Failure analysis of the tibial component baseplate after total knee arthroplasty

被引:4
|
作者
Oliveira, A. L. L. [1 ]
Cueva, E. G. [2 ]
Carvalho, R. T. [3 ]
机构
[1] Univ Sao Paulo, Dept Mech Engn, BR-2231 Sao Paulo, Brazil
[2] Univ Brasilia, UnB Gama Coll, Area Especial, Brasilia, DF, Brazil
[3] Orthopaed & Traumatol Hosp Publ Workers State Sao, Sao Paulo, Brazil
关键词
Total knee arthroplasty; Baseplate; Failure analysis; Fracture; Prostheses design; REPLACEMENT; FRACTURE; WEAR; TRAY; UHMWPE; DESIGN; JOINTS; PERFORMANCE; PROSTHESIS; WALKING;
D O I
10.1016/j.engfailanal.2013.10.012
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The present work was performed on a retrieved modular prosthetic tibial component, which presented a fracture in the medial portion of the baseplate. In order to understand the causes that led to the failure of the baseplate, the steel, the mechanical and structural characteristics and the fracture were analyzed and correlated with numerical studies related to the biomechanical loads and displacements transmitted to the knee joint during human gait. The results showed a set of adverse events that occurred before the fracture of the baseplate. The steel, used for the manufacture of tibial component, had a high ferrite content, low mechanical strength, and did not comply with standard specifications. Design errors of the baseplate and insert assembly were shown to be the cause for the damage to the polyethylene of the insert. This damage included wear, plastic deformation and plaque formation by delamination, which increased the gap between the fitting guides. These phenomena enabled the relative displacement between the components, and the misaligning of the mechanical axis, causing an increase of the bending moment in the medial region of the baseplate. The load peaks concentrated in the primary contact area with the medial femoral condyle exceeded the mechanical strength limits of the piece, causing the fracture of the metallic tibial baseplate. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 50 条
  • [31] Correlation of tibial component size and rotation with outcomes after total knee arthroplasty
    Klasan, Antonio
    Twiggs, Joshua G.
    Fritsch, Brett A.
    Miles, Brad P.
    Heyse, Thomas J.
    Solomon, Michael
    Parker, David A.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2020, 140 (11) : 1819 - 1824
  • [32] Various Types of Medial Tibial Bone Resorption after Total Knee Arthroplasty Using a Thick Cobalt Chromium Tibial Baseplate
    Song, Sang Jun
    Lee, Hyun Woo
    Kang, Se Gu
    Bae, Dae Kyung
    Park, Cheol Hee
    JOURNAL OF KNEE SURGERY, 2022, 35 (04) : 434 - 442
  • [33] Tibial Baseplate-Cement Interface Debonding in the ATTUNE Total Knee Arthroplasty System
    Torino, Daniel
    Damsgaard, Christopher
    Kolessar, David J.
    Hayes, Daniel S.
    Foster, Brian
    Constantino, Jesse
    Graham, Jove
    ARTHROPLASTY TODAY, 2022, 17 : 165 - 171
  • [34] Dissociation of the polyethylene insert from the tibial baseplate after total knee arthroplasty: a case report and review of the literature
    Wiley, Marcel R.
    Huo, Michael H.
    CURRENT ORTHOPAEDIC PRACTICE, 2018, 29 (02): : 180 - 183
  • [35] The effect of a tibial baseplate undersurface peripheral lip on cement penetration in total knee arthroplasty
    Vertullo, CJ
    Davey, JR
    JOURNAL OF ARTHROPLASTY, 2001, 16 (04): : 487 - 492
  • [36] Fixation of the tibial component of total knee arthroplasty after high tibial osteotomy -: A matched radiostereometric study
    Toksvig-Larsen, S
    Magyar, G
    Önsten, I
    Ryd, L
    Lindstrand, A
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02): : 295 - 297
  • [37] Varus tibial alignment is associated with greater tibial baseplate migration at 10 years following total knee arthroplasty
    Matthew G. Teeter
    Douglas D. Naudie
    Richard W. McCalden
    Xunhua Yuan
    David W. Holdsworth
    Steven J. MacDonald
    Brent A. Lanting
    Knee Surgery, Sports Traumatology, Arthroscopy, 2018, 26 : 1610 - 1617
  • [38] Osteollysis caused by tibial component debonding in total knee arthroplasty
    Cheng, Kevin
    Pruitt, Lisa
    Zaloudek, Charles
    Ries, Michael D.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (443) : 333 - 336
  • [39] Varus tibial alignment is associated with greater tibial baseplate migration at 10 years following total knee arthroplasty
    Teeter, Matthew G.
    Naudie, Douglas D.
    McCalden, Richard W.
    Yuan, Xunhua
    Holdsworth, David W.
    MacDonald, Steven J.
    Lanting, Brent A.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2018, 26 (06) : 1610 - 1617
  • [40] Hydroxyapatite enhances tibial component fixation in total knee arthroplasty
    Nelissen, RGHH
    Valstar, ER
    Rozing, PM
    SIROT 99, 1999, : 521 - 526