Ex vivo human bile duct radiofrequency ablation with a bipolar catheter

被引:15
|
作者
Atar, Mustafa [1 ]
Kadayifci, Abdurrahman [1 ,2 ]
Daglilar, Ebubekir [1 ]
Hagen, Catherine [3 ,4 ]
Fernandez-del Castillo, Carlos [5 ]
Brugge, William R. [1 ]
机构
[1] Harvard Med Sch, Massachusetts Gen Hosp, Div Gastroenterol, 3-H GI Associates,Zero Emerson Pl,Blossom St, Boston, MA 02114 USA
[2] Reyap Istanbul Hosp, Unit Gastroenterol, Istanbul, Turkey
[3] Harvard Med Sch, Massachusetts Gen Hosp, Dept Pathol, Boston, MA USA
[4] Med Coll Wisconsin, Dept Pathol, Milwaukee, WI 53226 USA
[5] Harvard Med Sch, Massachusetts Gen Hosp, Dept Surg, Boston, MA USA
关键词
Radiofrequency ablation; Biliary tumors; Common bile duct; Pancreas; Bipolar catheter; MALIGNANT BILIARY OBSTRUCTION;
D O I
10.1007/s00464-017-5984-0
中图分类号
R61 [外科手术学];
学科分类号
摘要
Management of the primary and secondary tumors of the bile ducts still remains as a major clinical challenge. Radiofrequency (RF) ablation (RFA) of these tumors is feasible but the effect of RF energy on the human common bile duct (CBD) and surrounding tissues has not been investigated. This pilot study aimed to determine the relationship between RF energy and the depth of ablation in the normal human CBD. The study was performed on fresh ex vivo human biliary-pancreatic tissue which had been resected for a pancreatic cyst or mass. The study was conducted within 15 min after resection. A bipolar Habib RFA catheter was placed into the middle of the intact CBD, and three different (5, 7, 10 W) power settings were applied over a 90-s period by an RF generator. Gross and histological examinations were performed. The depth of coagulation necrosis in CBD and the effect of RFA on CBD wall and surrounding pancreas tissue were determined by microscopic examination. The study included eight tissue samples. 5 W power was applied to three sites and RFA caused only focal epithelial necrosis limited to the CBD mucosa. 7 and 10 W were applied to five sites and coagulation necrosis occurred in all cases. Microscopically, necrosis was transmural, involved accessory bile duct glands, and extended to the surrounding pancreatic tissue in four of these cases. Macroscopically, RFA resulted in circumferential white-yellowish color change extending approximately 2 cm of the CBD. Bipolar RF energy application with 5 W resulted in limited ablation on CBD wall. However, 7 and 10 W generated tissue necrosis which extended through the CBD wall and into surrounding pancreas tissue. Endoscopic biliary RFA is an effective technique for local biliary tissue ablation but the use of high energy may injure surrounding tissue.
引用
收藏
页码:2808 / 2813
页数:6
相关论文
共 50 条
  • [41] Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force-Sensing Catheter
    Thiagalingam, Aravinda
    D'Avila, Andre
    Foley, Lori
    Guerrero, J. Luis
    Lambert, Hendrik
    Leo, Giovanni
    Ruskin, Jeremy N.
    Reddy, Vivek Y.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (07) : 806 - 811
  • [42] Catheter contact area strongly correlates with lesion area in radiofrequency cardiac ablation: an ex vivo porcine heart study
    Kriengsak Masnok
    Nobuo Watanabe
    Journal of Interventional Cardiac Electrophysiology, 2022, 63 : 561 - 572
  • [43] Evaluation of effects of a novel endoscopically applied radiofrequency ablation biliary catheter using an ex-vivo pig liver
    Itoi, Takao
    Isayama, Hiroyuki
    Sofuni, Atsushi
    Itokawa, Fumihide
    Tamura, Miho
    Watanabe, Yusuke
    Moriyasu, Fuminori
    Kahaleh, Michel
    Habib, Nagy
    Nagao, Toshitaka
    Yokoyama, Tomohisa
    Kasuya, Kazuhiko
    Kawakami, Hiroshi
    JOURNAL OF HEPATO-BILIARY-PANCREATIC SCIENCES, 2012, 19 (05) : 543 - 547
  • [44] Catheter contact area strongly correlates with lesion area in radiofrequency cardiac ablation: an ex vivo porcine heart study
    Masnok, Kriengsak
    Watanabe, Nobuo
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022, 63 (03) : 561 - 572
  • [45] An ex-vivo experimental study on optimization of bipolar radiofrequency liver ablation using perfusion-cooled electrodes
    Lee, JM
    Han, JK
    Kim, SH
    Lee, JY
    Shin, KS
    Choi, BI
    ACTA RADIOLOGICA, 2005, 46 (05) : 443 - 451
  • [46] Juxtametallic Bipolar Bone Radiofrequency Ablation: Thermal Monitoring in an Ex-Vivo Model with Specimen MRI and Histopathologic Correlation
    Sag, Alan A.
    Sperduto, William A. L.
    Eward, William
    Ronald, James
    Davis, Haley
    Jiang, Xiaoyin Sara
    Enterline, David S.
    Visgauss, Julia
    Brigman, Brian
    Goodwin, C. Rory
    Qadri, Yawar J.
    Kim, Charles Y.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 33 (12) : 1594 - 1600
  • [47] Hepatic bipolar radiofrequency ablation using perfused-cooled electrodes:: a comparative study in the ex vivo bovine liver
    Lee, JM
    Han, JK
    Kim, SH
    Lee, JY
    Choi, SH
    Choi, BI
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (923): : 944 - 949
  • [48] Optimization of Direct Current–Enhanced Radiofrequency Ablation: An Ex Vivo Study
    Toshihiro Tanaka
    Peter Isfort
    Philipp Bruners
    Tobias Penzkofer
    Kimihiko Kichikawa
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2010, 33 : 1028 - 1032
  • [49] In vivo Lesion Index (LSI) validation in percutaneous radiofrequency catheter ablation
    Themistoclakis, Sakis
    Calzolari, Vittorio
    De Mattia, Luca
    China, Paolo
    Dello Russo, Antonio
    Fassini, Gaetano
    Casella, Michela
    Caporaso, Igor
    Indiani, Stefano
    Addis, Alessandro
    Basso, Cristina
    Della Barbera, Mila
    Thiene, Gaetano
    Tondo, Claudio
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (05) : 874 - 882
  • [50] Bipolar radiofrequency ablation of the kidney: Comparison with monopolar radiofrequency ablation
    Nakada, SY
    Jerde, TJ
    Warner, TF
    Wright, AS
    Haemmerich, D
    Mahvi, DM
    Lee, FT
    JOURNAL OF ENDOUROLOGY, 2003, 17 (10) : 927 - 933