Numerical simulation of noninvasive blood pressure measurement

被引:5
|
作者
Hayashi, Satoru
Hayase, Toshiyuki
Shirai, Atsushi
Maruyama, Masaru
机构
[1] Tohoku Univ, Izumi Ku, Sendai, Miyagi 9813202, Japan
[2] Tohoku Univ, Inst Fluid Sci, Aoba Ku, Sendai, Miyagi 9808577, Japan
[3] Tohoku Elect Power Co Inc, Aoba Ku, Sendai, Miyagi 9808550, Japan
关键词
bio-fluid mechanics; noninvasive blood pressure measurement; Korotkoff sounds; collapsible tube; numerical analysis;
D O I
10.1115/1.2241592
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this paper a simulation model based on the partially pressurized collapsible tube model for reproducing noninvasive blood pressure measurement is presented. The model consists of a collapsible tube, which models the pressurized part of the artery, rigid pipes connected to the collapsible tube, which model proximal and distal region far from the pressurized part, and the Windkessel model, which represents the capacitance and the resistance of the distal part of the circulation. The blood flow is simplified to a one-dimensional system. Collapse and expansion of the tube is represented by the change in the cross-sectional area of the tube considering the force balance acting on the tube membrane in the direction normal to the tube axis. They are solved using the Runge-Kutta method. This simple model can easily reproduce the oscillation of inner fluid and corresponding tube collapse typical for the Korotkoff sounds generated by the cuff pressure. The numerical result is compared with the experiment and shows good agreement.
引用
收藏
页码:680 / 687
页数:8
相关论文
共 50 条
  • [41] NONINVASIVE LASER-BASED BLOOD-PRESSURE MEASUREMENT IN RABBITS
    HERROLD, EM
    GOLDWEIT, RS
    CARTER, JN
    ZUCCOTTI, G
    BORER, JS
    AMERICAN JOURNAL OF HYPERTENSION, 1992, 5 (03) : 197 - 202
  • [42] APPLICATION IN PSYCHOPHYSIOLOGY OF A NONINVASIVE, CONTINUOUS BLOOD-PRESSURE MEASUREMENT METHOD
    GODAERT, G
    HOMEOSTASIS IN HEALTH AND DISEASE, 1995, 36 (2-3): : 130 - 138
  • [43] Noninvasive versus invasive blood pressure measurement in normotensive and hypotensive baboons
    Hartford, CG
    Marcos, EF
    Rogers, GG
    LABORATORY ANIMAL SCIENCE, 1996, 46 (02): : 231 - 233
  • [44] A ring-type device for the noninvasive measurement of arterial blood pressure
    Zheng, DC
    Zhang, YT
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 3184 - 3187
  • [45] Inflationary noninvasive blood pressure measurement reduces the incidence of subcutaneous hemorrhage
    Nukita, Shizuko
    Nakano, Sumie
    Morimoto, Yasuhiro
    Shiramoto, Hiroko
    BLOOD PRESSURE MONITORING, 2020, 25 (01) : 39 - 41
  • [46] A Literature Review on Current and Proposed Technologies of Noninvasive Blood Pressure Measurement
    Mukherjee, Ramtanu
    Ghosh, Sanchita
    Gupta, Bharat
    Chakravarty, Tapas
    TELEMEDICINE AND E-HEALTH, 2018, 24 (03) : 185 - 193
  • [47] Noninvasive Continuous Blood Pressure Measurement with Wearable Millimeter Wave Device
    Liao, Catherine
    Shay, Oliver
    Gomes, Elizabeth
    Bikhchandani, Nikhil
    2021 IEEE 17TH INTERNATIONAL CONFERENCE ON WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS (BSN), 2021,
  • [48] A novel noninvasive method for automatic measurement of systolic and diastolic blood pressure
    Nitzan, M
    Babchenko, A
    Rosenfeld, C
    Milishanski, I
    JOURNAL OF HYPERTENSION, 2000, 18 : S43 - S43
  • [49] ACCURACY OF CONTINUOUS NONINVASIVE BLOOD-PRESSURE MEASUREMENT AT HIGH AND LOW BLOOD PRESSURES
    WONG, DH
    EATON, S
    CLACK, S
    ZACCARI, J
    SURESH, K
    BOGARD, D
    WATSON, TB
    ANESTHESIA AND ANALGESIA, 1994, 78 (02): : U257 - U257
  • [50] Iterative Learning Cascade Control of Continuous Noninvasive Blood Pressure Measurement
    Seel, Thomas
    Schauer, Thomas
    Weber, Sarah
    Affeld, Klaus
    2013 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN, AND CYBERNETICS (SMC 2013), 2013, : 2207 - 2212