Failure analysis of the tibial component baseplate after total knee arthroplasty

被引:4
|
作者
Oliveira, A. L. L. [1 ]
Cueva, E. G. [2 ]
Carvalho, R. T. [3 ]
机构
[1] Univ Sao Paulo, Dept Mech Engn, BR-2231 Sao Paulo, Brazil
[2] Univ Brasilia, UnB Gama Coll, Area Especial, Brasilia, DF, Brazil
[3] Orthopaed & Traumatol Hosp Publ Workers State Sao, Sao Paulo, Brazil
关键词
Total knee arthroplasty; Baseplate; Failure analysis; Fracture; Prostheses design; REPLACEMENT; FRACTURE; WEAR; TRAY; UHMWPE; DESIGN; JOINTS; PERFORMANCE; PROSTHESIS; WALKING;
D O I
10.1016/j.engfailanal.2013.10.012
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The present work was performed on a retrieved modular prosthetic tibial component, which presented a fracture in the medial portion of the baseplate. In order to understand the causes that led to the failure of the baseplate, the steel, the mechanical and structural characteristics and the fracture were analyzed and correlated with numerical studies related to the biomechanical loads and displacements transmitted to the knee joint during human gait. The results showed a set of adverse events that occurred before the fracture of the baseplate. The steel, used for the manufacture of tibial component, had a high ferrite content, low mechanical strength, and did not comply with standard specifications. Design errors of the baseplate and insert assembly were shown to be the cause for the damage to the polyethylene of the insert. This damage included wear, plastic deformation and plaque formation by delamination, which increased the gap between the fitting guides. These phenomena enabled the relative displacement between the components, and the misaligning of the mechanical axis, causing an increase of the bending moment in the medial region of the baseplate. The load peaks concentrated in the primary contact area with the medial femoral condyle exceeded the mechanical strength limits of the piece, causing the fracture of the metallic tibial baseplate. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 50 条
  • [21] Intraosseous Tibial Resuscitation After a Total Knee Arthroplasty Leading to Osteonecrosis and Loosening of the Tibial Component
    Wenzel, Alyssa N.
    Auld, Thomas
    Bautista, Anson
    Huso, Tait
    Khanuja, Harpal S.
    ARTHROPLASTY TODAY, 2023, 19
  • [22] ANALYSIS OF TOTAL KNEE ARTHROPLASTY AFTER UPPER TIBIAL OSTEOTOMY
    HENDERSON, IJP
    BRYAN, RS
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (02): : 290 - 290
  • [23] The stability of the cemented tibial component of total knee arthroplasty
    Catani, F
    Leardini, A
    Ensini, A
    Cucca, G
    Bragonzoni, L
    Toksvig-Larsen, S
    Giannini, S
    JOURNAL OF ARTHROPLASTY, 2004, 19 (06): : 775 - 782
  • [24] Radiolucency adjacent to the tibial component in total knee arthroplasty
    Lundberg-Jensen, A
    Kourakis, AH
    Rossen, B
    Kofoed, H
    ORTHOPEDICS, 2002, 25 (01) : 49 - 51
  • [25] Rotational alignment of the tibial component in total knee arthroplasty
    Bindelglass, DF
    ORTHOPEDICS, 2001, 24 (11) : 1049 - 1052
  • [26] Improving tibial component alignment in total knee arthroplasty
    G. Cinotti
    P. Sessa
    A. D’Arino
    F. R. Ripani
    G. Giannicola
    Knee Surgery, Sports Traumatology, Arthroscopy, 2015, 23 : 3563 - 3570
  • [27] Rotational alignment of the tibial component in total knee arthroplasty
    Indelli, Pier Francesco
    Graceffa, Angelo
    Marcucci, Massimiliano
    Baldini, Andrea
    ANNALS OF TRANSLATIONAL MEDICINE, 2016, 4 (01)
  • [28] Improving tibial component alignment in total knee arthroplasty
    Cinotti, G.
    Sessa, P.
    D'Arino, A.
    Ripani, F. R.
    Giannicola, G.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (12) : 3563 - 3570
  • [29] Correlation of tibial component size and rotation with outcomes after total knee arthroplasty
    Antonio Klasan
    Joshua G. Twiggs
    Brett A. Fritsch
    Brad P. Miles
    Thomas J. Heyse
    Michael Solomon
    David A. Parker
    Archives of Orthopaedic and Trauma Surgery, 2020, 140 : 1819 - 1824
  • [30] A new landmark for measuring tibial component rotation after total knee arthroplasty
    Mahmood, Fand F.
    Holloway, Nicholas J.
    Deakin, Angela H.
    Clarke, Jon V.
    Picard, Frederic
    KNEE, 2019, 26 (03): : 779 - 786