Evaluation of the ACR MRI phantom for quality assurance tests of 1.5 T MRI scanners in Estonian hospitals

被引:4
|
作者
Kaljuste, Doris [1 ]
Nigul, Mait [2 ]
机构
[1] Tallinn Univ Technol, Dept Biomed Engn, EE-19086 Tallinn, Estonia
[2] East Tallinn Cent Hosp, EE-10138 Tallinn, Estonia
关键词
magnetic resonance imaging; ACR MRI phantom; quality assurance; MRI image quality;
D O I
10.3176/proc.2014.3.06
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Since its discovery, magnetic resonance imaging (MRI) has been one of the main methods of imaging in radiology. So far, there are not many national or international guidelines for MRI quality assurance compared to imaging methods that are using ionizing radiation. American College of Radiology (ACR) has an accreditation program that includes a standardized image quality measurement protocol and a phantom. Seven important assessments of MRI image quality are included with acceptance criteria. The aim of this study was to determine whether using the ACR MRI phantom would be a suitable quality assurance method for MRI systems in Estonia. In order to determine if the ACR MRI phantom would be a suitable quality assurance method for MRI systems in Estonia, six MRI scanners were tested using the ACR MRI accreditation program phantom and method. Information about the image quality was obtained and 33% of the MRI scanners passed all seven assessments. In conclusion, the ACR MRI phantom is a suitable test body for some of the MRI scanners in Estonia and should be considered to be used for regular quality testing. Due to the fact that in some cases the ACR MRI phantom did not fit into the head coil, using the ACR MRI phantom cannot be nationally required and other quality assurance phantoms should also be assessed. Adding signal-to-noise ratio assessment to seven assessment criteria, provided by ACR quality assurance manual, and developing an automated ACR quality assurance procedure would make quality assurance with ACR MRI phantom more beneficial.
引用
收藏
页码:240 / 246
页数:7
相关论文
共 50 条
  • [21] Evaluation of MRI Acquisition Parameter Variations On Texture Feature Extraction Using ACR Phantom
    Xie, Y.
    Wang, J.
    Wang, C.
    Chang, Z.
    MEDICAL PHYSICS, 2016, 43 (06) : 3380 - 3380
  • [22] MR Simulator Quality Assurance Technique Using an ACR MRI Phantom Based on 3D MR Sequence to Accommodate Radiotherapy Applications
    Ho, C.
    Wong, O.
    Yuan, J.
    Cheung, K.
    Yu, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E544 - E544
  • [23] Image Quality Verification for Off-Center Variation Using ACR Phantom in Open MRI
    Hong, C. P.
    Lee, D. H.
    Han, B.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [24] Evaluating the effect of dielectric resonance on image uniformity at 1.5T and 3T in an ACR MRI phantom filled with mineral oil and silicone fluid
    Hatab, M.
    Goerner, F.
    Clarke, G.
    MEDICAL PHYSICS, 2007, 34 (06) : 2356 - 2356
  • [25] QUALITY ASSURANCE IN MRI ASSISTED TREATMENT PLANNING - COMPARISON OF IMAGING ACCURACY AT DIFFERENT MRI UNITS BASED ON PHANTOM MEASUREMENTS
    PROTT, FJ
    POTTER, R
    HAVERKAMP, U
    STOBER, U
    LENZEN, B
    WENDTNER, CM
    RADEMACHER, C
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 302 - 302
  • [26] Modification of the ACR Phantom and Analysis Procedures for 7T Clinical MRI Scantier Testing
    Lanners, D.
    Trester, P.
    Shur, Y.
    Ban, Z.
    Felmlee, J.
    Fagan, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E439 - E439
  • [27] A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing
    Singhrao, Kamal
    Fu, Jie
    Wu, Holden H.
    Hu, Peng
    Kishan, Amar U.
    Chin, Robert K.
    Lewis, John H.
    MEDICAL PHYSICS, 2020, 47 (04) : 1443 - 1451
  • [28] Cardiac Imaging Using Clinical 1.5 T MRI Scanners in a Murine Ischemia/Reperfusion Model
    Voelkl, Jakob G. J.
    Haubner, Bernhard J.
    Kremser, Christian
    Mayr, Agnes
    Klug, Gert
    Loizides, Alexander
    Mueller, Silvana
    Pachinger, Otmar
    Schocke, Michael
    Metzler, Bernhard
    JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2011,
  • [29] Preliminary study for a water-paraffin based phantom in MRI quality assurance test
    Lupi, Giulia
    Scorza, Andrea
    Rugiano, Maria Livia
    Sciuto, Salvatore Andrea
    Bini, Fabiano
    2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS, 2015, : 308 - 313
  • [30] Induced magnetic moment in stainless steel components of orthodontic appliances in 1.5 T MRI scanners
    Wang, Zhiyue J.
    Rollins, Nancy K.
    Liang, Hui
    Park, Yong Jong
    MEDICAL PHYSICS, 2015, 42 (10) : 5871 - 5878