Microstructure-based numerical computational method for the insertion torque of dental implant

被引:0
|
作者
Li, Luli [1 ,2 ]
Zhang, Song [1 ,2 ]
Li, Quhao [1 ,2 ]
Bian, Cuirong [3 ]
Zhang, Airong [1 ,2 ]
机构
[1] Shandong Univ, Sch Mech Engn, Jinan 250061, Shandong, Peoples R China
[2] Shandong Univ, Key Lab High Efficiency & Clean Mech Manufacture, Minist Educ, Jinan, Shandong, Peoples R China
[3] Shandong Univ, Dept Prosthodont, Qilu Hosp, Jinan 250012, Shandong, Peoples R China
关键词
Dental implant; Fabric tensor; Insertion torque; Microstructural solid model; Volume fraction; VERTEBRAL TRABECULAR BONE; LIMIT ANALYSIS; WOLFFS LAW; DENSITY; ARCHITECTURE; PREDICT; QUALITY; SITES;
D O I
10.1016/j.jmbbm.2019.06.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The bone quality has a significant effect on the insertion torque of dental implant. In most clinical studies, bone density is used as a gold standard in predicting insertion torque. By contrast, trabecular microstructure is ignored. In this study, a microstructure-based numerical computational method with high accuracy and efficiency for the insertion torque of dental implant was proposed by introducing two microscopic variables, namely, volume fraction and fabric tensor. First, two kinds of 3D microstructural solid models with same volume fraction and fabric tensor were established on the basis of the microstructural topology of six reference specimens. Second, a new numerical simulation method based on homogenous theory was used to explore the material models of these 3D microstructural solid models at the microscopic scale. Then, the anisotropic material models of specimens were developed on the basis of the mixture rule. Thereafter, a numerical simulation based on the anisotropic finite element (FE) model was carried out to acquire the insertion torque. To demonstrate the efficiency and accuracy of the simulation based on the anisotropic FE model, numerical simulations based on isotropic FE model and micro-computer tomography (micro-CT) FE models were also implemented as comparisons. Comparison of the simulated peak insertion torques of the anisotropic, isotropic, and micro-CT FE models with insertion experiments demonstrated the feasibility and potential of the proposed method. The anisotropic FE model reduced the time consumption by 91.85% and enhanced the accuracy by 11.82% compared with the micro-CT and isotropic FE models, respectively.
引用
收藏
页码:137 / 147
页数:11
相关论文
共 50 条
  • [31] Dental implant failure rates with low insertion torque with a nonsubmerged surgical approach: A retrospective clinical study
    Liu, Qiqi
    Yang, Jian
    Wang, Ruibin
    Yuan, Lingling
    Yu, Ke
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2023, 25 (01) : 118 - 123
  • [32] Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study
    Romero, Marta
    Herrero-Climent, Mariano
    Rios-Carrasco, Blanca
    Brizuela, Aritza
    Romero, Manuel Maria
    Gil, Javier
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (13)
  • [33] Advanced numerical method for generation of three-dimensional particles and its application in microstructure-based simulation of fatigue behavior
    Milesi, M.
    Bouchard, P. O.
    Cornuault, P. H.
    Chastel, Y.
    Bernacki, M.
    Hachem, E.
    COMPUTATIONAL MATERIALS SCIENCE, 2011, 50 (10) : 2836 - 2847
  • [34] Relevant Aspects of the Dental Implant Design on the Insertion Torque, Resonance Frequency Analysis (RFA) and Micromobility: An In Vitro Study
    Herrero-Climent, Mariano
    Falcao, Artur
    Tondela, Joao
    Brizuela, Aritza
    Rios-Carrasco, Blanca
    Gil, Javier
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (03)
  • [35] Development of a microstructure-based numerical approach for analyzing heat transfer within the asphalt mixture
    Mu, Ke
    Chen, Jianbing
    Mu, Qingyi
    Garg, Ankit
    Gao, Zhiwei
    Li, Yanwei
    CONSTRUCTION AND BUILDING MATERIALS, 2020, 260
  • [36] Numerical Analysis of the Microstructure-based Model for Layered Composites via MC and FEM Approaches
    Liu, Yunfang
    Zeng, Qingfeng
    Feng, Zhiqiang
    Cheng, Laifei
    Li, Liangjun
    Zhang, Litong
    BRAZILIAN JOURNAL OF PHYSICS, 2016, 46 (01) : 87 - 96
  • [37] Numerical Analysis of the Microstructure-based Model for Layered Composites via MC and FEM Approaches
    Yunfang Liu
    Qingfeng Zeng
    Zhiqiang Feng
    Laifei Cheng
    Liangjun Li
    Litong Zhang
    Brazilian Journal of Physics, 2016, 46 : 87 - 96
  • [39] Impact of dental implant insertion method on the periimplant epithelial seal experimental study on dogs
    Matic, Smiljana
    Stamatovic, Novak
    Lazic, Zoran
    Petkovic-Curcin, Aleksandra
    Bubalo, Marija
    Vojvodic, Danilo
    Djurdjevic, Dragan
    VOJNOSANITETSKI PREGLED, 2010, 67 (03) : 236 - 242
  • [40] Microstructure-based homogenization method for early-age creep of cement paste
    Yu, P.
    Duan, Y. H.
    Chen, E.
    Tang, S. W.
    Hanif, A.
    Fan, Y. L.