Background: The nasal cannula is considered a trusted and effective means of administering low-flow oxygen and is widely used for neonates and infants requiring oxygen therapy, despite an understanding that oxygen concentrations delivered to patients are variable. Methods: In the present study, realistic nasal airway replicas derived from medical scans of children less than 3 months old were used to measure the fraction of oxygen inhaled (FiO(2)) through nasal cannulas during low-flow oxygen delivery. Parameters influencing variability in FiO(2) were evaluated, as was the hypothesis that measured FiO(2) values could be predicted using a simple, flow-weighted calculation that assumes ideal mixing of oxygen with entrained room air. Tidal breathing through neonatal and infant nasal airway replicas was controlled using a lung simulator. Parameters for nasal cannula oxygen flow rate, nasal airway geometry, tidal volume, respiratory rate, inhalation/exhalation, or I:E ratio (t(i)/t(e)), breath waveform, and cannula prong insertion position were varied to determine their effect on measured FiO(2). In total, FiO(2) was measured for 384 different parameter combinations, with each combination repeated in triplicate. Analysis of variance (ANOVA) was used to assess the influence of parameters on measured FiO(2). Results: Measured FiO(2) was not appreciably affected by the breath waveform shape, the replica geometry, or the cannula position but was significantly influenced by the tidal volume, the inhalation time, and the nasal cannula flow rate. Conclusions: The flow-weighted calculation overpredicted FiO(2) for measured values above 60%, but an empirical correction to the calculation provided good agreement with measured FiO(2) across the full range of experimental data.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Nasal continuous positive airway pressure from high flow cannula versus Infant Flow for preterm infants
    D M Campbell
    P S Shah
    V Shah
    E N Kelly
    Journal of Perinatology, 2006, 26 : 546 - 549
  • [42] COMPARISON OF PULSED VS. CONTINUOUS OXYGEN DELIVERY USING REALISTIC ADULT NASAL AIRWAY REPLICAS
    Chen, John
    Martin, Andrew
    Finlay, Warren
    Katz, Ira
    Caillibotte, Georges
    Pichelin, Marine
    Noga, Michelle
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2017, 30 (03) : 9 - 9
  • [43] Non-rebreather mask and low-flow nasal cannula vs high-flow nasal cannula in severe COVID-19 pneumonia in the emergency department
    Kamil, Muhammad Khidir Mohd
    Yoong, Khadijah Poh Yuen
    Azhar, Abdul Muhaimin Noor
    Bustam, Aida
    Abdullah, Ahmad Hariz
    Yusuf, Mohd Hafyzuddin Md
    Zambri, Aliyah
    Zahedi, Ahmad Zulkarnain Ahmad
    Shafie, Hidayah
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2023, 63 : 86 - 93
  • [44] What are high-flow and low-flow oxygen delivery systems? Response
    Singhal, AB
    STROKE, 2005, 36 (10) : 2067 - 2067
  • [45] TRACHEAL GAS CONDITIONS DURING NASAL OXYGEN SUPPLEMENTATION: HEATED-HUMIDIFIED HIGH-FLOW NASAL OXYGEN VERSUS LOW-FLOW NASAL OXYGEN
    Kelly, Paul T.
    Jun, Ann
    Frampton, Chris
    Beckert, Lutz
    RESPIROLOGY, 2017, 22 : 10 - 10
  • [46] High flow nasal cannula as an oxygen delivery device during awake fibreoptic intubation
    Badiger, S.
    Fearnley, A.
    Ahmad, I.
    ANAESTHESIA, 2014, 69 : 12 - 12
  • [47] Simultaneous Administration Of Low Flow Nasal Cannula Oxygen Support And Pharmaceutical Aerosols
    Longest, P. W.
    Golshahi, L.
    Farkas, D.
    Tian, G.
    Behara, S. R. B.
    Hindle, M.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2015, 191
  • [48] High-flow nasal cannula oxygen therapy: more than a higher amount of oxygen delivery
    Maury, Eric
    Alves, Mikael
    Bige, Naike
    JOURNAL OF THORACIC DISEASE, 2016, 8 (10) : E1296 - E1300
  • [49] Airway management with Hi-flow nasal cannula oxygen in children with severe laryngeal obstruction
    Xiong, Ling
    Liu, Jianxia
    Li, Haisu
    Tan, Yanzhe
    Tang, Linlin
    Du, Min
    Xu, Ying
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2024, 176
  • [50] NEBULIZERS AND SPACERS FOR AEROSOL DELIVERY THROUGH ADULT NASAL CANNULA AT LOW OXYGEN FLOW RATE: AN IN-VITRO STUDY
    Madney, Yasmin M.
    Fathy, Maha
    Elberry, Ahmed A.
    Rabea, Hoda
    Abdelrahim, Mohamed
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2017, 30 (03) : 2 - 2