The syringe potentiometer: a low-cost device for pneumotachograph calibration

被引:4
|
作者
Cross, Troy J. [1 ,2 ]
Kelley, Eli F. [1 ]
Hardy, Tim A. [3 ]
Isautier, Jennifer M. J. [1 ]
Johnson, Bruce D. [1 ]
机构
[1] Mayo Clin, Dept Cardiovasc Dis, Rochester, MN 55902 USA
[2] Griffith Univ, Menzies Hlth Inst Queensland, Brisbane, Qld, Australia
[3] Univ Leeds, Sch Biomed Sci, Leeds, W Yorkshire, England
关键词
calibration; pneumotachograph; potentiometer; respiratory airflow; syringe; ACCURACY;
D O I
10.1152/japplphysiol.00150.2019
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this report was to 1) detail the construction of a low-cost device that provides a "reference" flow waveform for pneumotachograph (PNT) calibration, i.e., the "syringe potentiometer" (SP), and to compare the 2) accuracy and 3) practical performance of "calibration curves" obtained with the SP device to other more established methods of PNT calibration, i.e., the weighted averaging (WA) and polynomial least-squares (PolyLS) methods. Volume and flow waveforms obtained via the SP device were validated against a motion capture system and were deemed accurate surrogates of actual syringe volume and flows. The SP device was used to construct a calibration curve of a PNT by dividing the flow waveform of the SP by the analog output of the PNT amplifier. A total of 187 inspiratory and 187 expiratory strokes were collected. When the entire data set of expiratory strokes was used, the SP, WA, and PolyLS methods together demonstrated acceptable volume and flow errors as per American Thoracic Society/European Respiratory Society recommendations (less than +/- 3.5% and less than +/- 5.0% errors, respectively). The "practical" performance of each method was assessed with a nested subsampling procedure, whereby volume and flow errors were evaluated as the number of strokes was increased (in blocks of 5 strokes). To this end, the SP method demonstrated practical performance superior to that of the WA and PolyLS approaches, whereby acceptable volume and flow errors were achieved after only 5 calibration strokes; the WA and PolyLS methods required 15 and 20 strokes, respectively, to achieve the same level of volume and flow accuracy. NEW & NOTEWORTHY This report describes the construction and validation of a low-cost device for the purposes of pneumotachograph (PNT) calibration: the "syringe potentiometer" (SP). Calibration of a PNT with the SP device yielded acceptable volume and flow errors (<3.5% and 5%, respectively) across a wide range of flows (<0.5 to 15 L/s). The SP device offered superior "practical performance" over other established PNT calibration methods, whereby acceptable volume and flow errors were achieved after only five calibration strokes.
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [31] Calibration of a low-cost piezoelectrical mirror for DSPI
    Asmad, M
    Baldwin, G
    Rabal, H
    Arizaga, R
    Trivi, M
    OPTICS AND LASERS IN ENGINEERING, 2003, 40 (1-2) : 25 - 31
  • [32] GNSS Based Low-Cost Magnetometer Calibration
    Andel, Jan
    Simak, Vojtech
    Kanalikova, Alzbeta
    Pirnik, Rastislav
    SENSORS, 2022, 22 (21)
  • [33] A Low-Cost Calibration Method for Low-Cost MEMS Accelerometers Based on 3D Printing
    Garcia, Jesus A.
    Lara, Evangelina
    Aguilar, Leocundo
    SENSORS, 2020, 20 (22) : 1 - 19
  • [34] Low-Cost Device for Transcutaneous Visualization of Veins
    Goldan, Daniela
    Ursache, Teofil Ilie
    Rotariu, Cristian
    ADVANCES IN DIGITAL HEALTH AND MEDICAL BIOENGINEERING, VOL 1, EHB-2023, 2024, 109 : 343 - 349
  • [35] Low-cost device for constructing sorption isotherms
    Gibert, O.
    Meot, J. M.
    Marouze, C.
    Brouat, J.
    DRYING TECHNOLOGY, 2006, 24 (12) : 1697 - 1704
  • [36] Erratum to: A low-cost visual occlusion device
    Sangcheol Lee
    Sekee Kil
    Taewhan Kim
    Behavior Research Methods, 2014, 46 (4) : 949 - 949
  • [37] A low-cost device for cryoanesthesia of neonatal rodents
    Jamieson, Bradley B.
    Cano-Ferrer, Xavier
    Konstantinou, George
    de Launoit, Elisa
    Renier, Nicolas
    Imbert, Albane
    Kohl, Johannes
    HARDWAREX, 2023, 14
  • [38] A low-cost device for beer color measurement
    Di Caro, Domenico
    Liguori, Consolatina
    Pietrosanto, Antonio
    Sommella, Paolo
    2019 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR AGRICULTURE AND FORESTRY (METROAGRIFOR), 2019, : 222 - 226
  • [39] Low-Cost Interactive Device for Virtual Reality
    Asadzadeh, Afsoon
    Samad-Soltani, Taha
    Rezaei-hachesu, Peyman
    Salahzadeh, Zahra
    2020 6TH INTERNATIONAL CONFERENCE ON WEB RESEARCH (ICWR), 2020, : 38 - 42
  • [40] Development of low-cost photodynamic therapy device
    Momchilov, N.
    Bliznakova, I.
    Borisova, E.
    Troyanova, P.
    Avramov, L.
    ACTA PHYSICA POLONICA A, 2007, 112 (05) : 1125 - 1130