Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation

被引:46
|
作者
Harju, Jarkko [1 ]
Tarniceriu, Adrian [2 ]
Parak, Jakub [3 ,4 ]
Vehkaoja, Antti [3 ,4 ]
Yli-Hankala, Arvi [1 ,5 ]
Korhonen, Ilkka [3 ,4 ]
机构
[1] Tampere Univ Hosp, Dept Anesthesia, Tampere, Finland
[2] PulseOn SA, Neuchatel, Switzerland
[3] Tampere Univ Technol, Tampere, Finland
[4] Pulseon Oy, Espoo, Finland
[5] Univ Tampere, Med Sch, Tampere, Finland
关键词
heart rate; pulse rate; photoplethysmography; atrial fibrillation; perioperative monitoring; RATE-VARIABILITY; ACCURACY; PREVALENCE; FAILURE; WORN;
D O I
10.1088/1361-6579/aac9a9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: Atrial fibrillation (AF) causes marked risk for patients, while silent fibrillation may remain unnoticed if not suspected and screened. Development of comfortable yet accurate beat-to-beat heart rate (HR) monitoring with good AF detection sensitivity would facilitate screening and improve treatment. The purpose of this study was to evaluate whether a wrist-worn photoplethysmography (PPG) device can be used to monitor beat-to-beat HR accurately during post-operative treatment in patients suffering from AF and whether wrist-PPG can be used to distinguish AF from sinus rhythm (SR). Approach: Twenty-nine patients (14 with AF, 15 with SR, mean age 71.5 years) with multiple comorbidities were monitored during routine post-operative treatment. The monitoring included standard ECG, finger PPG monitoring and a wrist-worn PPG monitor with green and infrared light sources. The HR from PPG sensors was compared against ECG-derived HR. Main results: The wrist PPG technology had very good HR and beat detection accuracy when using green light. For the SR group, the mean absolute error (MAE) for HR was 1.50 bpm, and for the interbeat intervals (IBI), the MAE was 7.64 ms. For the AF group, the MAE for HR was 4.28 bpm and for IBI, the MAE was 14.67 ms. Accuracy for the infrared (IR) channel was worse. Finger PPG provided similar accuracy for HR and better accuracy for the IBI. AF detection sensitivity using green light was 99.0% and the specificity was 93.0%. Performance can be improved by discarding unreliable IBI periods. Significance: Results suggest that wrist PPG measurement allows accurate HR and beat-to-beat HR monitoring also in AF patients, and could be used for differentiating between SR and AF with very good sensitivity.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] Effects of azimilide on heart rate and ECG conduction intervals during sinus rhythm in patients with a history of atrial fibrillation
    Pritchett, ELC
    Schulte, MC
    Schnell, D
    Marcello, SR
    Wilkinson, WE
    Page, RL
    Connolly, SJ
    JOURNAL OF CLINICAL PHARMACOLOGY, 2002, 42 (04): : 388 - 394
  • [32] Influence of heart rate on prognosis of patients with heart failure and atrial fibrillation
    Mareev, Y.
    Petruchina, A.
    Goryunova, T.
    Sokolov, S.
    Mareev, V.
    EUROPEAN JOURNAL OF HEART FAILURE, 2014, 16 : 29 - 30
  • [33] HEART RATE IS NOT RELATED TO SURVIVAL IN PATIENTS WITH HEART FAILURE AND ATRIAL FIBRILLATION
    Hawkey, Sean
    Levin, Daniel
    Mohan, Mohapradeep
    Baig, Fatima
    Iskandar, Muhammad
    Rutherford, Lynn
    Lang, Chim
    Choy, Anna
    HEART, 2015, 101 : A15 - A16
  • [34] Heart rate is not related to survival in patients with heart failure and atrial fibrillation
    Hawkey, S.
    Levin, D.
    Mohan, M.
    Baig, F.
    Rutherford, L.
    Iskandar, M.
    Lang, C. H. I. M.
    Choy, A.
    EUROPEAN JOURNAL OF HEART FAILURE, 2015, 17 : 177 - 177
  • [35] Ventricular rate and beat-to-beat variation of stroke volume in atrial fibrillation
    Kerr, AJ
    Williams, MJA
    Stewart, RAH
    AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (09): : 1116 - 1119
  • [36] Heart rate variability in patients with atrial fibrillation of sinus rhythm or atrial fibrillation: chaos or merit?
    Zhang, Lifan
    Li, Bingxun
    Wu, Lin
    ANNALS OF MEDICINE, 2025, 57 (01)
  • [37] Innovative approaches in investigating inter-beat intervals: Graph theoretical method suggests altered autonomic functioning in adolescents with ADHD
    Kvadsheim, Elisabet
    Fasmer, Ole Bernt
    Fasmer, Erlend Eindride
    R. Hauge, Erik
    Thayer, Julian F.
    Osnes, Berge
    Haavik, Jan
    Koenig, Julian
    Adolfsdottir, Steinunn
    Plessen, Kerstin Jessica
    Sorensen, Lin
    PSYCHOPHYSIOLOGY, 2022, 59 (06)
  • [38] Association between Atrial Fibrillatory Rate and Heart Rate Variability in Patients with Atrial Fibrillation and Congestive Heart Failure
    Corino, Valentina D. A.
    Cygankiewicz, Iwona
    Mainardi, Luca T.
    Stridh, Martin
    Vasquez, Rafael
    de Luna, Antonio Bayes
    Holmqvist, Fredrik
    Zareba, Wojciech
    Platonov, Pyotr G.
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2013, 18 (01) : 41 - 50
  • [39] Automated Home Monitoring of Atrial Fibrillation in Heart Failure Patients
    Dovancescu, Silviu
    Babaeizadeh, Saeed
    2015 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), 2015, 42 : 261 - 264
  • [40] DropConnected neural networks trained on time-frequency and inter-beat features for classifying heart sounds
    Kay, Edmund
    Agarwal, Anurag
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (08) : 1645 - 1657