On-Line Dialysate Infusion to Estimate Absolute Blood Volume in Dialysis Patients

被引:29
|
作者
Schneditz, Daniel [1 ]
Schilcher, Gernot [2 ]
Ribitsch, Werner [2 ]
Krisper, Peter [2 ]
Haditsch, Bernd [1 ]
Kron, Joachim [3 ]
机构
[1] Med Univ Graz, Inst Physiol, A-8010 Graz, Austria
[2] Med Univ Graz, Clin Div Nephrol, Dept Internal Med, A-8010 Graz, Austria
[3] KfH Kidney Ctr Berlin Kopenick, KfH Kuratorium Dialyse & Nierentransplantat, Berlin, Germany
关键词
ultra-pure dialysate; indicator dilution; volume kinetics; blood volume; half-life; TOTAL PROTEIN-CONCENTRATION; CARDIOPULMONARY RECIRCULATION; SOUND-SPEED; HEMODIALYSIS; HYPOTENSION; PRESSURE; HEMATOCRIT; DENSITY; FLUID;
D O I
10.1097/MAT.0000000000000086
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It was the aim to measure the distribution volume and the elimination of ultra-pure dialysate in stable hemodialysis patients during on-line hemodiafiltration (HDF). Dialysate was automatically infused as a volume indicator using standard on-line HDF equipment. Indicator concentration was noninvasively measured in the arterial blood-line (using the blood volume monitor, Fresenius Medical Care, Bad Homburg vor der H he, Germany), and its time course was analyzed to obtain the elimination rate and the distribution volume V-t at the time of dilution. Blood volume at treatment start (V-0) was calculated accounting for the degree of intradialytic hemoconcentration. Five patients (two females) were studied during 15 treatments. Two to six measurements using indicator volumes ranging from 60 to 210 ml were done in each treatment. V-0 was 4.59 +/- 1.15 L and larger than the volume of 4.08 +/- 0.48 L estimated from anthropometric relationships. The mean half-life of infused volume was 17.2 +/- 29.7 min. Given predialysis volume expansion V-0 was consistent with blood volume determined from anthropometric measurements. Information on blood volume could substantially improve volume management in hemodialysis patients and fluid therapy in intensive care patients undergoing extracorporeal blood treatment. The system has the potential for complete automation using proper control inputs for BVM and HDF modules of the dialysis machine.
引用
收藏
页码:436 / 442
页数:7
相关论文
共 50 条
  • [41] Post-Dilution on Line Haemodiafiltration with Citrate Dialysate: First Clinical Experience in Chronic Dialysis Patients
    Panichi, Vincenzo
    Fiaccadori, Enrico
    Rosati, Alberto
    Fanelli, Roberto
    Bernabini, Giada
    Scatena, Alessia
    Pizzarelli, Francesco
    SCIENTIFIC WORLD JOURNAL, 2013,
  • [42] ON-LINE VOLUME FLOW-RATE AND VELOCITY PROFILE MEASUREMENT FOR BLOOD IN MICROVESSELS
    BAKER, M
    WAYLAND, H
    MICROVASCULAR RESEARCH, 1974, 7 (01) : 131 - 143
  • [43] High volume on line hemodiafiltration in dialysis pregnant patients: predilutional or postdilutional?
    Puddu, Marcelo
    Migueliz, Maria Lucia
    Rosa-Diez, Guillermo
    Crucelegui, Soledad
    Martinez, Agustina Heredia
    Luxardo, Rosario
    Amalfi, Hector
    Egolf, Alicia
    Fernandez, Maria Isabel
    Nunez, Natalia
    Canaud, Bernard
    JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (03) : 392 - 397
  • [44] High volume on line hemodiafiltration in dialysis pregnant patients: predilutional or postdilutional?
    Marcelo Puddu
    María Lucía Migueliz
    Guillermo Rosa-Diez
    Soledad Crucelegui
    Agustina Heredia Martínez
    Rosario Luxardo
    Héctor Amalfi
    Alicia Egolf
    María Isabel Fernandez
    Natalia Nuñez
    Bernard Canaud
    Journal of Artificial Organs, 2021, 24 : 392 - 397
  • [45] Screening for sleep apnea syndrome in dialysis patients by on-line hematocrit monitoring.
    Magnone, M
    vonAlbertini, B
    Despland, PA
    Oudy, C
    Wauters, JP
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1997, 8 : A1127 - A1127
  • [46] Effect of hemodiafiltration with on-line regeneration of ultrafiltrate (HFR) on oxidative stress in dialysis patients
    Calo, Lorenzo
    Naso, Agostino
    Carraro, Gianni
    Wratten, Mary Lou
    Pagnin, Elisa
    Bertipaglia, Lara
    Piccoli, Antonio
    D'Angelo, Angela
    Cascone, Carmelo
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2006, 21 : 244 - 244
  • [47] Influence of Variable Inflow Volume, Dwell Duration, and Dialysate Glucose Concentration on Ultrafiltration Volume in Patients on Automated Peritoneal Dialysis: pPD Study
    De Los Rios, Tatiana
    Garcia, Isabel
    Schmitt, Jana
    Stauss-Grabo, Manuela
    Atiye, Saynab
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2024, 35 (10):
  • [48] Clinical Effectiveness of Intermittent Infusion Hemodiafiltration Using Backfiltration of Ultrapure Dialysis Fluid Compared with Predilution On-Line Hemodiafiltration
    Mineshima, Michio
    Eguchi, Kei
    Shishido, Kanji
    Takahashi, Susumu
    Kubo, Tsukasa
    Kawaguchi, Hiroshi
    Shitomi, Kouzo
    Shibagaki, Keigo
    Suga, Kiichi
    Nagao, Hirotomo
    Takada, Mikihiko
    Taoka, Masahiro
    Sato, Takashi
    SCIENTIFIC ASPECTS OF DIALYSIS THERAPY: JSDT/ISBP ANNIVERSARY EDITION, 2017, 189 : 24 - 29
  • [49] COMPARISON OF LEUKOCYTE (WBC) COUNTS (CTS) IN BLOOD AND DIALYSATE IN CHRONIC PERITONEAL-DIALYSIS PATIENTS WITH PERITONITIS
    VASSA, N
    NOLPH, KD
    PROWANT, BF
    MOORE, HL
    KHANNA, R
    TWARDOWSKI, ZJ
    CLINICAL RESEARCH, 1994, 42 (03): : A383 - A383
  • [50] Absolute Blood Volume in Hemodialysis Patients: Why Is It Relevant, and How to Measure It?
    Thijssen, Stephan
    Kappel, Franz
    Kotanko, Peter
    BLOOD PURIFICATION, 2013, 35 (1-3) : 63 - 71