Evaluation of the Abbott m2000rt RealTime™ HIV-1 assay with manual sample preparation compared with the ROCHE COBAS® AmpliPrep™/AMPLICOR™ HIV-1 MONITOR® v1.5 using specimens from East Africa

被引:31
|
作者
Crump, John A. [1 ,2 ,3 ,4 ]
Scott, Lesley E. [5 ]
Msuya, Emma [3 ]
Morrissey, Anne B. [1 ]
Kimaro, Ekyafyose E. [3 ]
Shao, John F. [3 ,4 ]
Stevens, Wendy S. [5 ,6 ]
机构
[1] Duke Univ, Med Ctr, Dept Med, Div Infect Dis & Int Health, Durham, NC 27710 USA
[2] Duke Univ, Duke Global Hlth Inst, Durham, NC 27710 USA
[3] Kilimanjaro Christian Med Ctr, Moshi, Tanzania
[4] Tumaini Univ, Kilimanjaro Christian Med Coll, Moshi, Tanzania
[5] Univ Witwatersrand, Dept Mol Med & Haematol, Sch Pathol, Fac Hlth Sci, ZA-2050 Wits, South Africa
[6] Natl Hlth Lab Serv, Johannesburg, South Africa
基金
美国国家卫生研究院;
关键词
HIV; Reverse transcriptase polymerase chain reaction; Viral load; Validation studies; Tanzania; GENETICALLY DIVERSE PANEL; DRIED BLOOD SPOTS; VIRUS TYPE-1 RNA; QUANTITATIVE DETECTION; PERFORMANCE; NUCLISENS; PLASMA; TANZANIA; PCR; QUANTIFICATION;
D O I
10.1016/j.jviromet.2009.08.013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of the Abbott m2000rt RealTime (TM) HIV-1 assay (RealTime HIV-1) with manual sample preparation was compared against the ROCHE COBAS (R) AmpliPrep (TM)/AMPLICOR (TM) HIV-1 MONITOR Test v0.5 (CAP/CA HIV-1) using samples collected from 100 donors infected with HIV and 20 donors not infected with HIV in northern Tanzania where HIV-1 subtypes A, C, D, and their recombinant forms predominate. The RealTime HIV-1 appeared to have more within-run variability at high HIV-1 RNA concentrations, but total assay variability over the dynamic range tested was within the manufacturer's claim of < 0.3 SD copies/mL Accuracy studies showed 100% concordance for positive and negative values. When continuous values were examined, CAP/CA HIV-1 yielded higher values than the RealTime HIV-1 at higher nominal HIV-1 RNA concentrations. The RealTime HIV-1 assay showed excellent linearity between 2.5 and 7.0 log copies/mL Of negative samples, 100% showed negative results, and > 95% of samples with nominal concentrations of 40 copies/mL were detected at >= 40 copies/mL by RealTime HIV-1. Manual sample preparation may contribute to higher total assay variability. This study suggests that the Abbott m2000rt RealTime HIV-1 assay with manual sample preparation is an acceptable and feasible alternative to the conventional ROCHE COBAS AmpliPrep/AMLICOR HIV-1 Monitor v1.5 assay and that the RealTime HIV-1 assay performs well on samples from East Africa. (c) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:218 / 222
页数:5
相关论文
共 43 条
  • [41] Comparative performance of the Biocentric Generic Viral Load, Roche CAP/CTM v1.5, Roche CAP/CTM v2.0 and m2000 Abbott assays for quantifying HIV-1 B and non-B strains: Underestimation of some CRF02 strains
    Avettand-Fenoel, V
    Melard, A.
    Gueudin, M.
    Maillard, A.
    Dina, J.
    Gousset, M.
    Chaix, M. L.
    Lerolle, N.
    Viard, J. P.
    Meyer, L.
    Plantier, J. C.
    Rouzioux, C.
    JOURNAL OF CLINICAL VIROLOGY, 2019, 110 : 36 - 41
  • [42] Improved Testing of Recent HIV-1 Infections with the BioRad Avidity Assay Compared to the Limiting Antigen Avidity Assay and BED Capture Enzyme Immunoassay: Evaluation Using Reference Sample Panels from the German Seroconverter Cohort
    Hauser, Andrea
    Santos-Hoevener, Claudia
    Meixenberger, Karolin
    Zimmermann, Ruth
    Somogyi, Sybille
    Fiedler, Stefan
    Hofmann, Alexandra
    Bartmeyer, Barbara
    Jansen, Klaus
    Hamouda, Osamah
    Bannert, Norbert
    Kuecherer, Claudia
    PLOS ONE, 2014, 9 (06):
  • [43] Evaluation of the accuracy of the Asante assay as a point-of-care rapid test for HIV-1 recent infections using serum bank specimens from blood donors in South Africa, July 2018-August 2021
    Singh, B.
    Mthombeni, J.
    Olorunfemi, G.
    Goosen, M.
    Cutler, E.
    Julius, H.
    Brukwe, Z.
    Puren, A.
    SAMJ SOUTH AFRICAN MEDICAL JOURNAL, 2023, 113 (10): : 1437 - 1443