Imaging of Soft Tissues Adjacent to Orthopedic Hardware: Comparison of 3-T and 1.5-T MRI

被引:28
|
作者
Farrelly, Cormac [1 ]
Davarpanah, Amir [2 ]
Brennan, Stephen [3 ]
Sampson, Mathew [4 ]
Eustace, Stephen J. [1 ]
机构
[1] Cappagh Natl Orthopaed Hosp, Dept Radiol, Dublin 11, Ireland
[2] Northwestern Univ Hosp, Dept Radiol, Feinberg Sch Med, Chicago, IL USA
[3] Cappagh Natl Orthopaed Hosp, Dept Orthoped Surg, Dublin 11, Ireland
[4] Sports Surg Clin, Dept Radiol, Dublin, Ireland
关键词
1.5; T; 3; artifact reduction; MRI; susceptibility; TOTAL HIP-ARTHROPLASTY; MUSCULOSKELETAL MRI; METALLIC IMPLANTS; SPIN-ECHO; ARTIFACTS; REPLACEMENT; EXPERIENCE; REDUCTION; IMAGES; 3.0T;
D O I
10.2214/AJR.08.1740
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare metal artifact reduction techniques at 1.5-T and 3-T MRI. MATERIALS AND METHODS. A titanium plate with steel screws was placed in a freshly harvested pig leg. The leg was imaged with 1.5-T and 3-T MRI. A T2-weighted turbo spin-echo sequence was used with echo-train lengths of 8, 16, 32, and 64 and a constant readout bandwidth of 31.2 kHz. The images were compared qualitatively, and the optimal echotrain length was selected. Images were acquired at the optimal echo-train length with four different readout bandwidths. Artifact was measured quantitatively, and image quality was ranked qualitatively. The qualitatively best image acquired at 1.5 T was compared with the qualitatively highest-ranked image acquired at 3 T. RESULTS. At both 1.5 T and 3 T, optimal images of equal quality were produced at echotrain lengths of 8 and 16. At higher readout bandwidths, there was quantitatively less artifact. The qualitatively best images were acquired at a readout bandwidth of 31.2 kHz at 1.5 T and 62.5 kHz at 3 T (Cronbach's alpha = 1.00). The optimal image at 3 T was qualitatively superior to that at 1.5 T. CONCLUSION. Optimizing image acquisition parameters in this phantom model resulted in similar quantitative susceptibility artifact at 3 T and 1.5 T and better qualitative images at 3 T than at 1.5 T.
引用
收藏
页码:W60 / W64
页数:5
相关论文
共 50 条
  • [1] Qualitative comparison of 3-T and 1.5-T MRI in the evaluation of epilepsy
    Phal, Pramit M.
    Usmanov, Alexander
    Nesbit, Gary M.
    Anderson, James C.
    Spencer, David
    Wang, Paul
    Helwig, Jonathan A.
    Roberts, Colin
    Hamilton, Bronwyn E.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (03) : 890 - 895
  • [2] Comparison Between 1.5-T and 3-T MRI for Fetal Imaging: Is There an Advantage to Imaging With a Higher Field Strength?
    Victoria, Teresa
    Johnson, Ann M.
    Edgar, J. Christopher
    Zarnow, Deborah M.
    Vossough, Arastoo
    Jaramillo, Diego
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 206 (01) : 195 - 201
  • [3] Comparison of the Accuracy Rates of 3-T and 1.5-T MRI of the Knee in the Diagnosis of Meniscal Tear
    Grossman, Jeffrey W.
    De Smet, Arthur A.
    Shinki, Kazuhiko
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (02) : 509 - 514
  • [4] Paramagnetic Rim Lesions in Multiple Sclerosis: Comparison of Visualization at 1.5-T and 3-T MRI
    Hemond, Christopher C.
    Reich, Daniel S.
    Dundamadappa, Sathish K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (01) : 120 - 129
  • [5] Conventional 3-T MRI and 1.5-T MR Arthrography of Femoroacetabular Impingement
    Robinson, Philip
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (03) : 509 - 515
  • [6] Quantification of cardiac iron in patients with thalassemia with 3-T MRI calibrated by 1.5-T MRI
    Xu, Fengming
    Luo, Chaotian
    Li, Meicheng
    Guan, Kaiming
    Peng, Fei
    Yang, Gaohui
    Peng, Peng
    ACTA RADIOLOGICA, 2023, 64 (06) : 2096 - 2103
  • [7] MR imaging of the brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary experience
    Tagliafico, Alberto
    Succio, Giulia
    Neumaier, Carlo Emanuele
    Serafini, Giovanni
    Ghidara, Matteo
    Calabrese, Massimo
    Martinoli, Carlo
    SKELETAL RADIOLOGY, 2011, 40 (06) : 717 - 724
  • [8] MR imaging of the brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary experience
    Alberto Tagliafico
    Giulia Succio
    Carlo Emanuele Neumaier
    Giovanni Serafini
    Matteo Ghidara
    Massimo Calabrese
    Carlo Martinoli
    Skeletal Radiology, 2011, 40 : 717 - 724
  • [9] Computational and experimental studies of an orthopedic implant: MRI-related heating at 1.5-T/64-MHz and 3-T/128-MHz
    Liu, Yan
    Chen, Ji
    Shellock, Frank G.
    Kainz, Wolfgang
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 37 (02) : 491 - 497
  • [10] Assessment of the available evidence for the use of 7-Tesla (T) magnetic resonance imaging (MRI) in neurological and musculoskeletal disorders, with comparison to 3-T and 1.5-T MRI: A systematic scoping review
    Radojewski, Piotr
    Piredda, Gian Franco
    Bonanno, Gabriele
    Lovblad, Karl-Olof
    Vargas, Maria Isabel
    Sutter, Reto
    Nanz, Daniel
    Karrer, Tanya
    Salanti, Georgia
    Wiest, Roland
    EUROPEAN JOURNAL OF NEUROLOGY, 2025, 32 (01)