Comparison of Axxent-Xoft, 192Ir and 60Co high-dose-rate brachytherapy sources for image-guided brachytherapy treatment planning for cervical cancer

被引:11
|
作者
Mobit, P. N. [1 ,2 ]
Packianathan, S. [1 ]
He, R. [1 ]
Yang, C. C. [1 ]
机构
[1] Univ Mississippi, Med Ctr, Dept Radiat Oncol, Jackson, MS 39216 USA
[2] Cameroon Oncol Ctr, Douala, Cameroon
来源
BRITISH JOURNAL OF RADIOLOGY | 2015年 / 88卷 / 1052期
关键词
50-KV ELECTRONIC BRACHYTHERAPY; PARAMETERS; CARCINOMA;
D O I
10.1259/bjr.20150010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the dosimetric differences and similarities between treatment plans generated with Axxent-Xoft electronic brachytherapy source (Xoft-EBS), Ir-192 and Co-60 for tandem and ovoids (T&O) applicators. Methods: In this retrospective study, we replanned 10 patients previously treated with 192Ir high-dose-rate brachytherapy. Prescription was 7G x 34 fractions to Point A. For each original plan, we created two additional plans with Xoft-EBS and Co-60. The dose to each organ at risk (OAR) was evaluated in terms of V-35% and V-50%, the percentage volume receiving 35% and 50% of the prescription dose, respectively, and D-2cc, highest dose to a 2cm(3) volume of an OAR. Results: There was no difference between plans generated by Ir-192 and Co-60, but the plans generated using Xoft-EBS showed a reduction of up to 50% in V-35%, V-50% and D-2cc. The volumes of the 200% and 150% isodose lines, however, were 74% and 34% greater than the comparable volumes generated with the Ir-192 source. Point B dose was on average only 16% of the Point A dose for plans generated with Xoft-EBS compared with 30% for plans generated with Ir-192 or Co-60. Conclusion: The Xoft-EBS can potentially replace either Ir-192 or Co-60 in T&O treatments. Xoft-EBS offers either better sparing of the OARs compared with Ir-192 or Co-60 or at least similar sparing. Xoft-EBS-generated plans had higher doses within the target volume than Ir-192-or Co-60-generated ones. Advances in knowledge: This work presents newer knowledge in dosimetric comparison between Xoft-EBS, Ir-192 or Co-60 sources for T&O implants.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Dosimetric comparison of brachytherapy sources for high-dose-rate treatment of endometrial cancer: 192Ir, 60Co and an electronic brachytherapy source
    Mobit, Paul N.
    Nguyen, Alex
    Packianathan, Satyaseelan
    He, Rui
    Yang, Claus C.
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1059):
  • [2] Dosimetric Comparison of Brachytherapy Sources for High Dose Rate Treatment of Endometrial Cancer: Electronic Brachytherapy, 192Ir and 60Co Sources
    Lozares, S.
    Font Gomez, J.
    Gandia, A.
    Miranda Burgos, A.
    Gascon, M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [3] Comparative Analysis of 60Co and 192Ir Sources in High Dose Rate Brachytherapy for Cervical Cancer
    Wen, Aiping
    Wang, Xianliang
    Wang, Bingjie
    Yan, Chuanjun
    Luo, Jingyue
    Wang, Pei
    Li, Jie
    CANCERS, 2022, 14 (19)
  • [4] Dosimetric comparison of 60Co and 192Ir high dose rate source used in brachytherapy treatment of cervical cancer
    Shukla, Arvind Kumar
    Jangid, Pawan Kumar
    Rajpurohit, Vikram Singh
    Verma, Atul
    Dangayach, Suresh Kumar
    Gagrani, Vaibhav
    Rathore, Narendra Kumar
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (06) : 1212 - 1215
  • [5] A dosimetric comparison of Xoft Axxent Electronic Brachytherapy and iridium-192 high-dose-rate brachytherapy in the treatment of endometrial cancer
    Dickler, Adam
    Kirk, Michael C.
    Coon, Alan
    Bernard, Damian
    Zusag, Tom
    Rotmensch, Jacob
    Wazer, David E.
    BRACHYTHERAPY, 2008, 7 (04) : 351 - 354
  • [6] Comparison of 60Co and 192Ir sources in HDR brachytherapy
    Strohmaier, Stefan
    Zwierzchowski, Grzegorz
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2011, 3 (04) : 199 - 208
  • [7] Radiobiological and dosimetric comparison of 60Co versus 192Ir high-dose-rate intracavitary-interstitial brachytherapy for cervical cancer
    Wen, Aiping
    Wang, Xianliang
    Wang, Bingjie
    Yan, Chuanjun
    Luo, Jingyue
    Wang, Pei
    Li, Jie
    RADIATION ONCOLOGY, 2022, 17 (01)
  • [8] Radiobiological and dosimetric comparison of 60Co versus 192Ir high-dose-rate intracavitary-interstitial brachytherapy for cervical cancer
    Aiping Wen
    Xianliang Wang
    Bingjie Wang
    Chuanjun Yan
    Jingyue Luo
    Pei Wang
    Jie Li
    Radiation Oncology, 17
  • [9] Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources
    Baltas, D
    Geramani, K
    Ioannidis, GT
    Hierholz, K
    Rogge, B
    Kolotas, C
    Müller-Sievers, K
    Milickovic, N
    Kober, B
    Zamboglou, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03): : 653 - 661
  • [10] Comparison of 192Ir, 169Yb, and 60Co high-dose rate brachytherapy sources for skin cancer treatment
    Safigholi, Habib
    Meigooni, Ali S.
    Song, William Y.
    MEDICAL PHYSICS, 2017, 44 (09) : 4426 - 4436