Failure analysis of the tibial component baseplate after total knee arthroplasty

被引:4
|
作者
Oliveira, A. L. L. [1 ]
Cueva, E. G. [2 ]
Carvalho, R. T. [3 ]
机构
[1] Univ Sao Paulo, Dept Mech Engn, BR-2231 Sao Paulo, Brazil
[2] Univ Brasilia, UnB Gama Coll, Area Especial, Brasilia, DF, Brazil
[3] Orthopaed & Traumatol Hosp Publ Workers State Sao, Sao Paulo, Brazil
关键词
Total knee arthroplasty; Baseplate; Failure analysis; Fracture; Prostheses design; REPLACEMENT; FRACTURE; WEAR; TRAY; UHMWPE; DESIGN; JOINTS; PERFORMANCE; PROSTHESIS; WALKING;
D O I
10.1016/j.engfailanal.2013.10.012
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The present work was performed on a retrieved modular prosthetic tibial component, which presented a fracture in the medial portion of the baseplate. In order to understand the causes that led to the failure of the baseplate, the steel, the mechanical and structural characteristics and the fracture were analyzed and correlated with numerical studies related to the biomechanical loads and displacements transmitted to the knee joint during human gait. The results showed a set of adverse events that occurred before the fracture of the baseplate. The steel, used for the manufacture of tibial component, had a high ferrite content, low mechanical strength, and did not comply with standard specifications. Design errors of the baseplate and insert assembly were shown to be the cause for the damage to the polyethylene of the insert. This damage included wear, plastic deformation and plaque formation by delamination, which increased the gap between the fitting guides. These phenomena enabled the relative displacement between the components, and the misaligning of the mechanical axis, causing an increase of the bending moment in the medial region of the baseplate. The load peaks concentrated in the primary contact area with the medial femoral condyle exceeded the mechanical strength limits of the piece, causing the fracture of the metallic tibial baseplate. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 50 条
  • [41] Effect of tibial component varus on wear in total knee arthroplasty
    Srivastava, Ajay
    Lee, Gregory Y.
    Steklov, Nikolai
    Colwell, Clifford W., Jr.
    Ezzet, Kace A.
    D'Lima, Darryl D.
    KNEE, 2012, 19 (05): : 560 - 563
  • [42] Obliquity of tibial component after unicompartmental knee arthroplasty
    Asada, Shigeki
    Inoue, Shinji
    Tsukamoto, Ichiro
    Mori, Shigeshi
    Akagi, Masao
    KNEE, 2019, 26 (02): : 410 - 415
  • [43] Varus/valgus alignment of the tibial component in total knee arthroplasty
    Jeffcote, B
    Shakespeare, D
    KNEE, 2003, 10 (03): : 243 - 247
  • [44] Relationship between Tibial Baseplate Design and Rotational Alignment Landmarks in Primary Total Knee Arthroplasty
    Indelli, Pier Francesco
    Graceffa, Angelo
    Baldini, Andrea
    Payne, Brielle
    Pipino, Gennaro
    Marcucci, Massimiliano
    ARTHRITIS, 2015,
  • [45] Total knee arthroplasty after high tibial osteotomy
    Meding, JB
    Keating, EM
    Ritter, MA
    Faris, PM
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2000, (375) : 175 - 184
  • [46] Total Knee Arthroplasty After High Tibial Osteotomy
    Cerciello, Simone
    Vasso, Michele
    Maffulli, Nicola
    Neyret, Philippe
    Corona, Katia
    Panni, Alfredo Schiavone
    ORTHOPEDICS, 2014, 37 (03) : 191 - 198
  • [47] Fracture of the tibial tray after total knee arthroplasty
    Altintas, F
    Sener, N
    Ugutmen, E
    JOURNAL OF ARTHROPLASTY, 1999, 14 (01): : 112 - 114
  • [48] TOTAL KNEE ARTHROPLASTY AFTER HIGH TIBIAL OSTEOTOMY
    HOFMANN, AA
    KANE, KR
    ORTHOPEDICS, 1994, 17 (09) : 887 - 890
  • [49] A proposed new rotating reference axis for the tibial component after proximal tibial resection in total knee arthroplasty
    Ohmori, Takaaki
    Kabata, Tamon
    Kajino, Yoshitomo
    Inoue, Daisuke
    Taga, Tadashi
    Yamamoto, Takashi
    Takagi, Tomoharu
    Yoshitani, Junya
    Ueno, Takuro
    Ueoka, Ken
    Tsuchiya, Hiroyuki
    PLOS ONE, 2018, 13 (12):
  • [50] Finite element analysis of the tibial component alignment in a transverse plane in total knee arthroplasty
    Popescu, Roman
    Cristea, Stefan
    Oleksik, Valentin
    Pascu, Adrian Marius
    Haritinian, Emil George
    JOURNAL OF APPLIED BIOMEDICINE, 2021, 19 (04) : 234 - 239