Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation

被引:44
|
作者
Ding, Xiaorong [1 ]
Zhang, Yuanting [1 ]
Tsang, Hon Ki [1 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Shatin, Hong Kong, Peoples R China
关键词
heart disease; calibration interval; pulse transit time; blood pressure; estimation accuracy; FINGER PHOTOPLETHYSMOGRAM; ARRIVAL-TIME; VALIDATION; HYPERTENSION; DEVICES;
D O I
10.1088/0967-3334/37/2/227
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Continuous blood pressure (BP) measurement without a cuff is advantageous for the early detection and prevention of hypertension. The pulse transit time (PTT) method has proven to be promising for continuous cuffless BP measurement. However, the problem of accuracy is one of the most challenging aspects before the large-scale clinical application of this method. Since PTT-based BP estimation relies primarily on the relationship between PTT and BP under certain assumptions, estimation accuracy will be affected by cardiovascular disorders that impair this relationship and by the calibration frequency, which may violate these assumptions. This study sought to examine the impact of heart disease and the calibration interval on the accuracy of PTT-based BP estimation. The accuracy of a PTT-BP algorithm was investigated in 37 healthy subjects and 48 patients with heart disease at different calibration intervals, namely 15 min, 2 weeks, and 1 month after initial calibration. The results showed that the overall accuracy of systolic BP estimation was significantly lower in subjects with heart disease than in healthy subjects, but diastolic BP estimation was more accurate in patients than in healthy subjects. The accuracy of systolic and diastolic BP estimation becomes less reliable with longer calibration intervals. These findings demonstrate that both heart disease and the calibration interval can influence the accuracy of PTT-based BP estimation and should be taken into consideration to improve estimation accuracy.
引用
收藏
页码:227 / 237
页数:11
相关论文
共 50 条
  • [1] Pulse Transit Time-based Blood Pressure Estimation Using Hilbert-Huang Transform
    Zhang, Qiao
    Shi, Yang
    Teng, Daniel
    Dinh, Anh
    Ko, Seok-Bum
    Chen, Li
    Basran, Jenny
    Dal Bello-Haas, Vanina
    Choi, Younhee
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 1785 - +
  • [2] VALIDATION OF THE FREESCAN PULSE TRANSIT TIME-BASED BLOOD PRESSURE MONITOR
    Wu, Chih-Cheng
    Chao, Paul C. -P.
    JOURNAL OF HYPERTENSION, 2016, 34 : E142 - E142
  • [3] Blood Pressure Estimation Based on Pulse Transit Time with LANDMARC
    Wang, Yang
    Liu, Zhiwen
    2017 IEEE LIFE SCIENCES CONFERENCE (LSC), 2017, : 141 - 144
  • [4] Validation of Two Pulse Transit Time-Based Blood Pressure Measurement Devices
    Bard, Dylan M.
    Hahnen, Christina
    Joseph, Jeffrey I.
    van Helmond, Noud
    FASEB JOURNAL, 2019, 33
  • [5] Perturbationless Calibration of Pulse Transit Time to Blood Pressure
    Gao, Mingwu
    Mukkamala, Ramakrishna
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 232 - 235
  • [6] Validation of Pulse Transit Time-Based Blood pressure measurement in heart failure Patients with non-invasive Ventilation
    Spiesshoefer, J.
    Heinrich, J.
    Eckert, S.
    Bitter, T.
    Horstkotte, D.
    Oldenburg, O.
    INTERNIST, 2013, 54 : 106 - 106
  • [7] Cuff-Less High-Accuracy Calibration-Free Blood Pressure Estimation Using Pulse Transit Time
    Kachuee, Mohamad
    Kiani, Mohammad Mahdi
    Mohammadzade, Hoda
    Shabany, Mahdi
    2015 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2015, : 1006 - 1009
  • [8] Evaluation of transit time-based models in wearable central aortic blood pressure estimation
    Fierro, German
    Armentano, Ricardo
    Silveira, Fernando
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (03)
  • [9] A Novel Calibration Procedure of Pulse Transit Time based Blood Pressure measurement with Heart Rate and Respiratory Rate
    Lui, Hin-Wai
    Chow, King-Lau
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 4318 - 4322
  • [10] Noninvasive Continuous Blood Pressure Estimation From Pulse Transit Time: A Review of the Calibration Models
    Barvik, Daniel
    Cerny, Martin
    Penhaker, Marek
    Noury, Norbert
    IEEE REVIEWS IN BIOMEDICAL ENGINEERING, 2022, 15 : 138 - 151