Comparison of Axxent-Xoft, 192Ir and 60Co high-dose-rate brachytherapy sources for image-guided brachytherapy treatment planning for cervical cancer

被引:11
|
作者
Mobit, P. N. [1 ,2 ]
Packianathan, S. [1 ]
He, R. [1 ]
Yang, C. C. [1 ]
机构
[1] Univ Mississippi, Med Ctr, Dept Radiat Oncol, Jackson, MS 39216 USA
[2] Cameroon Oncol Ctr, Douala, Cameroon
来源
BRITISH JOURNAL OF RADIOLOGY | 2015年 / 88卷 / 1052期
关键词
50-KV ELECTRONIC BRACHYTHERAPY; PARAMETERS; CARCINOMA;
D O I
10.1259/bjr.20150010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the dosimetric differences and similarities between treatment plans generated with Axxent-Xoft electronic brachytherapy source (Xoft-EBS), Ir-192 and Co-60 for tandem and ovoids (T&O) applicators. Methods: In this retrospective study, we replanned 10 patients previously treated with 192Ir high-dose-rate brachytherapy. Prescription was 7G x 34 fractions to Point A. For each original plan, we created two additional plans with Xoft-EBS and Co-60. The dose to each organ at risk (OAR) was evaluated in terms of V-35% and V-50%, the percentage volume receiving 35% and 50% of the prescription dose, respectively, and D-2cc, highest dose to a 2cm(3) volume of an OAR. Results: There was no difference between plans generated by Ir-192 and Co-60, but the plans generated using Xoft-EBS showed a reduction of up to 50% in V-35%, V-50% and D-2cc. The volumes of the 200% and 150% isodose lines, however, were 74% and 34% greater than the comparable volumes generated with the Ir-192 source. Point B dose was on average only 16% of the Point A dose for plans generated with Xoft-EBS compared with 30% for plans generated with Ir-192 or Co-60. Conclusion: The Xoft-EBS can potentially replace either Ir-192 or Co-60 in T&O treatments. Xoft-EBS offers either better sparing of the OARs compared with Ir-192 or Co-60 or at least similar sparing. Xoft-EBS-generated plans had higher doses within the target volume than Ir-192-or Co-60-generated ones. Advances in knowledge: This work presents newer knowledge in dosimetric comparison between Xoft-EBS, Ir-192 or Co-60 sources for T&O implants.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Image-Guided Interstitial High-Dose-Rate Brachytherapy in Hepatocellular Carcinoma
    Mohnike, Konrad
    Wieners, Gero
    Pech, Maciej
    Seidensticker, Max
    Ruehl, Ricarda
    Lopez-Haenninen, Enrique
    Ricke, Jens
    DIGESTIVE DISEASES, 2009, 27 (02) : 170 - 174
  • [42] Biological Planning for High-Dose-Rate Brachytherapy: Application to Cervical Cancer Treatment
    Lee, Eva K.
    Yuan, Fan
    Templeton, Alistair
    Yao, Rui
    Kiel, Krystyna
    Chu, James C. H.
    INTERFACES, 2013, 43 (05) : 462 - 476
  • [43] Evaluation of bone dose arising from skin cancer brachytherapy: A comparison between 192Ir and 60Co sources through Monte Carlo simulations
    Sheikholeslami, Sahar
    Khodaverdian, Shaghayegh
    Hashemzaei, Fatemeh
    Ghobadi, Parvin
    Ghorbani, Mahdi
    Farhood, Bagher
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 205
  • [44] Clinical application of 192Ir High-Dose-Rate brachytherapy in metastatic lymph nodes of the neck
    Lu, Hongling
    Sun, Yunchuan
    Gao, Yan
    Xiao, Li
    Zhou, Jianxi
    Yin, Xiaoming
    Guo, Wei
    Fan, Kui
    DISCOVER ONCOLOGY, 2023, 14 (01)
  • [45] Clinical application of 192Ir High-Dose-Rate brachytherapy in metastatic lymph nodes of the neck
    Hongling Lu
    Yunchuan Sun
    Yan Gao
    Li Xiao
    Jianxi Zhou
    Xiaoming Yin
    Wei Guo
    Kui Fan
    Discover Oncology, 14
  • [46] Image-guided interstitial high-dose-rate brachytherapy in the local treatment of intrathoracic malignancies
    Ferentinos, K.
    Tselis, N.
    Kolotas, C.
    Schirren, J.
    Baltas, D.
    Antonakakis, A.
    Ackermann, H.
    Zamboglou, N.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 : 57 - 57
  • [47] Evaluation of 192Ir, 60Co and 169Yb Sources for High Dose Rate Prostate Brachytherapy Inverse Planning Using An Interior Point Constraint Generation Algorithm
    Chung, E. Mok Tsze
    Safigholi, H.
    Nicolae, A.
    Davidson, M.
    Ravi, A.
    Aleman, D.
    Song, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3464 - 3465
  • [48] Direction modulated brachytherapy (DMBT) for treatment of cervical cancer: A planning study with 192 Ir, 60Co, and 169Yb HDR sources
    Safigholi, Habib
    Han, Dae Yup
    Mashouf, Shahram
    Soliman, Abraam
    Meigooni, Ali S.
    Owrangi, Amir
    Song, William Y.
    MEDICAL PHYSICS, 2017, 44 (12) : 6538 - 6547
  • [49] Treatment results of image-guided high-dose-rate interstitial brachytherapy for pelvic recurrence of uterine cancer
    Yoshida, Ken
    Yamazaki, Hideya
    Kotsuma, Tadayuki
    Takenaka, Tadashi
    Masui, Koji
    Yoshioka, Yasuo
    Uesugi, Yasuo
    Shimbo, Taiju
    Yoshikawa, Nobuhiko
    Yoshioka, Hiroto
    Narumi, Yoshifumi
    Tatsumi, Keiji
    Tanaka, Eiichi
    BRACHYTHERAPY, 2015, 14 (04) : 440 - 448
  • [50] In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancer
    Anagnostopoulos, G
    Baltas, D
    Geretschlaeger, A
    Martin, T
    Papagiannis, P
    Tselis, N
    Zamboglou, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04): : 1183 - 1191