On-Line Dialysate Infusion to Estimate Absolute Blood Volume in Dialysis Patients

被引:29
|
作者
Schneditz, Daniel [1 ]
Schilcher, Gernot [2 ]
Ribitsch, Werner [2 ]
Krisper, Peter [2 ]
Haditsch, Bernd [1 ]
Kron, Joachim [3 ]
机构
[1] Med Univ Graz, Inst Physiol, A-8010 Graz, Austria
[2] Med Univ Graz, Clin Div Nephrol, Dept Internal Med, A-8010 Graz, Austria
[3] KfH Kidney Ctr Berlin Kopenick, KfH Kuratorium Dialyse & Nierentransplantat, Berlin, Germany
关键词
ultra-pure dialysate; indicator dilution; volume kinetics; blood volume; half-life; TOTAL PROTEIN-CONCENTRATION; CARDIOPULMONARY RECIRCULATION; SOUND-SPEED; HEMODIALYSIS; HYPOTENSION; PRESSURE; HEMATOCRIT; DENSITY; FLUID;
D O I
10.1097/MAT.0000000000000086
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It was the aim to measure the distribution volume and the elimination of ultra-pure dialysate in stable hemodialysis patients during on-line hemodiafiltration (HDF). Dialysate was automatically infused as a volume indicator using standard on-line HDF equipment. Indicator concentration was noninvasively measured in the arterial blood-line (using the blood volume monitor, Fresenius Medical Care, Bad Homburg vor der H he, Germany), and its time course was analyzed to obtain the elimination rate and the distribution volume V-t at the time of dilution. Blood volume at treatment start (V-0) was calculated accounting for the degree of intradialytic hemoconcentration. Five patients (two females) were studied during 15 treatments. Two to six measurements using indicator volumes ranging from 60 to 210 ml were done in each treatment. V-0 was 4.59 +/- 1.15 L and larger than the volume of 4.08 +/- 0.48 L estimated from anthropometric relationships. The mean half-life of infused volume was 17.2 +/- 29.7 min. Given predialysis volume expansion V-0 was consistent with blood volume determined from anthropometric measurements. Information on blood volume could substantially improve volume management in hemodialysis patients and fluid therapy in intensive care patients undergoing extracorporeal blood treatment. The system has the potential for complete automation using proper control inputs for BVM and HDF modules of the dialysis machine.
引用
收藏
页码:436 / 442
页数:7
相关论文
共 50 条
  • [1] Variable-Volume Kinetic Model to Estimate Absolute Blood Volume in Patients on Dialysis Using Dialysate Dilution
    Samandari, Hamed
    Schneditz, Daniel
    Germain, Michael J.
    Horowitz, Joseph
    Hollot, Christopher V.
    Chait, Yossi
    ASAIO JOURNAL, 2018, 64 (01) : 77 - 85
  • [2] An improved method to estimate absolute blood volume based on dialysate dilution
    Kron, Susanne
    Schneditz, Daniel
    Keane, David F.
    Leimbach, Til
    Kron, Joachim
    ARTIFICIAL ORGANS, 2021, 45 (09) : E359 - E363
  • [3] Estimation of absolute blood volume in hemodialysis patients: A numerical algorithm for assessing blood volume increase after dialysate bolus infusion
    Pstras, Leszek
    Krenn, Simon
    Waniewski, Jacek
    Schmiedecker, Michael
    Mussnig, Sebastian
    Niknam, Janosch
    Wabel, Peter
    Mayer, Christopher C.
    Schneditz, Daniel
    Hecking, Manfred
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 88
  • [4] On-line blood and dialysate sampling system for multiparametric monitoring during dialysis
    Lurzer, W
    Scharfetter, H
    Hutten, H
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 77 - 78
  • [5] DIALYSATE CONSUMPTION AND DIALYSIS DOSE IN ON-LINE HEMODIAFILTRATION WITH AUTOMATED COUPLING DIALYSATE/BLOOD FLOW FUNCTION (AUTOFLOW FUNCTION)
    Park, Yongki
    Jeon, Jimin
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 206 - 207
  • [6] Dialysate/infusate composition and infusion mode in on-line hemodiafiltration
    Pedrini, LA
    De Cristofaro, V
    Pagliari, B
    Ruggiero, P
    HEMODIALYSIS TECHNOLOGY, 2002, 137 : 344 - 349
  • [7] «On-line» Preparation of Bicarbonate-containing Dialysate for Use in Peritoneal Dialysis
    Ing, T.S.
    Quon, M.J.
    Daugirdas, J.T.
    Liu, P.
    Gandhi, V.C.
    Reid, R.R.
    International Journal of Artificial Organs, 1981, 4 (06): : 308 - 309
  • [8] Feasibility of Dialysate Bolus-Based Absolute Blood Volume Estimation in Maintenance Hemodialysis Patients
    Krenn, Simon
    Schmiedecker, Michael
    Schneditz, Daniel
    Hoedlmoser, Sebastian
    Mayer, Christopher C.
    Wassertheurer, Siegfried
    Omic, Haris
    Schernhammer, Eva
    Wabel, Peter
    Hecking, Manfred
    FRONTIERS IN MEDICINE, 2022, 9
  • [9] Can Bioimpedance Analysis Be Used to Estimate Absolute Blood Volume in Hemodialysis Patients?
    Kron, Joachim
    Volkenandt, John
    Broszeit, Stefanie
    Leimbach, Til
    Kron, Susanne
    HEMODIALYSIS INTERNATIONAL, 2025,
  • [10] Estimation of delivered dialysis dose by on-line monitoring of the ultraviolet absorbance in the spent dialysate
    Uhlin, F
    Fridolin, I
    Lindberg, LG
    Magnusson, M
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2003, 41 (05) : 1026 - 1036