Radiation dose calculations for CT scans with tube current modulation using the approach to equilibrium function

被引:10
|
作者
Li, Xinhua
Zhang, Da
Liu, Bob [1 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Div Diagnost Imaging Phys, Boston, MA 02114 USA
关键词
computed tomography (CT); dosimetry; tube current modulation; DOSIMETRY; EQUATIONS;
D O I
10.1118/1.4898124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The approach to equilibrium function has been used previously to calculate the radiation dose to a shift-invariant medium undergoing CT scans with constant tube current [Li, Zhang, and Liu, Med. Phys. 39, 5347-5352 (2012)]. The authors have adapted this method to CT scans with tube current modulation (TCM). Methods: For a scan with variable tube current, the scan range was divided into multiple subscan ranges, each with a nearly constant tube current. Then the dose calculation algorithm presented previously was applied. For a clinical CT scan series that presented tube current per slice, the authors adopted an efficient approach that computed the longitudinal dose distribution for one scan length equal to the slice thickness, which center was at z = 0. The cumulative dose at a specific point was a summation of the contributions from all slices and the overscan. Results: The dose calculations performed for a total of four constant and variable tube current distributions agreed with the published results of Dixon and Boone [Med. Phys. 40, 111920 (14pp.) (2013)]. For an abdomen/pelvis scan of an anthropomorphic phantom (model ATOM 701-B, CIRS, Inc., VA) on a GE Lightspeed Pro 16 scanner with 120 kV, N x T = 20 mm, pitch = 1.375, z axis current modulation (auto mA), and angular current modulation (smart mA), dose measurements were performed using two lines of optically stimulated luminescence dosimeters, one of which was placed near the phantom center and the other on the surface. Dose calculations were performed on the central and peripheral axes of a cylinder containing water, whose cross-sectional mass was about equal to that of the ATOM phantom in its abdominal region, and the results agreed with the measurements within 28.4%. Conclusions: The described method provides an effective approach that takes into account subject size, scan length, and constant or variable tube current to evaluate CT dose to a shift-invariant medium. For a clinical CT scan, dose calculations may be performed with a water-containing cylinder whose cross-sectional mass is equal to that of the subject. This method has the potential to substantially improve evaluations of patient dose from clinical CT scans, compared to CTDIvol, size-specific dose estimate (SSDE), or the dose evaluated for a TCM scan with a constant tube current equal to the average tube current of the TCM scan. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Examining the Effects Tube Current Modulation On Fetal Dose From Abdomen/Pelvis CT Scans
    Hardy, A.
    Bostani, M.
    Angel, F.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2018, 45 (06) : E691 - E691
  • [2] Pediatric cardiovascular CT angiography: Radiation dose reduction using automatic anatomic tube current modulation
    Herzog, Christopher
    Mulvihill, Denise M.
    Nguyen, Shaun A.
    Savino, Giancarlo
    Schmidt, Bernhard
    Costello, Philip
    Vogl, Thomas J.
    Schoepf, U. Joseph
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (05) : 1232 - 1240
  • [3] Estimating Organ Dose From CT Scans Performed with Tube Current Modulated Scans
    Khatonabadi, M.
    Chu, G.
    Oria, D.
    Cagnon, C.
    DeMarco, J.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [4] Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study
    Sabarudin, Akmal
    Mustafa, Zakira
    Nassir, Khadijah Mohd
    Hamid, Hamzaini Abdul
    Sun, Zhonghua
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 319 - 328
  • [5] Estimating Radiation Dose From Low-Dose Lung Cancer Screening CT Exams Using Tube Current Modulation
    Hardy, A.
    Bostani, M.
    McMillan, K.
    Zankl, M.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2016, 43 (06) : 3773 - 3773
  • [6] Potential for Radiation Dose Savings in Abdominal and Chest CT Using Automatic Tube Voltage Selection in Combination With Automatic Tube Current Modulation
    Mayer, Caroline
    Meyer, Mathias
    Fink, Christian
    Schmidt, Bernhard
    Sedlmair, Martin
    Schoenberg, Stefan O.
    Henzler, Thomas
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (02) : 292 - 299
  • [7] Quantitative Effects of Clinical Practice On the Efficiency of CT Tube Current Modulation for Radiation Dose Reduction
    Zhang, J.
    Cornett, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3635 - 3635
  • [8] Calculation of dose reduction by tube current modulation in pediatric CT
    Schmidt, B
    Greess, H
    Wolf, H
    Kalender, WA
    RADIOLOGY, 2001, 221 : 365 - 366
  • [9] Patient Specific Tube Current Modulation for CT Dose Reduction
    Jin, Yannan
    Yin, Zhye
    Yao, Yangyang
    Wang, Hui
    Wu, Mingye
    Kalra, Mannudeep
    De Man, Bruno
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [10] Absorption dose reduction by tube current modulation in CT system
    Nagaoka, T
    Miyazaki, O
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 719 - 731