Dual-exposure technique for extending the dynamic range of x-ray flat panel detectors

被引:10
|
作者
Sisniega, A. [1 ,2 ]
Abella, M. [1 ,2 ]
Desco, M. [1 ,2 ,3 ]
Vaquero, J. J. [1 ,2 ]
机构
[1] Univ Carlos III Madrid, Dept Bioingn & Ingn Aeroespacial, Madrid, Spain
[2] IiSGM, Madrid, Spain
[3] Ctr Invest Red Salud Mental CIBERSAM, Madrid, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 02期
关键词
COMPUTED-TOMOGRAPHY; BREAST CT; RADIOGRAPHY; DESIGN; SYSTEM; NOISE;
D O I
10.1088/0031-9155/59/2/421
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work presents an approach to extend the dynamic range of x-ray flat panel detectors by combining two acquisitions of the same sample taken with two different x-ray photon flux levels and the same beam spectral configuration. In order to combine both datasets, the response of detector pixels was modelled in terms of mean and variance using a linear model. The model was extended to take into account the effect of pixel saturation. We estimated a joint probability density function (j-pdf) of the pixel values by assuming that each dataset follows an independent Gaussian distribution. This j-pdf was used for estimating the final pixel value of the high-dynamic-range dataset using amaximum likelihood method. The suitability of the pixel model for the representation of the detector signal was assessed using experimental data from a small-animal cone-beam micro-CT scanner equipped with a flat panel detector. The potential extension in dynamic range offered by our method was investigated for generic flat panel detectors using analytical expressions and simulations. The performance of the proposed dual-exposure approach in realistic imaging environments was compared with that of a regular single-exposure technique using experimental data from two different phantoms. Image quality was assessed in terms of signal-to-noise ratio, contrast, and analysis of profiles drawn on the images. The dynamic range, measured as the ratio between the exposure for saturation and the exposure equivalent to instrumentation noise, was increased from 76.9 to 166.7 when using our method. Dual-exposure results showed higher contrast-to-noise ratio and contrast resolution than the single-exposure acquisitions for the same x-ray dose. In addition, image artifacts were reduced in the combined dataset. This technique to extend the dynamic range of the detector without increasing the dose is particularly suited to image samples that contain both low and high attenuation regions.
引用
收藏
页码:421 / 439
页数:19
相关论文
共 50 条
  • [41] Signal and noise analysis of flat-panel sandwich detectors for single-shot dual-energy x-ray imaging
    Kim, Dong Woon
    Kim, Ho Kyung
    Youn, Hanbean
    Yun, Seungman
    Han, Jong Chul
    Kim, Junwoo
    Kam, Soohwa
    Tanguay, Jesse
    Cunningham, Ian A.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [42] Review X-ray photoconductors and stabilized a-Se for direct conversion digital flat-panel X-ray image-detectors
    S. O. Kasap
    J. A. Rowlands
    Journal of Materials Science: Materials in Electronics, 2000, 11 : 179 - 198
  • [43] Non-uniformity emendation technique for amorphous silicon flat-panel detectors used for industrial X-ray digital radiography
    Wang, Xiao
    Han, Yan
    Si, Jinjuan
    MEASUREMENT, 2008, 41 (07) : 817 - 822
  • [44] An X-ray tomograph based on a flat panel detector
    Constantin, F.
    Pavel, C.
    Bugoi, R.
    Toderas, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 685 - 689
  • [45] Development of flat panel X-ray image sensors
    Izumi, Y.
    Teranuma, O.
    Sato, T.
    Uehara, K.
    Okada, H.
    Tokuda, S.
    Sato, T.
    Shapu Giho/Sharp Technical Journal, 2001, (80): : 25 - 30
  • [46] The design and imaging characteristics of dynamic, solid-state, flat-panel x-ray image detectors for digital fluoroscopy and fluorography
    Cowen, A. R.
    Davies, A. G.
    Sivananthan, M. U.
    CLINICAL RADIOLOGY, 2008, 63 (10) : 1073 - 1085
  • [47] Progress report on the performance of real time selenium flat panel detectors for direct X-ray Imaging
    Tousignant, O
    Choquette, M
    Demers, Y
    Laperrière, L
    Leboeuf, J
    Honda, M
    Nishiki, M
    Takahashi, A
    Tsukamoto, A
    MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 503 - 510
  • [48] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Schieber, M
    Hermon, H
    Zentai, G
    Partain, L
    Street, RA
    Gilboa, H
    RADIOLOGY, 2001, 221 : 464 - 464
  • [49] A 128-Channel CMOS Charge Readout ASIC for Flat-Panel X-Ray Detectors
    Adachi, Susumu
    Nishimura, Akihiro
    Yoshimlita, Toshinori
    Tanabe, Koichi
    Okamura, Shoichi
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (06) : 3673 - 3683
  • [50] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Gilboa, H
    Zuck, A
    Dagan, O
    Vilensky, A
    Breen, BN
    Taieb, A
    Reisman, B
    Hermon, H
    Zentai, G
    Partain, L
    Street, R
    Ready, S
    X-RAY AND GAMMA-RAY DETECTORS AND APPLICATIONS IV, 2002, 4784 : 315 - 325