The syringe potentiometer: a low-cost device for pneumotachograph calibration

被引:4
|
作者
Cross, Troy J. [1 ,2 ]
Kelley, Eli F. [1 ]
Hardy, Tim A. [3 ]
Isautier, Jennifer M. J. [1 ]
Johnson, Bruce D. [1 ]
机构
[1] Mayo Clin, Dept Cardiovasc Dis, Rochester, MN 55902 USA
[2] Griffith Univ, Menzies Hlth Inst Queensland, Brisbane, Qld, Australia
[3] Univ Leeds, Sch Biomed Sci, Leeds, W Yorkshire, England
关键词
calibration; pneumotachograph; potentiometer; respiratory airflow; syringe; ACCURACY;
D O I
10.1152/japplphysiol.00150.2019
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this report was to 1) detail the construction of a low-cost device that provides a "reference" flow waveform for pneumotachograph (PNT) calibration, i.e., the "syringe potentiometer" (SP), and to compare the 2) accuracy and 3) practical performance of "calibration curves" obtained with the SP device to other more established methods of PNT calibration, i.e., the weighted averaging (WA) and polynomial least-squares (PolyLS) methods. Volume and flow waveforms obtained via the SP device were validated against a motion capture system and were deemed accurate surrogates of actual syringe volume and flows. The SP device was used to construct a calibration curve of a PNT by dividing the flow waveform of the SP by the analog output of the PNT amplifier. A total of 187 inspiratory and 187 expiratory strokes were collected. When the entire data set of expiratory strokes was used, the SP, WA, and PolyLS methods together demonstrated acceptable volume and flow errors as per American Thoracic Society/European Respiratory Society recommendations (less than +/- 3.5% and less than +/- 5.0% errors, respectively). The "practical" performance of each method was assessed with a nested subsampling procedure, whereby volume and flow errors were evaluated as the number of strokes was increased (in blocks of 5 strokes). To this end, the SP method demonstrated practical performance superior to that of the WA and PolyLS approaches, whereby acceptable volume and flow errors were achieved after only 5 calibration strokes; the WA and PolyLS methods required 15 and 20 strokes, respectively, to achieve the same level of volume and flow accuracy. NEW & NOTEWORTHY This report describes the construction and validation of a low-cost device for the purposes of pneumotachograph (PNT) calibration: the "syringe potentiometer" (SP). Calibration of a PNT with the SP device yielded acceptable volume and flow errors (<3.5% and 5%, respectively) across a wide range of flows (<0.5 to 15 L/s). The SP device offered superior "practical performance" over other established PNT calibration methods, whereby acceptable volume and flow errors were achieved after only five calibration strokes.
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [21] Sitec develops low-cost device
    不详
    CIM BULLETIN, 2004, 97 (1079): : 19 - 19
  • [22] Low-Cost NIR Measurement Device
    Mustafa, Luka
    Cerncic, Eva
    2015 4TH INTERNATIONAL WORKSHOP ON OPTICAL WIRELESS COMMUNICATIONS (IWOW), 2015, : 127 - 131
  • [23] A low-cost visual occlusion device
    Sancheol Lee
    Sekee Kil
    Taewhan Kim
    Behavior Research Methods, 2014, 46 : 935 - 948
  • [24] Development of Low-cost Wire Type Linear Potentiometer for Flexible Spherical Actuator
    Matsui, Yasuko
    Akagi, Tetsuya
    Dohta, Shujiro
    2016 IEEE INTERNATIONAL CONFERENCE ON ADVANCED INTELLIGENT MECHATRONICS (AIM), 2016, : 1017 - 1021
  • [25] Calibration method enabling low-cost SDR
    Debaillie, Bjoern
    Van Wesemael, Peter
    Craninckx, Jan
    2008 IEEE INTERNATIONAL CONFERENCE ON COMMUNICATIONS, PROCEEDINGS, VOLS 1-13, 2008, : 4899 - 4903
  • [26] The calibration and deployment of a low-cost methane sensor
    Riddick, Stuart N.
    Mauzerall, Denise L.
    Celia, Michael
    Allen, Grant
    Pitt, Joseph
    Kang, Mary
    Riddick, John C.
    ATMOSPHERIC ENVIRONMENT, 2020, 230
  • [27] Evaluation of low-cost formaldehyde sensors calibration
    Alonso, Maria Justo
    Madsen, Henrik
    Liu, Peng
    Jorgensen, Rikke Bramming
    Jorgensen, Thomas Berg
    Christiansen, Even Johan
    Myrvang, Olav Aleksander
    Bastien, Diane
    Mathisen, Hans Martin
    BUILDING AND ENVIRONMENT, 2022, 222
  • [28] Automatic calibration of low-cost digital cameras
    Habib, AF
    Morgan, MF
    OPTICAL ENGINEERING, 2003, 42 (04) : 948 - 955
  • [29] Calibration of Low-cost Triaxial Inertial Sensors
    Rohac, Jan
    Sipos, Martin
    Simanek, Jakub
    IEEE INSTRUMENTATION & MEASUREMENT MAGAZINE, 2015, 18 (06) : 32 - 38
  • [30] LOW-COST CALIBRATION METHOD FOR THE INFRARED CAMERA
    Uemura, Derek Koji
    Shah, Sanjay B.
    Regmi, Prafulla
    Grimes, Jesse
    Wang-Li, Lingjuan
    APPLIED ENGINEERING IN AGRICULTURE, 2023, 39 (05) : 529 - 534