A Practical, Automated Quality Assurance Method for Measuring Spatial Resolution in PET

被引:29
|
作者
Lodge, Martin A. [1 ]
Rahmim, Arman [1 ]
Wahl, Richard L. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Sch Med, Div Nucl Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD USA
[2] Johns Hopkins Univ, Sch Med, Sidney Kimmel Comprehens Canc Ctr Johns Hopkins, Baltimore, MD USA
关键词
spatial resolution; quality assurance; multicenter trials; PET; IMAGE-RECONSTRUCTION; EMISSION-TOMOGRAPHY; PERFORMANCE;
D O I
10.2967/jnumed.108.060079
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The use of different scanners, acquisition protocols, and reconstruction algorithms has been identified as a problem that limits the use of PET in multicenter trials. The aim of this project was to aid standardization of data collection by developing a quality assurance method for measuring the spatial resolution achieved with clinical imaging protocols. Methods: A commercially available 68Ge cylinder phantom (diameter, 20 cm) with a uniform activity concentration was positioned in the center of the PET field of view, and an image was acquired using typical clinical parameters. Spatial resolution was measured by artificially generating an object function (O) with uniform activity within a 20-cm-diameter cylinder, assuming no noise and perfect spatial resolution, centered on the original image (I); dividing F[I] by F[O], where F indicates a 2-dimensional Fourier transform, to produce a modulation transfer function; and taking the inverse Fourier transform of the modulation transfer function to produce a point-spread function in image space. The method was validated using data acquired on 4 different commercial PET systems. Results: Spatial resolution on the Discovery LS was measured at 5.75 +/- 0.58 mm, compared with 5.54 +/- 0.19 mm from separate point source measurements. Variability of the resolution measurements differed between scanners and protocols, but the typical SD was approximately 0.15 mm when iterative reconstruction was used. The potential for predicting resolution recovery coefficients for small objects was also demonstrated. Conclusion: The proposed method does not require elaborate phantom preparation and is practical to perform, and data analysis is fully automated. This approach is useful for evaluating clinical reconstruction protocols across varying scanners and reconstruction algorithms and should greatly aid standardization of data collection between centers.
引用
收藏
页码:1307 / 1314
页数:8
相关论文
共 50 条
  • [32] QUALITY ASSURANCE FOR MEASURING AND TESTING FACILITIES
    ROMMERSKIRCH, W
    F&M-FEINWERKTECHNIK & MESSTECHNIK, 1975, 83 (07): : 349 - 353
  • [33] Pet Food Quality Assurance and Safety and Quality Assurance Survey within the Costa Rican Pet Food Industry
    Leiva, Astrid
    Molina, Andrea
    Redondo-Solano, Mauricio
    Artavia, Graciela
    Rojas-Bogantes, Lizeth
    Granados-Chinchilla, Fabio
    ANIMALS, 2019, 9 (11):
  • [34] A PRACTICAL METHOD FOR MEASURING THE SPATIAL FREQUENCY RESPONSE OF LIGHT FIELD CAMERAS
    Firmenich, Damien
    Baboulaz, Loic
    Suesstrunk, Sabine
    2014 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2014, : 5776 - 5780
  • [35] Measuring the effect of count rate on PET spatial resolution using a uniform cylinder phantom
    Lodge, Martin
    Rahmim, Arman
    Wong, Dean
    Wahl, R.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [36] A novel method for harmonization of PET image spatial resolution without phantoms
    Carbonell, Felix
    Zijdenbos, Alex P.
    Hempel, Evan
    Hajos, Mihaly
    Bedell, Barry J.
    EJNMMI PHYSICS, 2025, 12 (01):
  • [37] Respiratory Gating Quality Assurance: A Simple Method to Achieve Millisecond Temporal Resolution
    McCabe, B.
    Wiersma, R.
    MEDICAL PHYSICS, 2014, 41 (06) : 185 - 185
  • [38] Quality Assurance for Spatial Research Data
    Wagner, Michael
    Henzen, Christin
    ISPRS INTERNATIONAL JOURNAL OF GEO-INFORMATION, 2022, 11 (06)
  • [39] Method for measuring the spatial characteristics of high-resolution diffraction gratings
    Shishova, M. V.
    Odinokov, S. B.
    Zherdev, A. Y.
    Popov, V. V.
    HOLOGRAPHY, DIFFRACTIVE OPTICS, AND APPLICATIONS VIII, 2018, 10818
  • [40] Quality assurance in cardiovascular CT: a practical guide
    White, S. K.
    Castellano, E.
    Gartland, N.
    Patel, T.
    Padley, S. P. G.
    Rubens, M. B.
    Nicol, E. D.
    CLINICAL RADIOLOGY, 2016, 71 (08) : 729 - 738