Evaluation of the Abbott m2000rt RealTime™ HIV-1 assay with manual sample preparation compared with the ROCHE COBAS® AmpliPrep™/AMPLICOR™ HIV-1 MONITOR® v1.5 using specimens from East Africa
被引:31
|作者:
Crump, John A.
[1
,2
,3
,4
]
Scott, Lesley E.
[5
]
Msuya, Emma
[3
]
Morrissey, Anne B.
[1
]
Kimaro, Ekyafyose E.
[3
]
Shao, John F.
[3
,4
]
Stevens, Wendy S.
[5
,6
]
机构:
[1] Duke Univ, Med Ctr, Dept Med, Div Infect Dis & Int Health, Durham, NC 27710 USA
[2] Duke Univ, Duke Global Hlth Inst, Durham, NC 27710 USA
[3] Kilimanjaro Christian Med Ctr, Moshi, Tanzania
[4] Tumaini Univ, Kilimanjaro Christian Med Coll, Moshi, Tanzania
[5] Univ Witwatersrand, Dept Mol Med & Haematol, Sch Pathol, Fac Hlth Sci, ZA-2050 Wits, South Africa
[6] Natl Hlth Lab Serv, Johannesburg, South Africa
基金:
美国国家卫生研究院;
关键词:
HIV;
Reverse transcriptase polymerase chain reaction;
Viral load;
Validation studies;
Tanzania;
GENETICALLY DIVERSE PANEL;
DRIED BLOOD SPOTS;
VIRUS TYPE-1 RNA;
QUANTITATIVE DETECTION;
PERFORMANCE;
NUCLISENS;
PLASMA;
TANZANIA;
PCR;
QUANTIFICATION;
D O I:
10.1016/j.jviromet.2009.08.013
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
The performance of the Abbott m2000rt RealTime (TM) HIV-1 assay (RealTime HIV-1) with manual sample preparation was compared against the ROCHE COBAS (R) AmpliPrep (TM)/AMPLICOR (TM) HIV-1 MONITOR Test v0.5 (CAP/CA HIV-1) using samples collected from 100 donors infected with HIV and 20 donors not infected with HIV in northern Tanzania where HIV-1 subtypes A, C, D, and their recombinant forms predominate. The RealTime HIV-1 appeared to have more within-run variability at high HIV-1 RNA concentrations, but total assay variability over the dynamic range tested was within the manufacturer's claim of < 0.3 SD copies/mL Accuracy studies showed 100% concordance for positive and negative values. When continuous values were examined, CAP/CA HIV-1 yielded higher values than the RealTime HIV-1 at higher nominal HIV-1 RNA concentrations. The RealTime HIV-1 assay showed excellent linearity between 2.5 and 7.0 log copies/mL Of negative samples, 100% showed negative results, and > 95% of samples with nominal concentrations of 40 copies/mL were detected at >= 40 copies/mL by RealTime HIV-1. Manual sample preparation may contribute to higher total assay variability. This study suggests that the Abbott m2000rt RealTime HIV-1 assay with manual sample preparation is an acceptable and feasible alternative to the conventional ROCHE COBAS AmpliPrep/AMLICOR HIV-1 Monitor v1.5 assay and that the RealTime HIV-1 assay performs well on samples from East Africa. (c) 2009 Elsevier B.V. All rights reserved.