摘要
Background. Flattening filter-free (FFF) beams are an emerging technology that has not yet been widely implemented as standard practice in radiotherapy centers. To facilitate the clinical implementation of FFF, we attempted to elucidate the difference in plan quality and treatment delivery time compared to flattening filter beams (i.e. standard, STD) for several patient groups. We hypothesize that the treatment plan quality is comparable while the treatment delivery time of volumetric modulated arc therapy (VMAT) is considerably shorter using FFF beams, especially for stereotactic treatments. Methods. A total of 120 patients treated for head and neck (H&N) tumors, high-grade glioma, prostate cancer, early stage lung cancer and intra-cranial metastatic disease (both single and multiple metastases) were included in the study. For each cohort, 20 consecutive patients were selected. The plans were generated using STD-and FFF-VMAT for both 6 MV and 10 MV, and were compared with respect to plan quality, monitor units and delivery time using Wilcoxon signed rank tests. Results. For H&N and high-grade gliomas, there was a significant difference in homogeneity index in favor for STD-VMAT (p < 0.001). For the stereotactic sites there were no differences in plan conformity. Stereotactic FFF-VMAT plans required significantly shorter delivery time compared to STD-VMAT plans (p < 0.001) for higher dose per fraction, on average 54.5% for 6 MV and 71.4% for 10 MV. FFF-VMAT generally required a higher number of MU/Gy (p < 0.001), on average 7.0% for 6 MV and 8.4% for 10 MV. Conclusion. It was generally possible to produce FFF-VMAT plans with the same target dose coverage and doses to organs at risk as STD-VMAT plans. Target dose homogeneity tended to be somewhat inferior for FFF-VMAT for the larger targets investigated. For stereotactic radiotherapy, FFF-VMAT resulted in a considerable time gain while maintaining similar plan quality compared to STD beams.
引用
收藏
页码:1005 / 1011
页数:7
相关论文
共 50 条
  • [21] Dosimetric Comparison of Unmatched Flattening Filter-free and Flattened Beams in Volumetric Arc Therapy Plans for Head-and-neck Cancer
    Kumar, Arvind
    Sharma, Kiran
    Shan, Chandi Prasad
    Garg, Abhishek
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (04) : 338 - 344
  • [22] Robustness quantification methods comparison in volumetric modulated arc therapy to treat head and neck cancer
    Liu, Wei
    Patel, Samir H.
    Shen, Jiajian
    Hu, Yanle
    Harrington, Daniel P.
    Ding, Xiaoning
    Halyard, Michele Y.
    Schild, Steven E.
    Wong, William W.
    Ezzell, Gary A.
    Bues, Martin
    PRACTICAL RADIATION ONCOLOGY, 2016, 6 (06) : E269 - E275
  • [23] Comparison of Treatment Planning Systems in Elekta Volumetric Modulated Arc Therapy (Elekta VMAT) - Prostate Cancer Study
    Haga, A.
    Kida, S.
    Okano, Y.
    Itoh, S.
    Matsuura, T.
    Saotome, N.
    Sakumi, A.
    Kotoku, J.
    Kenshiro, S.
    Nakagawa, K.
    Iwai, Y.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [24] Comparing four volumetric modulated arc therapy beam arrangements for the treatment of early-stage prostate cancer
    Elith, Craig A.
    Dempsey, Shane E.
    Warren-Forward, Helen M.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2014, 61 (02): : 91 - 101
  • [25] A planning comparison of 3-dimensional conformal multiple static field, conformal arc, and volumetric modulated arc therapy for the delivery of stereotactic body radiotherapy for early stage lung cancer
    Dickey, Mike
    Roa, Wilson
    Drodge, Suzanne
    Ghosh, Sunita
    Murray, Brad
    Scrimger, Rufus
    Gabos, Zsolt
    MEDICAL DOSIMETRY, 2015, 40 (04) : 347 - 351
  • [26] Volumetric modulated arc radiation therapy in the treatment of lung cancer: images of treatment planning case
    Andaloussi-Saghir, Khalid
    Mansouri, Hamid
    PAN AFRICAN MEDICAL JOURNAL, 2014, 18
  • [27] Real-time interactive planning for radiotherapy of head and neck cancer with volumetric modulated arc therapy
    Baker, Lindsey
    Olson, Robert
    Braich, Taran
    Koulis, Theodora
    Ye, Allison
    Ahmed, Nisar
    Tran, Eric
    Lawyer, Kim
    Otto, Karl
    Smith, Sally
    Mestrovic, Ante
    Matthews, Quinn
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2019, 9 : 83 - 88
  • [28] A dosimetric evaluation of flattening filter-free volumetric modulated arc therapy for postoperative treatment of cervical cancer
    Fuli Zhang
    Huayong Jiang
    Weidong Xu
    Yadi Wang
    Junmao Gao
    Qingzhi Liu
    Ping Wang
    Na Lu
    Diandian Chen
    Bo Yao
    Jun Hou
    Heliang He
    Jianping Chen
    Oncology and Translational Medicine, 2016, 2 (04) : 179 - 184
  • [29] Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer
    Alvarez-Moret, Judith
    Pohl, Fabian
    Koelbl, Oliver
    Dobler, Barbara
    RADIATION ONCOLOGY, 2010, 5
  • [30] Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer
    Judith Alvarez-Moret
    Fabian Pohl
    Oliver Koelbl
    Barbara Dobler
    Radiation Oncology, 5