Dual-exposure technique for extending the dynamic range of x-ray flat panel detectors

被引:10
|
作者
Sisniega, A. [1 ,2 ]
Abella, M. [1 ,2 ]
Desco, M. [1 ,2 ,3 ]
Vaquero, J. J. [1 ,2 ]
机构
[1] Univ Carlos III Madrid, Dept Bioingn & Ingn Aeroespacial, Madrid, Spain
[2] IiSGM, Madrid, Spain
[3] Ctr Invest Red Salud Mental CIBERSAM, Madrid, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 02期
关键词
COMPUTED-TOMOGRAPHY; BREAST CT; RADIOGRAPHY; DESIGN; SYSTEM; NOISE;
D O I
10.1088/0031-9155/59/2/421
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work presents an approach to extend the dynamic range of x-ray flat panel detectors by combining two acquisitions of the same sample taken with two different x-ray photon flux levels and the same beam spectral configuration. In order to combine both datasets, the response of detector pixels was modelled in terms of mean and variance using a linear model. The model was extended to take into account the effect of pixel saturation. We estimated a joint probability density function (j-pdf) of the pixel values by assuming that each dataset follows an independent Gaussian distribution. This j-pdf was used for estimating the final pixel value of the high-dynamic-range dataset using amaximum likelihood method. The suitability of the pixel model for the representation of the detector signal was assessed using experimental data from a small-animal cone-beam micro-CT scanner equipped with a flat panel detector. The potential extension in dynamic range offered by our method was investigated for generic flat panel detectors using analytical expressions and simulations. The performance of the proposed dual-exposure approach in realistic imaging environments was compared with that of a regular single-exposure technique using experimental data from two different phantoms. Image quality was assessed in terms of signal-to-noise ratio, contrast, and analysis of profiles drawn on the images. The dynamic range, measured as the ratio between the exposure for saturation and the exposure equivalent to instrumentation noise, was increased from 76.9 to 166.7 when using our method. Dual-exposure results showed higher contrast-to-noise ratio and contrast resolution than the single-exposure acquisitions for the same x-ray dose. In addition, image artifacts were reduced in the combined dataset. This technique to extend the dynamic range of the detector without increasing the dose is particularly suited to image samples that contain both low and high attenuation regions.
引用
收藏
页码:421 / 439
页数:19
相关论文
共 50 条
  • [21] A Simple Technique for Signal Compression in High Dynamic Range, High Speed X-ray Pixel Detectors
    Fiorini, C.
    Nasri, B.
    Facchinetti, S.
    Bombelli, L.
    Fischer, P.
    Porro, M.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (05) : 2595 - 2600
  • [22] A Simple Technique for Signal Compression in High Dynamic Range, High Speed X-ray Pixel Detectors
    Fiorini, C.
    Nasri, B.
    Facchinetti, S.
    Bombelli, L.
    Fischer, P.
    Porro, M.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [23] THE OPTIMAL GAIN AND DYNAMIC RANGE OF CCD DETECTORS FOR X-RAY DIFFRACTION
    Durst, R. D.
    Mueller, P.
    Campana, C.
    Foundling, S.
    Jin, Z.
    Phillips, J.
    Pressprich, M.
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 1999, 55 : 239 - 239
  • [24] Effective dose reduction in dual-energy flat panel X-ray imaging:: Technique and clinical evaluation
    Avinash, GB
    Jabri, KN
    Uppaluri, R
    Rader, A
    Fischbach, F
    Ricke, J
    Teichgräber, U
    MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3, 2002, 4684 : 1048 - 1059
  • [25] Dual-Energy, Dual-Exposure PA and Lateral Chest Radiograph: Not Your Father's Chest X-Ray
    Boswell, Gilbert E.
    Wolfgramm, Sione T.
    Fong, Raynard K.
    Hawley, Daniel B.
    MILITARY MEDICINE, 2023, 188 (1-2) : 12 - 15
  • [26] Dependence of x-ray sensitivity of direct conversion x-ray detectors on x-ray exposure and exposure history
    Kabir, MZ
    Yunus, M
    Kasap, SO
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 170 - 176
  • [27] X-Ray Radiographic Imaging Technique with High Dynamic Range
    Liu Bin
    Wang Li-ming
    Su Xin-yan
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34 (04) : 1104 - 1108
  • [28] Photoconductor selection for digital flat panel x-ray image detectors based on the dark current
    Kasap, SO
    Rowlands, JA
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 2000, 18 (02): : 615 - 620
  • [29] Performance tests on a-Si TFT arrays for flat panel digital x-ray detectors
    Couture, A
    Albagli, D
    Possin, G
    Hudspeth, H
    Janiszewski, P
    Zoeller, M
    Granfors, P
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 243 - 250
  • [30] A nonlinear lag correction algorithm for a-Si flat-panel x-ray detectors
    Starman, Jared
    Star-Lack, Josh
    Virshup, Gary
    Shapiro, Edward
    Fahrig, Rebecca
    MEDICAL PHYSICS, 2012, 39 (10) : 6035 - 6047