which in turn depends on beam energy but is independent of monitor unit rate (MU/min). For plane-parallel and pin-point chambers, whose d(el) is minimal, the k(s) factor is negligible, whereas for Farmer type chambers the recombination could be important and hence the correction is to be applied (also at voltages higher than the recommended one). (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 50 条
  • [31] Secondary MU Calculation for Flattening Filter Free (FFF) Photon Beams
    Kim, M.
    Zhu, T.
    MEDICAL PHYSICS, 2015, 42 (06) : 3495 - 3495
  • [32] Commissioning Measurements of Flattening Filter and Flattening Filter Free Photon Beams Using a TrueBeam Stx® Linear Accelerator
    Saidi K.
    El Baydaoui R.
    El Gouach H.
    Mesradi M.R.
    Naim A.
    Iranian Journal of Medical Physics, 2021, 18 (01) : 49 - 62
  • [33] Radiosurgery with flattening-filter-free techniques in the treatment of brain metastases
    Rieber, J.
    Tonndorf-Martini, E.
    Schramm, O.
    Rhein, B.
    Stefanowicz, S.
    Kappes, J.
    Hoffmann, H.
    Lindel, K.
    Debus, J.
    Rieken, S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 789 - 796
  • [34] Depth Dose Enhancement on Flattening-Filter-Free Photon Beam: A Monte Carlo Study in Nanoparticle-Enhanced Radiotherapy
    Chow, James C. L.
    APPLIED SCIENCES-BASEL, 2020, 10 (20): : 1 - 10
  • [35] An online replanning method using warm start optimization and aperture morphing for flattening-filter-free beams
    Ahunbay, Ergun E.
    Ates, O.
    Li, X. A.
    MEDICAL PHYSICS, 2016, 43 (08) : 4575 - 4584
  • [36] Flattening Filter Free Photon Beams for Stereotactic Radiosurgery of Trigeminal Neuralgia: A Plan Comparison Study
    Mynampati, D.
    Jeong, K.
    Garg, M.
    Yaparpalvi, R.
    Tome, W.
    MEDICAL PHYSICS, 2019, 46 (06) : E656 - E656
  • [37] Contrast Enhancement for Portal Imaging in Nanoparticle-Enhanced Radiotherapy: A Monte Carlo Phantom Evaluation Using Flattening-Filter-Free Photon Beams
    Abdulle, Aniza
    Chow, James C. L.
    NANOMATERIALS, 2019, 9 (07)
  • [38] Evaluation of Ion Recombination Correction for Indigenously Developed Farmer Ion Chamber in Flattening Filter-Free Photon Beams
    Patwe, Parimal T.
    Deshpande, Sudesh S.
    Mahajan, Gajanan R.
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (02) : 279 - 284
  • [39] Penumbra Measurement for a Flattening-Filter-Free Artiste Linear Accelerator
    Gayou, O.
    Johnson, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [40] Technical Report: Evaluation of peripheral dose for flattening filter free photon beams
    Covington, E. L.
    Ritter, T. A.
    Moran, J. M.
    Owrangi, A. M.
    Prisciandaro, J. I.
    MEDICAL PHYSICS, 2016, 43 (08) : 4789 - 4796