Radiation dose calculations for CT scans with tube current modulation using the approach to equilibrium function

被引:10
|
作者
Li, Xinhua
Zhang, Da
Liu, Bob [1 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Div Diagnost Imaging Phys, Boston, MA 02114 USA
关键词
computed tomography (CT); dosimetry; tube current modulation; DOSIMETRY; EQUATIONS;
D O I
10.1118/1.4898124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The approach to equilibrium function has been used previously to calculate the radiation dose to a shift-invariant medium undergoing CT scans with constant tube current [Li, Zhang, and Liu, Med. Phys. 39, 5347-5352 (2012)]. The authors have adapted this method to CT scans with tube current modulation (TCM). Methods: For a scan with variable tube current, the scan range was divided into multiple subscan ranges, each with a nearly constant tube current. Then the dose calculation algorithm presented previously was applied. For a clinical CT scan series that presented tube current per slice, the authors adopted an efficient approach that computed the longitudinal dose distribution for one scan length equal to the slice thickness, which center was at z = 0. The cumulative dose at a specific point was a summation of the contributions from all slices and the overscan. Results: The dose calculations performed for a total of four constant and variable tube current distributions agreed with the published results of Dixon and Boone [Med. Phys. 40, 111920 (14pp.) (2013)]. For an abdomen/pelvis scan of an anthropomorphic phantom (model ATOM 701-B, CIRS, Inc., VA) on a GE Lightspeed Pro 16 scanner with 120 kV, N x T = 20 mm, pitch = 1.375, z axis current modulation (auto mA), and angular current modulation (smart mA), dose measurements were performed using two lines of optically stimulated luminescence dosimeters, one of which was placed near the phantom center and the other on the surface. Dose calculations were performed on the central and peripheral axes of a cylinder containing water, whose cross-sectional mass was about equal to that of the ATOM phantom in its abdominal region, and the results agreed with the measurements within 28.4%. Conclusions: The described method provides an effective approach that takes into account subject size, scan length, and constant or variable tube current to evaluate CT dose to a shift-invariant medium. For a clinical CT scan, dose calculations may be performed with a water-containing cylinder whose cross-sectional mass is equal to that of the subject. This method has the potential to substantially improve evaluations of patient dose from clinical CT scans, compared to CTDIvol, size-specific dose estimate (SSDE), or the dose evaluated for a TCM scan with a constant tube current equal to the average tube current of the TCM scan. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM)
    Layman, Rick R.
    Hardy, Anthony J.
    Kim, Hyun J.
    Chou, Ei Ne
    Bostani, Maryam
    Cagnon, Chris
    Cody, Dianna
    McNitt-Gray, Michael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (05): : 97 - 109
  • [32] MDCT of the coronary arteries: Feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose
    Abada, Hicham T.
    Larchez, Christophe
    Daoud, Beatrice
    Sigal-Cinqualbre, Anne
    Paul, Jean-Francois
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) : S387 - S390
  • [33] Eye Lens Dose Reduction From CT Scan Using Organ Based Tube Current Modulation
    Liu, H.
    Liu, T.
    Xu, X.
    Wu, J.
    Zhuo, W.
    MEDICAL PHYSICS, 2015, 42 (06) : 3250 - 3250
  • [34] ORGAN DOSE EVALUATIONS BASED ON MONTE CARLO SIMULATION FOR CT EXAMINATIONS USING TUBE CURRENT MODULATION
    Fujii, K.
    Nomura, K.
    Muramatsu, Y.
    Obara, S.
    Akahane, K.
    Kusumoto, M.
    RADIATION PROTECTION DOSIMETRY, 2017, 174 (03) : 387 - 394
  • [35] Dose reduction in CT examinations of children by anatomically adapted tube current modulation
    Greess, H
    Noemayr, A
    Wolf, H
    Janka, R
    Kalender, WA
    Bautz, WA
    RADIOLOGY, 1999, 213P : 294 - 294
  • [36] Head CT dose reduction with organ-based tube current modulation
    Ota, Joji
    Yokota, Hajime
    Kobayashi, Takenori
    Ogata, Yuki
    Kubo, Takumi
    Chida, Koichi
    Masuda, Yoshitada
    Uno, Takashi
    MEDICAL PHYSICS, 2022, 49 (03) : 1964 - 1971
  • [37] Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6309 - 6326
  • [38] Add Tube Current Modulation to a Low Dose Simulation Tool for CT Systems
    Ding, Y.
    Wen, G.
    Brown, K.
    Klahr, P.
    Dhanantwari, A.
    MEDICAL PHYSICS, 2015, 42 (06) : 3621 - 3621
  • [39] Dose conversion coefficients for CT examinations of adults with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20): : 6243 - 6261
  • [40] The impact on CT dose of the variability in tube current modulation technology: a theoretical investigation
    Li, Xiang
    Segars, W. Paul
    Samei, Ehsan
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (16): : 4525 - 4548