Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation

被引:88
|
作者
Greess, H
Lutze, J
Nömayr, A
Wolf, H
Hothorn, T
Kalender, WA
Bautz, W
机构
[1] Univ Erlangen Nurnberg, Inst Radiol, D-91054 Erlangen, Germany
[2] Siemens AG Med Solut, Forchheim, Germany
[3] Univ Erlangen Nurnberg, Dept Med Informat Biometry & Epidemiol, Erlangen, Germany
[4] Univ Erlangen Nurnberg, Inst Med Phys, Erlangen, Germany
关键词
multislice spiral CT; mA s reduction; children; image quality; exposure to patients;
D O I
10.1007/s00330-004-2301-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The potential of online tube current modulation in subsecond multislice spiral CT (MSCT) examinations of children to reduce the dose without a loss in image quality is investigated in a controlled patient study. The dose can be reduced for oval patient sectional view without an increase in noise if the tube current is reduced where the patient diameter and, consequently, attenuation are small. We investigated a product version of an online control for tube current in a SOMATOM Sensation 4 (Siemens, Forchheim). We evaluated image quality, noise and dose reduction for examinations with online tube current modulation in 30 MSCT of thorax/abdomen and abdomen and compared mA s for tube current modulation to the mA s in standard weight-adapted children protocols. Image quality was rated as "very good," "good," "diagnostic" and "poor" in a consensus by three radiologists. Noise was assessed in comparison to 24 MSCT examinations without tube current modulation measured as SD in ROIs. The dose was reduced from 26 to 43% (mean 36%), depending on the patient's geometry and weight. In general, no loss of image quality was observed. Measured noise showed a decrease up to 26% and an increase up to 36%, although there was no decrease of image quality. Online tube current modulation is now used as a standard in MSCT at our institution. Dose in MSCT examinations of children can be reduced substantially in routine examinations by online tube current modulation without a loss of image quality.
引用
收藏
页码:995 / 999
页数:5
相关论文
共 50 条
  • [41] Evaluating Effective Doses and Image Quality for Multislice CT Using Automatic Tube Current Modulation
    Tsai, S. F.
    Tsai, H. Y.
    Tyan, Y. S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [42] Influence of CT automatic tube current modulation on uncertainty in effective dose
    Sookpeng, S.
    Martin, C. J.
    Gentle, D. J.
    RADIATION PROTECTION DOSIMETRY, 2016, 168 (01) : 46 - 54
  • [43] Prospective estimation of organ dose in CT under tube current modulation
    Tian, Xiaoyu
    Li, Xiang
    Segars, W. Paul
    Frush, Donald P.
    Samei, Ehsan
    MEDICAL PHYSICS, 2015, 42 (04) : 1575 - 1585
  • [44] Implementation of Tube Current Modulation into An Organ Dose Calculator for CT Patients
    Lee, C.
    Yeom, Y.
    Atkinson, J.
    Folio, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E264 - E264
  • [45] Patient Dose Monitoring in CT with Tube Current Modulation and Multiple Series
    Boone, J.
    Dixon, R.
    Li, B.
    Zhang, D.
    Li, X.
    MEDICAL PHYSICS, 2019, 46 (06) : E374 - E374
  • [46] Dose reduction in pediatric spiral CT
    Rogalla, P
    Stoever, B
    Juran, R
    Hamm, BK
    RADIOLOGY, 1996, 201 : 1605 - 1605
  • [47] Strategies for reduction of radiation dose in cardiac multislice CT
    Jean-François Paul
    Hicham T. Abada
    European Radiology, 2007, 17 : 2028 - 2037
  • [48] Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study
    Sabarudin, Akmal
    Mustafa, Zakira
    Nassir, Khadijah Mohd
    Hamid, Hamzaini Abdul
    Sun, Zhonghua
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 319 - 328
  • [49] Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CT
    Leswick, David A.
    Hunt, Megan M.
    Webster, Steven T.
    Fladeland, Derek A.
    RADIOLOGY, 2008, 249 (02) : 572 - 580
  • [50] Strategies for reduction of radiation dose in cardiac multislice CT
    Paul, Jean-Francois
    Abada, Hicham T.
    EUROPEAN RADIOLOGY, 2007, 17 (08) : 2028 - 2037