The principle and calibration of the third gear flow- meter

被引:3
|
作者
Zhang Jun [1 ]
Niu Meng-meng [1 ]
Tang Hong Mei [1 ]
Li Xian Hua [1 ]
Tang Cunren [1 ]
机构
[1] Anhui Univ Sci & Technol, Sch Mech Engn, Huainan 232001, Peoples R China
关键词
gear flow-meter; calibration experiment; pressure;
D O I
10.4028/www.scientific.net/AMM.397-400.1713
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
At present, the domestic development of the gear flow-meter is far behind the developed countries, especially in the micro gear flow-meter. This paper proposes a typical structure named the third gear flow-meter. The structural characteristics and working principle of the third gear flow-meter was introduced in detail, the prototype of the third gear flow-meter was designed and processed, and the calibration tests and pressure experiments of the third gear flow-meter prototype was made at last. We can draw that the three gear flow-meter can be used in hydraulic system pressure range is less than 25Mpa, the flow rate was 0.15 similar to 0.2m(3)/h can meet the high requirements of the stability of measurement, and the measurement accuracy in the whole flow range is 0.2 class. This article provides a reliable experimental data for the design of the micro gear flow-meter.
引用
收藏
页码:1713 / 1717
页数:5
相关论文
共 50 条
  • [21] UNCERTAINTY EVALUATION OF FLOW METER CALIBRATION BY GRAVIMETRIC WATER FLOW STANDARDS AT KRISS
    Chun, Sejong
    Yoon, Byung-Ro
    PROCEEDINGS OF THE ASME 2020 FLUIDS ENGINEERING DIVISION SUMMER MEETING (FEDSM2020), VOL 1, 2020,
  • [22] Numerical calibration and unsteady cavitation simulation of turbine flow meter
    Chen, Gang
    Cao, Guang-Jun
    Fu, Su-Hong
    Li, Ji-Feng
    Zhang, Dong-Sheng
    Kung Cheng Je Wu Li Hsueh Pao/Journal of Engineering Thermophysics, 2010, 31 (SUPPL.): : 269 - 272
  • [23] CALIBRATION OF A FLOSCOPE PEAK FLOW METER AND A VITALOGRAPH PULMONARY MONITOR
    FISHER, J
    SHAW, A
    ANAESTHESIA, 1981, 36 (09) : 886 - 888
  • [24] Bayesian estimation of a polynomial calibration function associated to a flow meter
    Yardin, C.
    Amar, S.
    Fischer, N.
    Sancandi, M.
    Keller, M.
    ADVANCED MATHEMATICAL AND COMPUTATIONAL TOOLS IN METROLOGY AND TESTING XI, 2019, 89 : 417 - 426
  • [25] Simulation of flow meter calibration factors for various installation effects
    Lund Inst of Technology, Lund, Sweden
    Meas J Int Meas Confed, 4 (235-244):
  • [26] Line in flow-meter calibration advance methods and solutions
    Sagi, H
    Roberts, WG
    SUCCESS IN THE 21ST CENTURY DEPENDS ON MODERN METROLOGY, VOLS 1-2, 1997, : 383 - 399
  • [27] ANALYTICAL PROCEDURE OF CALIBRATION FOR PULSED ULTRASONIC DOPPLER FLOW METER
    JORGENSEN, JE
    GARBINI, JL
    JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 1974, 96 (02): : 158 - 167
  • [28] Blow-down calibration of a large ultrasonic flow meter
    Johnson, Aaron N.
    Harman, Eric
    Boyd, Joey T.
    FLOW MEASUREMENT AND INSTRUMENTATION, 2021, 77 (77)
  • [29] SIMULTANEOUS COMPARISON OF RENAL BLOOD FLOW AS MEASURED BY THE FICK PRINCIPLE AND BY THE BUBBLE FLOW METER
    CONN, HL
    MARKLEY, K
    AMERICAN JOURNAL OF PHYSIOLOGY, 1950, 160 (03): : 547 - 551
  • [30] Performance of a gas flow meter calibration system utilizing critical flow venturi standards
    Michael Carter
    William Johansen
    Charles Britton
    MAPAN, 2011, 26 : 247 - 254