The Effect of Repetitive Pilot-Hole Use on the Insertion Torque and Pullout Strength of Vertebral System Screws

被引:29
|
作者
Defino, Helton L. A. [1 ]
Rosa, Rodrigo Cesar [1 ]
Silva, Patricia [1 ]
Shimano, Antonio Carlos [1 ]
Volpon, Jose Batista [1 ]
Albuquerque de Paula, Francisco Jose [2 ]
Schleicher, Philipp [3 ]
Schnake, Klaus [3 ]
Kandziora, Frank [3 ]
机构
[1] Univ Sao Paulo, Dept Biomech Med & Rehabil Locomotor Apparatus, Sao Paulo, Brazil
[2] Univ Sao Paulo, Dept Internal Med, Sao Paulo, Brazil
[3] Charite Univ Med Berlin, Ctr Musculoskeletal Surg, Berlin, Germany
关键词
spine; bone screws; biomechanics; insertion torque; retentive strength; PEDICLE SCREWS; SPINAL INSTRUMENTATION; LUMBOSACRAL SPINE; LUMBAR SPINE; OUT STRENGTH; FIXATION; TESTS;
D O I
10.1097/BRS.0b013e31819e3556
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vitro biomechanical investigation of the screw-holding capacity. Objective. To evaluate the effect of repetitive screw-hole use on the insertional torque and retentive strength of vertebral system screws. Summary and Background Data. Placement and removal of vertebral system screws is sometimes necessary during the surgical procedures in order to assess the walls of the pilot hole. This procedure may compromise the holding capacity of the implant. Methods. Screws with outer diameter measuring 5, 6, and 7 mm were inserted into wood, polyurethane, polyethylene, and cancellous bone cylindrical blocks. The pilot holes were made with drills of a smaller, equal, or wider diameter than the inner screw diameter. Three experimental groups were established based on the number of insertions and reinsertions of the screws and subgroups were created according to the outer diameter of the screw and the diameter of the pilot hole used. Results. A reduction of screw-holding capacity was observed between the first and the following insertions regardless the anchorage material. The pattern of reduction of retentive strength was not similar to the pattern of torque reduction. The pullout strength was more pronounced between the first and the last insertions, while the torque decreased more proportionally from the first to the last insertions. Conclusion. Insertion and reinsertion of the screws of the vertebral fixation system used in the present study reduced the insertion torque and screw purchase.
引用
收藏
页码:871 / 876
页数:6
相关论文
共 21 条
  • [1] Effect of pilot hole diameter and tapping on insertion torque and axial pullout strength of 4.0-mm cancellous bone screws
    Kunkel, Kevin A. R.
    Suber, Jonathan T.
    Gerard, Patrick D.
    Kowaleski, Michael P.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2011, 72 (12) : 1660 - 1665
  • [2] The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone
    Battula, Suneel
    Schoenfeld, Andrew J.
    Sahai, Vivek
    Vrabec, Gregory A.
    Tank, Jason
    Njus, Glen O.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04): : 990 - 995
  • [3] CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS
    DAFTARI, TK
    HORTON, WC
    HUTTON, WC
    JOURNAL OF SPINAL DISORDERS, 1994, 7 (02): : 139 - 145
  • [4] Effect of pilot-hole size on the pullout strength of flexor digitorum longus transfer fixed with a bioabsorbable screw
    Donley, Brian G.
    Jambor, Clinton
    Erdermier, Ahmet
    Sferra, James
    Cavanagh, Peter
    FOOT & ANKLE INTERNATIONAL, 2007, 28 (10) : 1078 - 1081
  • [5] Insertion torque and pullout force of rescue screws for anterior cervical plate fixation in a fatigued initial pilot hole
    Pitzen, T
    Franta, F
    Barbier, D
    Steudel, WI
    JOURNAL OF NEUROSURGERY-SPINE, 2004, 1 (02) : 198 - 201
  • [6] Divergent Configuration Improves Insertion Torque and Pullout Strength of Anterior Cervical Screws
    Xavier, Fred
    Saha, Subrata
    2016 32ND SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE (SBEC), 2016, : 98 - 98
  • [7] Effect of insertion torque on bone screw pullout strength
    Lawson, KJ
    Brems, J
    ORTHOPEDICS, 2001, 24 (05) : 451 - 454
  • [8] Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws
    Heidemann, W
    Gerlach, KL
    Grobel, KH
    Kollner, HG
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 1998, 26 (01) : 50 - 55
  • [9] The effect of insertion technique and surgeon experience on the pullout strength of orthopaedic screws
    Meals, Clifton
    Fox, Bradley
    Zhang, Lijie Grace
    Neviaser, Robert J.
    CURRENT ORTHOPAEDIC PRACTICE, 2016, 27 (01): : 69 - 71
  • [10] How pilot-hole size affects bone-screw pullout strength in human cadaveric cancellous bone
    Steeves, M
    Stone, C
    Mogaard, J
    Byrne, S
    CANADIAN JOURNAL OF SURGERY, 2005, 48 (03) : 207 - 212