The syringe potentiometer: a low-cost device for pneumotachograph calibration

被引:4
|
作者
Cross, Troy J. [1 ,2 ]
Kelley, Eli F. [1 ]
Hardy, Tim A. [3 ]
Isautier, Jennifer M. J. [1 ]
Johnson, Bruce D. [1 ]
机构
[1] Mayo Clin, Dept Cardiovasc Dis, Rochester, MN 55902 USA
[2] Griffith Univ, Menzies Hlth Inst Queensland, Brisbane, Qld, Australia
[3] Univ Leeds, Sch Biomed Sci, Leeds, W Yorkshire, England
关键词
calibration; pneumotachograph; potentiometer; respiratory airflow; syringe; ACCURACY;
D O I
10.1152/japplphysiol.00150.2019
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this report was to 1) detail the construction of a low-cost device that provides a "reference" flow waveform for pneumotachograph (PNT) calibration, i.e., the "syringe potentiometer" (SP), and to compare the 2) accuracy and 3) practical performance of "calibration curves" obtained with the SP device to other more established methods of PNT calibration, i.e., the weighted averaging (WA) and polynomial least-squares (PolyLS) methods. Volume and flow waveforms obtained via the SP device were validated against a motion capture system and were deemed accurate surrogates of actual syringe volume and flows. The SP device was used to construct a calibration curve of a PNT by dividing the flow waveform of the SP by the analog output of the PNT amplifier. A total of 187 inspiratory and 187 expiratory strokes were collected. When the entire data set of expiratory strokes was used, the SP, WA, and PolyLS methods together demonstrated acceptable volume and flow errors as per American Thoracic Society/European Respiratory Society recommendations (less than +/- 3.5% and less than +/- 5.0% errors, respectively). The "practical" performance of each method was assessed with a nested subsampling procedure, whereby volume and flow errors were evaluated as the number of strokes was increased (in blocks of 5 strokes). To this end, the SP method demonstrated practical performance superior to that of the WA and PolyLS approaches, whereby acceptable volume and flow errors were achieved after only 5 calibration strokes; the WA and PolyLS methods required 15 and 20 strokes, respectively, to achieve the same level of volume and flow accuracy. NEW & NOTEWORTHY This report describes the construction and validation of a low-cost device for the purposes of pneumotachograph (PNT) calibration: the "syringe potentiometer" (SP). Calibration of a PNT with the SP device yielded acceptable volume and flow errors (<3.5% and 5%, respectively) across a wide range of flows (<0.5 to 15 L/s). The SP device offered superior "practical performance" over other established PNT calibration methods, whereby acceptable volume and flow errors were achieved after only five calibration strokes.
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [1] The "Syringe Potentiometer": A Low-cost Solution For Accurate and Precise Calibration Of Pneumotachographs
    Cross, Troy J.
    Kelley, Eli
    Hardy, Timothy
    Johnson, Bruce
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2019, 51 (06): : 655 - 655
  • [2] LOW-COST MICROVOLT POTENTIOMETER
    WOOD, WH
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1957, 28 (03): : 202 - 203
  • [3] LOW-COST THERMOCOUPLE POTENTIOMETER
    WORRELL, FT
    AMERICAN JOURNAL OF PHYSICS, 1963, 31 (03) : 216 - &
  • [4] A Fast Calibration Method for Pneumotachograph with a 3L Syringe
    Li, Yueqi
    Qiu, Xin
    Zhang, Hao
    Xu, Lirui
    Lu, Saihu
    Du, Lidong
    Chen, Xianxiang
    Fang, Zhen
    BIOENGINEERING-BASEL, 2023, 10 (09):
  • [5] CMOS IC makes low-cost digital potentiometer
    Cossio, JC
    EDN, 1996, 41 (02) : 99 - 100
  • [6] EFFECTIVE, LOW-COST POLE-SYRINGE
    MULDER, JB
    VETERINARY MEDICINE & SMALL ANIMAL CLINICIAN, 1977, 72 (06): : 1001 - 1001
  • [7] Low-cost, low-power, clockwork syringe pump
    Pooke, Francis
    Payne, Matthew
    Holder-Pearson, Lui
    Heaton, Doug
    Campbell, Jake
    Chase, J. Geoffrey
    HARDWAREX, 2023, 16
  • [8] Validation of Chloe SED, a Low-Cost, Reusable Syringe Extension Device for the Provision of Paracervical Analgesia
    Ramanathan, Aparna
    OBSTETRICS AND GYNECOLOGY, 2022, 139 : 7S - 7S
  • [9] A low-cost spacer device
    Gordon, S
    TROPICAL DOCTOR, 1999, 29 (02) : 116 - 116
  • [10] A LOW-COST ROTARY OPTICAL POTENTIOMETER MECHANISM BASED ON OPTOELECTRONICS TECHNIQUE
    Kwon, Young-shin
    Kim, Won-jong
    7TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2014, VOL 3, 2014,