Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws

被引:44
|
作者
Heidemann, W
Gerlach, KL
Grobel, KH
Kollner, HG
机构
[1] Otto Von Guericke Univ, Dept Maxillofacial Surg, D-39120 Magdeburg, Germany
[2] Otto Von Guericke Univ, Inst Mech, D-39120 Magdeburg, Germany
关键词
D O I
10.1016/S1010-5182(98)80035-8
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
When screws are inserted in thick cortical bone, a small pilot hole size, corresponding to the care diameter of the screw, can result in high torsional stress, leading to screw fracture, The aim of this study was to enlarge the drill size up to a critical pilot hale size (CPHS) which, if exceeded, means a rapid decrease in the screw holding power, 1.5 and 2 mm titanium screws were inserted in discs of polyvinylchloride (PVC), wood and porcine mandibular bone with thicknesses differing between 2 to 4 mm, using an increasing pilot hole size between 66% and 95% of the screw external diameter, Torque measurements and pullout tests were performed and the CPHS was calculated, In torque measurements, the CPHS of microscrews ranged between 83% and 85% of the screw external diameter (SED). The CPHS of miniscrews lay between 80% and 90% of SED. In pullout analysis, the CPHS of microscrews ranged between 83% and 89% of SED; the CPHS of miniscrews lay between 79% and 91% of SED, The mean of the CPHS was calculated to be approximately 85% of the SED, Up to this critical point, the pilot hole size may be increased without affecting the holding power of the screws.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 41 条
  • [1] The Effect of Repetitive Pilot-Hole Use on the Insertion Torque and Pullout Strength of Vertebral System Screws
    Defino, Helton L. A.
    Rosa, Rodrigo Cesar
    Silva, Patricia
    Shimano, Antonio Carlos
    Volpon, Jose Batista
    Albuquerque de Paula, Francisco Jose
    Schleicher, Philipp
    Schnake, Klaus
    Kandziora, Frank
    SPINE, 2009, 34 (09) : 871 - 876
  • [2] Insertion torque and pullout force of rescue screws for anterior cervical plate fixation in a fatigued initial pilot hole
    Pitzen, T
    Franta, F
    Barbier, D
    Steudel, WI
    JOURNAL OF NEUROSURGERY-SPINE, 2004, 1 (02) : 198 - 201
  • [3] CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS
    DAFTARI, TK
    HORTON, WC
    HUTTON, WC
    JOURNAL OF SPINAL DISORDERS, 1994, 7 (02): : 139 - 145
  • [4] Effect of pilot hole diameter and tapping on insertion torque and axial pullout strength of 4.0-mm cancellous bone screws
    Kunkel, Kevin A. R.
    Suber, Jonathan T.
    Gerard, Patrick D.
    Kowaleski, Michael P.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2011, 72 (12) : 1660 - 1665
  • [5] The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone
    Battula, Suneel
    Schoenfeld, Andrew J.
    Sahai, Vivek
    Vrabec, Gregory A.
    Tank, Jason
    Njus, Glen O.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04): : 990 - 995
  • [6] Auswirkung verschiedener Bohrlochdurchmesser auf die Haltekraft von OsteosyntheseschraubenThe effect of various pilot hole sizes on the holding power of osteosynthesis screws
    W. Heidemann
    K. L. Gerlach
    K. -H. Gröbel
    H. -G. Köllner
    Mund-, Kiefer- und Gesichtschirurgie, 1998, 2 (3): : 136 - 140
  • [7] GEOMETRICAL ANALYSIS OF HOLE SIZES FOR THREAD FORMING SCREWS
    LAWSER, JJ
    SAE TRANSACTIONS, 1969, 78 : 75 - &
  • [8] Effect of different radial hole designs on pullout and structural strength of cannulated pedicle screws
    Chen, Hsin-Chang
    Lai, Yu Shu
    Chen, Wen-Chuan
    Chen, Jou-Wen
    Chang, Chia-Ming
    Chen, Yi-Long
    Wang, Shih-Tien
    Cheng, Cheng-Kung
    MEDICAL ENGINEERING & PHYSICS, 2015, 37 (08) : 746 - 751
  • [9] TORQUE AND PULLOUT ANALYSIS OF 6 CURRENTLY AVAILABLE SELF-TAPPING AND EMERGENCY SCREWS
    BOYLE, JM
    FROST, DE
    FOLEY, WL
    GRADY, JJ
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1993, 51 (01) : 45 - 50
  • [10] Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: Influences of radial hole and pilot hole tapping
    Chen, Lih-Huei
    Tai, Ching-Lung
    Lai, Po-Liang
    Lee, De-Mei
    Tsai, Tsung-Tin
    Fu, Tsai-Sheng
    Niu, Chi-Chien
    Chen, Wen-Jer
    CLINICAL BIOMECHANICS, 2009, 24 (08) : 613 - 618